{"title":"Comparison of Volume Measurements and Bioimpedance Spectroscopy Using A Stand-on Device for Assessment of Unilateral Breast Cancer-Related Lymphedema.","authors":"L. C. Ward, Belinda Thompson, K. Gaitatzis, Louise A. Koelmeyer","doi":"10.4274/ejbh.galenos.2024.2023-12-8","DOIUrl":null,"url":null,"abstract":"Objective\nBreast cancer related lymphedema (BCRL) may be assessed through objective measurement of limb swelling with common techniques including volumetric measurement using a tape measure or perometry, and measurement of extracellular water using bioimpedance spectroscopy (BIS). This study aimed to evaluate the performance of a stand-on BIS device for detection of BCRL, introduce a novel graphical method to compare volumetric and BIS methods alongside traditional specificity and sensitivity analysis, and determine and compare BIS thresholds with those published previously.\n\n\nMaterials and Methods\nFemale participants with indocyanine green lymphography confirmed unilateral arm lymphedema (n = 197) and healthy controls (n = 267) were assessed using a cross-sectional study design. BIS and volumetric measures were obtained in a single session.\n\n\nResults\nThe BIS lymphedema index (L-Dex) method had a significantly higher sensitivity than the excess volume approach (area under the curve = 0.832 vs. 0.649, p = 0.0001). A threshold of L-Dex 6.5 had a higher true positive rate (70.6%) than L-Dex 10 (68.5%) although false positive rate increased from 0.4% to 2.6%. A threshold of 5% excess volume improved the true positive rate (68.5%) compared with 10% excess volume (49.7%) however the false positive rate increased to an unacceptable 47%. The L-Dex ranges in this study were not significantly different from previously published ranges.\n\n\nConclusion\nBIS was superior for identifying BCRL compared with volume measurements, reaffirming the value of this technique. However, it is recommended that BIS be used in conjunction with comprehensive evaluation of symptoms and clinical presentation. The proposed graphical method provides a simple and easily interpretable approach to compare and define concordance between the two commonly used methods for BCRL assessment namely limb volume and BIS L-Dex indices. The existing BIS (L-Dex) thresholds for presence of BCRL were also validated.","PeriodicalId":1,"journal":{"name":"Accounts of Chemical Research","volume":"70 21","pages":"141-148"},"PeriodicalIF":17.7000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Accounts of Chemical Research","FirstCategoryId":"0","ListUrlMain":"https://doi.org/10.4274/ejbh.galenos.2024.2023-12-8","RegionNum":1,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
引用次数: 0
Abstract
Objective
Breast cancer related lymphedema (BCRL) may be assessed through objective measurement of limb swelling with common techniques including volumetric measurement using a tape measure or perometry, and measurement of extracellular water using bioimpedance spectroscopy (BIS). This study aimed to evaluate the performance of a stand-on BIS device for detection of BCRL, introduce a novel graphical method to compare volumetric and BIS methods alongside traditional specificity and sensitivity analysis, and determine and compare BIS thresholds with those published previously.
Materials and Methods
Female participants with indocyanine green lymphography confirmed unilateral arm lymphedema (n = 197) and healthy controls (n = 267) were assessed using a cross-sectional study design. BIS and volumetric measures were obtained in a single session.
Results
The BIS lymphedema index (L-Dex) method had a significantly higher sensitivity than the excess volume approach (area under the curve = 0.832 vs. 0.649, p = 0.0001). A threshold of L-Dex 6.5 had a higher true positive rate (70.6%) than L-Dex 10 (68.5%) although false positive rate increased from 0.4% to 2.6%. A threshold of 5% excess volume improved the true positive rate (68.5%) compared with 10% excess volume (49.7%) however the false positive rate increased to an unacceptable 47%. The L-Dex ranges in this study were not significantly different from previously published ranges.
Conclusion
BIS was superior for identifying BCRL compared with volume measurements, reaffirming the value of this technique. However, it is recommended that BIS be used in conjunction with comprehensive evaluation of symptoms and clinical presentation. The proposed graphical method provides a simple and easily interpretable approach to compare and define concordance between the two commonly used methods for BCRL assessment namely limb volume and BIS L-Dex indices. The existing BIS (L-Dex) thresholds for presence of BCRL were also validated.
期刊介绍:
Accounts of Chemical Research presents short, concise and critical articles offering easy-to-read overviews of basic research and applications in all areas of chemistry and biochemistry. These short reviews focus on research from the author’s own laboratory and are designed to teach the reader about a research project. In addition, Accounts of Chemical Research publishes commentaries that give an informed opinion on a current research problem. Special Issues online are devoted to a single topic of unusual activity and significance.
Accounts of Chemical Research replaces the traditional article abstract with an article "Conspectus." These entries synopsize the research affording the reader a closer look at the content and significance of an article. Through this provision of a more detailed description of the article contents, the Conspectus enhances the article's discoverability by search engines and the exposure for the research.