{"title":"Imaging of low Z masked with high Z (Pb, U) materials using 14 MeV neutron","authors":"S. Bishnoi, T. Patel, P.S. Sarkar, L. Pant","doi":"10.1088/1748-0221/19/05/p05022","DOIUrl":null,"url":null,"abstract":"\n An experimental study has been performed using 14 MeV neutrons for imaging of low Z material (particularly composed of C, H, O elements) masked with thick layers of dense and high Z materials. The experimental setup consists of a D-T neutron generator, a metallic collimator and an imaging system. The imaging system is designed with a polypropylene zinc sulphide scintillator screen integrated with a lens coupled 16-bit ICCD camera. Imaging capability of the system was investigated using iron test samples with holes and line pair features. The minimum hole size of 2 mm could be imaged at a contrast of 36% and a line of 2 mm width visible at a contrast of 24% indicating the system's resolution of ∼ mm. Low Z samples such as water (H2O) and polyethylene (C2H2)\n n\n placed behind thick layers of Pb (40 mm) and Uranium (35 mm), were imaged successfully. These images reveal the system's ability towards low Z material imaging in the presence of heavier metals. Good contrast images acquired at a lower neutron yield of ∼ 5 × 108 n/sec of D-T neutron generator has provided a possibility to realise fast neutron imaging having moderate resolution (∼ mm) with a smaller footprint and an economical system design for field applications.","PeriodicalId":16184,"journal":{"name":"Journal of Instrumentation","volume":null,"pages":null},"PeriodicalIF":1.3000,"publicationDate":"2024-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Instrumentation","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1088/1748-0221/19/05/p05022","RegionNum":4,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0
Abstract
An experimental study has been performed using 14 MeV neutrons for imaging of low Z material (particularly composed of C, H, O elements) masked with thick layers of dense and high Z materials. The experimental setup consists of a D-T neutron generator, a metallic collimator and an imaging system. The imaging system is designed with a polypropylene zinc sulphide scintillator screen integrated with a lens coupled 16-bit ICCD camera. Imaging capability of the system was investigated using iron test samples with holes and line pair features. The minimum hole size of 2 mm could be imaged at a contrast of 36% and a line of 2 mm width visible at a contrast of 24% indicating the system's resolution of ∼ mm. Low Z samples such as water (H2O) and polyethylene (C2H2)
n
placed behind thick layers of Pb (40 mm) and Uranium (35 mm), were imaged successfully. These images reveal the system's ability towards low Z material imaging in the presence of heavier metals. Good contrast images acquired at a lower neutron yield of ∼ 5 × 108 n/sec of D-T neutron generator has provided a possibility to realise fast neutron imaging having moderate resolution (∼ mm) with a smaller footprint and an economical system design for field applications.
期刊介绍:
Journal of Instrumentation (JINST) covers major areas related to concepts and instrumentation in detector physics, accelerator science and associated experimental methods and techniques, theory, modelling and simulations. The main subject areas include.
-Accelerators: concepts, modelling, simulations and sources-
Instrumentation and hardware for accelerators: particles, synchrotron radiation, neutrons-
Detector physics: concepts, processes, methods, modelling and simulations-
Detectors, apparatus and methods for particle, astroparticle, nuclear, atomic, and molecular physics-
Instrumentation and methods for plasma research-
Methods and apparatus for astronomy and astrophysics-
Detectors, methods and apparatus for biomedical applications, life sciences and material research-
Instrumentation and techniques for medical imaging, diagnostics and therapy-
Instrumentation and techniques for dosimetry, monitoring and radiation damage-
Detectors, instrumentation and methods for non-destructive tests (NDT)-
Detector readout concepts, electronics and data acquisition methods-
Algorithms, software and data reduction methods-
Materials and associated technologies, etc.-
Engineering and technical issues.
JINST also includes a section dedicated to technical reports and instrumentation theses.