Yaoling Li, Jing Li, Huiyu Wang, Mingyang Du, Lirong Wei, Teng Su, Gang Ding, Xuehan Qian, Ning Hua
{"title":"The Performance of Spot Photoscreener in 6 to 10 Weeks Infants in China: A Cross-Sectional Study","authors":"Yaoling Li, Jing Li, Huiyu Wang, Mingyang Du, Lirong Wei, Teng Su, Gang Ding, Xuehan Qian, Ning Hua","doi":"10.1155/2024/8817530","DOIUrl":null,"url":null,"abstract":"<i>Purpose</i>. To compare the refractive errors measured by the Spot photoscreener (with or without cycloplegia) to cycloplegic retinoscopy in 6- to 10-week-old infants. <i>Materials and Methods</i>. 101 right eyes from 101 healthy infants aged 6 to 10 weeks were recruited for this cross-sectional observational study. Refractive errors were measured using Spot photoscreener before and after cycloplegia, as well as cycloplegic retinoscopy. Comparisons between the refractive measurements were performed using one-way ANOVA with the post hoc Tukey HSD test or Kruskal–Wallis test with the Steel–Dwass test according to the data normality. Pearson’s correlation test and 95% confidence intervals were calculated. The agreement was evaluated using a Bland–Altman plot with 95% limits of agreement of the differences. <i>Results</i>. Spot photoscreener was found to underestimate the spherical equivalent by 2.33 Diopters (D) in these infants. Following the induction of cycloplegia, the spherical equivalent measured by Spot photoscreener was in excellent agreement with cycloplegic retinoscopy with the mean difference of 0.01 D. Spot photoscreener overestimated cylindrical parameter by 0.2 D with poor agreement with cycloplegic retinoscopy no matter whether cycloplegia was induced. It had good agreement with cycloplegic retinoscopy in the <svg height=\"11.927pt\" style=\"vertical-align:-3.291101pt\" version=\"1.1\" viewbox=\"-0.0498162 -8.6359 10.0959 11.927\" width=\"10.0959pt\" xmlns=\"http://www.w3.org/2000/svg\" xmlns:xlink=\"http://www.w3.org/1999/xlink\"><g transform=\"matrix(.013,0,0,-0.013,0,0)\"></path></g><g transform=\"matrix(.0091,0,0,-0.0091,5.031,3.132)\"></path></g></svg> vector than the <svg height=\"11.927pt\" style=\"vertical-align:-3.291101pt\" version=\"1.1\" viewbox=\"-0.0498162 -8.6359 14.5442 11.927\" width=\"14.5442pt\" xmlns=\"http://www.w3.org/2000/svg\" xmlns:xlink=\"http://www.w3.org/1999/xlink\"><g transform=\"matrix(.013,0,0,-0.013,0,0)\"><use xlink:href=\"#g113-75\"></use></g><g transform=\"matrix(.0091,0,0,-0.0091,5.031,3.132)\"></path></g><g transform=\"matrix(.0091,0,0,-0.0091,9.463,3.132)\"></path></g></svg> vector measurement. <i>Conclusions</i>. With the induction of cycloplegia, Spot photoscreener can accurately evaluate spherical equivalent in hyperopic infants with mild-to-moderate astigmatism. While it may provide valuable measurements of astigmatism, discrepancies in cylinder and axis should be taken into account.","PeriodicalId":16674,"journal":{"name":"Journal of Ophthalmology","volume":"21 1","pages":""},"PeriodicalIF":1.8000,"publicationDate":"2024-05-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1155/2024/8817530","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose. To compare the refractive errors measured by the Spot photoscreener (with or without cycloplegia) to cycloplegic retinoscopy in 6- to 10-week-old infants. Materials and Methods. 101 right eyes from 101 healthy infants aged 6 to 10 weeks were recruited for this cross-sectional observational study. Refractive errors were measured using Spot photoscreener before and after cycloplegia, as well as cycloplegic retinoscopy. Comparisons between the refractive measurements were performed using one-way ANOVA with the post hoc Tukey HSD test or Kruskal–Wallis test with the Steel–Dwass test according to the data normality. Pearson’s correlation test and 95% confidence intervals were calculated. The agreement was evaluated using a Bland–Altman plot with 95% limits of agreement of the differences. Results. Spot photoscreener was found to underestimate the spherical equivalent by 2.33 Diopters (D) in these infants. Following the induction of cycloplegia, the spherical equivalent measured by Spot photoscreener was in excellent agreement with cycloplegic retinoscopy with the mean difference of 0.01 D. Spot photoscreener overestimated cylindrical parameter by 0.2 D with poor agreement with cycloplegic retinoscopy no matter whether cycloplegia was induced. It had good agreement with cycloplegic retinoscopy in the vector than the vector measurement. Conclusions. With the induction of cycloplegia, Spot photoscreener can accurately evaluate spherical equivalent in hyperopic infants with mild-to-moderate astigmatism. While it may provide valuable measurements of astigmatism, discrepancies in cylinder and axis should be taken into account.
期刊介绍:
Journal of Ophthalmology is a peer-reviewed, Open Access journal that publishes original research articles, review articles, and clinical studies related to the anatomy, physiology and diseases of the eye. Submissions should focus on new diagnostic and surgical techniques, instrument and therapy updates, as well as clinical trials and research findings.