{"title":"Percutaneous Surgery for Trigger Finger Treatment Using a Novel Surgical Device: An Experimental Study on Fresh Cadavers","authors":"","doi":"10.1016/j.jhsg.2024.03.002","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><p>Trigger finger, a stenosing tenosynovitis of the flexor tendon at the A1 pulley, can cause pain and impair daily activities. Despite common surgical interventions, postsurgical complications are frequent, prompting the search for less invasive techniques.</p></div><div><h3>Methods</h3><p>An experimental study was conducted on fresh cadavers to compare three techniques: the first using a PulleyCut without ultrasound guidance, the second using a PulleyCut with ultrasound guidance, and the third using a percutaneous needle technique. The complete release of the A1 pulley, integrity of the A2 pulley, flexor tendons, and neurovascular bundles were assessed.</p></div><div><h3>Results</h3><p>The new device group and the ultrasound-guided group demonstrated 100% complete release of the A1 pulley, whereas the needle group achieved only 38% success. There were no A2 pulley injuries in any group. Flexor tendons were injured in 7% of cases in the new device group and 77% in the needle group. A neurovascular injury occurred in the needle group.</p></div><div><h3>Conclusions</h3><p>Compared with the percutaneous needle technique, the new device proved safe and effective for A1 pulley release, minimizing damage to flexor tendons and neurovascular structures. Ultrasound did not provide significant advantages, suggesting that the new device can be confidently used without ultrasound assistance. The PulleyCut represents a promising percutaneous technique for trigger finger treatment, demonstrating superiority over the needle technique in terms of efficacy and safety. These results encourage future clinical investigations to validate its practical application.</p></div><div><h3>Type of study/level of evidence</h3><p>Therapeutic IIc.</p></div>","PeriodicalId":36920,"journal":{"name":"Journal of Hand Surgery Global Online","volume":"6 4","pages":"Pages 494-499"},"PeriodicalIF":0.0000,"publicationDate":"2024-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2589514124000562/pdfft?md5=7cb43ccc44cc2831a94bc6200f62e186&pid=1-s2.0-S2589514124000562-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Hand Surgery Global Online","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2589514124000562","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose

Trigger finger, a stenosing tenosynovitis of the flexor tendon at the A1 pulley, can cause pain and impair daily activities. Despite common surgical interventions, postsurgical complications are frequent, prompting the search for less invasive techniques.

Methods

An experimental study was conducted on fresh cadavers to compare three techniques: the first using a PulleyCut without ultrasound guidance, the second using a PulleyCut with ultrasound guidance, and the third using a percutaneous needle technique. The complete release of the A1 pulley, integrity of the A2 pulley, flexor tendons, and neurovascular bundles were assessed.

Results

The new device group and the ultrasound-guided group demonstrated 100% complete release of the A1 pulley, whereas the needle group achieved only 38% success. There were no A2 pulley injuries in any group. Flexor tendons were injured in 7% of cases in the new device group and 77% in the needle group. A neurovascular injury occurred in the needle group.

Conclusions

Compared with the percutaneous needle technique, the new device proved safe and effective for A1 pulley release, minimizing damage to flexor tendons and neurovascular structures. Ultrasound did not provide significant advantages, suggesting that the new device can be confidently used without ultrasound assistance. The PulleyCut represents a promising percutaneous technique for trigger finger treatment, demonstrating superiority over the needle technique in terms of efficacy and safety. These results encourage future clinical investigations to validate its practical application.

Type of study/level of evidence

Therapeutic IIc.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
使用新型手术装置进行经皮手术治疗扳机指:新鲜尸体实验研究
目的扳机指是 A1 滑轮处屈肌腱的一种狭窄性腱鞘炎,可引起疼痛并影响日常活动。方法在新鲜尸体上进行了一项实验研究,比较了三种技术:第一种是在没有超声引导的情况下使用 PulleyCut,第二种是在超声引导下使用 PulleyCut,第三种是使用经皮穿刺针技术。对 A1 滑轮的完全松解、A2 滑轮的完整性、屈肌腱和神经血管束进行了评估。结果 新装置组和超声引导组 100%完全松解了 A1 滑轮,而针组仅成功了 38%。各组均无 A2 滑轮损伤。新装置组中有 7% 的病例损伤了屈肌腱,而针组中有 77% 的病例损伤了屈肌腱。结论与经皮穿刺针技术相比,新装置在 A1 滑轮松解术中安全有效,最大程度地减少了对屈肌腱和神经血管结构的损伤。超声波的优势并不明显,这表明新装置无需超声波辅助即可放心使用。PulleyCut 是一种很有前景的经皮治疗扳机指技术,在疗效和安全性方面都优于针刺技术。研究类型/证据级别治疗 IIc.
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
1.10
自引率
0.00%
发文量
111
审稿时长
12 weeks
期刊最新文献
Table of Contents Scaphoid Fractures and Nonunion: A Survey-based Review of Hand Surgeon’s Practice and the Evidence Hook of Hamate Regrowth After Surgical Excision: A Report of Two Cases Possible Bupivacaine Induced Hepatitis Following Wide Awake, Local Anesthesia, No Tourniquet Carpal Tunnel Surgery: A Case Report Digital Papillary Adenocarcinoma: A Case Presentation and Literature Review
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1