{"title":"Effect of arm position on image quality and radiation dose during thorax and abdomen computed tomography scans","authors":"E.Ø. Buchhave , C.D. Rasmussen , R. Lentz , E.H. Frederiksen , J.O. Kipp , M.Ø. Leftes","doi":"10.1016/j.radi.2024.04.015","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p>During Computed Tomography (CT) scans of the Thorax-Abdomen-Pelvis (TAP) the patient's arms should be positioned above the head to obtain optimal image quality and expose the patient to the lowest possible radiation dose. This may be impossible with patients with shoulder problems leading to arms being positioned in other ways. This study aimed to investigate differences in objective image quality and estimated effective dose <em>(E</em>), when positioning the arms below shoulder level in CT-TAP.</p></div><div><h3>Methods</h3><p>An anthropomorphic phantom with cadaver arms was used. Four arm positions were tested: Along the torso (A), on the pelvis (B), on a pillow on the pelvis (C), and one arm on pillow on the pelvis and the other arm on the pelvis (D). A Siemens SOMATOM Definition Flash CT-scanner with CareDose 4D was used. The dose length product was read to estimate <em>E</em>. Image quality was assessed objectively by measuring noise within the region of interest in the liver and urinary bladder.</p></div><div><h3>Results</h3><p>Significant differences in <em>E</em> between all arm positions were seen (p = 0.005). The lowest <em>E</em> was obtained in position C, reducing <em>E</em> by 8.42%. Position A provided the best image quality but the highest <em>E.</em></p></div><div><h3>Conclusion</h3><p>This study showed no unequivocal optimal positioning of arms in CT-TAP. Position A provided the best object image quality, while position C yielded the lowest <em>E</em>. These results may impact the planning of diagnostic CT where positioning of arms may influence optimal image quality and radiation dose.</p></div><div><h3>Implication for practice</h3><p>This study illustrates tendencies for objective image quality and <em>E</em> when arms are positioned below shoulder level. Further research is needed to assess the clinical relevance with the diagnostic potential.</p></div>","PeriodicalId":47416,"journal":{"name":"Radiography","volume":null,"pages":null},"PeriodicalIF":2.5000,"publicationDate":"2024-05-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiography","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1078817424000920","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction
During Computed Tomography (CT) scans of the Thorax-Abdomen-Pelvis (TAP) the patient's arms should be positioned above the head to obtain optimal image quality and expose the patient to the lowest possible radiation dose. This may be impossible with patients with shoulder problems leading to arms being positioned in other ways. This study aimed to investigate differences in objective image quality and estimated effective dose (E), when positioning the arms below shoulder level in CT-TAP.
Methods
An anthropomorphic phantom with cadaver arms was used. Four arm positions were tested: Along the torso (A), on the pelvis (B), on a pillow on the pelvis (C), and one arm on pillow on the pelvis and the other arm on the pelvis (D). A Siemens SOMATOM Definition Flash CT-scanner with CareDose 4D was used. The dose length product was read to estimate E. Image quality was assessed objectively by measuring noise within the region of interest in the liver and urinary bladder.
Results
Significant differences in E between all arm positions were seen (p = 0.005). The lowest E was obtained in position C, reducing E by 8.42%. Position A provided the best image quality but the highest E.
Conclusion
This study showed no unequivocal optimal positioning of arms in CT-TAP. Position A provided the best object image quality, while position C yielded the lowest E. These results may impact the planning of diagnostic CT where positioning of arms may influence optimal image quality and radiation dose.
Implication for practice
This study illustrates tendencies for objective image quality and E when arms are positioned below shoulder level. Further research is needed to assess the clinical relevance with the diagnostic potential.
RadiographyRADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
4.70
自引率
34.60%
发文量
169
审稿时长
63 days
期刊介绍:
Radiography is an International, English language, peer-reviewed journal of diagnostic imaging and radiation therapy. Radiography is the official professional journal of the College of Radiographers and is published quarterly. Radiography aims to publish the highest quality material, both clinical and scientific, on all aspects of diagnostic imaging and radiation therapy and oncology.