Tony Realini, Kevin Halenda, Joel Palko, Brian McMillan, Goundappa K Balasubramani
{"title":"Test-Retest Reliability of Intraocular Pressure Measurements With Office-Based Versus Home-Based Rebound Tonometers.","authors":"Tony Realini, Kevin Halenda, Joel Palko, Brian McMillan, Goundappa K Balasubramani","doi":"10.1097/IJG.0000000000002441","DOIUrl":null,"url":null,"abstract":"<p><strong>Prcis: </strong>Intraocular pressure (IOP) measurements obtained with the HOME2 rebound tonometer are interchangeable with measurements obtained using the office-based IC100 and IC200 rebound tonometers, making characterization of circadian IOP achievable with no inter-device sources of variability.</p><p><strong>Objective: </strong>To evaluate test-retest reliability between IOP measurements obtained with a home rebound tonometer operated by patients and those obtained with office rebound tonometers operated by experienced clinical staff.</p><p><strong>Methods: </strong>After training and practice with a home rebound tonometer (iCare HOME2) in a randomly selected eye, patients with established or suspected open angle glaucoma underwent IOP measurement in the fellow eye as follows in randomized order: 3 good quality measurements with the iCare IC100 operated by experienced clinical staff, 3 good quality measurements with the iCare IC200 operated by experienced clinical staff, and 3 good quality measurements with the HOME2 operated by the patient. The Shrout-Fleisch intraclass correlation coefficient (ICC; 2,1) was the measure of reliability utilized based on the model using 2-way random effects, absolute agreement, and single measurement.</p><p><strong>Results: </strong>Thirty eyes of 30 patients were evaluated. The Pearson correlation coefficient ( r ) for IOP measurements taken with the IC100 and HOME2 was 0.97, with the IC200 and HOME2 was 0.96, and with the IC100 and IC200 was 0.97. Test-retest reliability between HOME2 and IC100 (ICC: 0.90), HOME2 and IC200 (ICC: 0.92), and IC100 and IC200 (ICC: 0.94) all reflected excellent test-retest reliability.</p><p><strong>Conclusion: </strong>The test-retest reliability of IOP measurements taken by patients using the iCare HOME2 self-tonometer compared with IOP measurements taken by trained operators using the iCare IC100 and IC200 tonometers was uniformly excellent (ICCs all ≥0.9). These findings indicate that these three devices should be considered interchangeable for the clinical assessment of circadian IOP.</p>","PeriodicalId":15938,"journal":{"name":"Journal of Glaucoma","volume":" ","pages":"758-762"},"PeriodicalIF":2.0000,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Glaucoma","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/IJG.0000000000002441","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/5/27 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Prcis: Intraocular pressure (IOP) measurements obtained with the HOME2 rebound tonometer are interchangeable with measurements obtained using the office-based IC100 and IC200 rebound tonometers, making characterization of circadian IOP achievable with no inter-device sources of variability.
Objective: To evaluate test-retest reliability between IOP measurements obtained with a home rebound tonometer operated by patients and those obtained with office rebound tonometers operated by experienced clinical staff.
Methods: After training and practice with a home rebound tonometer (iCare HOME2) in a randomly selected eye, patients with established or suspected open angle glaucoma underwent IOP measurement in the fellow eye as follows in randomized order: 3 good quality measurements with the iCare IC100 operated by experienced clinical staff, 3 good quality measurements with the iCare IC200 operated by experienced clinical staff, and 3 good quality measurements with the HOME2 operated by the patient. The Shrout-Fleisch intraclass correlation coefficient (ICC; 2,1) was the measure of reliability utilized based on the model using 2-way random effects, absolute agreement, and single measurement.
Results: Thirty eyes of 30 patients were evaluated. The Pearson correlation coefficient ( r ) for IOP measurements taken with the IC100 and HOME2 was 0.97, with the IC200 and HOME2 was 0.96, and with the IC100 and IC200 was 0.97. Test-retest reliability between HOME2 and IC100 (ICC: 0.90), HOME2 and IC200 (ICC: 0.92), and IC100 and IC200 (ICC: 0.94) all reflected excellent test-retest reliability.
Conclusion: The test-retest reliability of IOP measurements taken by patients using the iCare HOME2 self-tonometer compared with IOP measurements taken by trained operators using the iCare IC100 and IC200 tonometers was uniformly excellent (ICCs all ≥0.9). These findings indicate that these three devices should be considered interchangeable for the clinical assessment of circadian IOP.
期刊介绍:
The Journal of Glaucoma is a peer reviewed journal addressing the spectrum of issues affecting definition, diagnosis, and management of glaucoma and providing a forum for lively and stimulating discussion of clinical, scientific, and socioeconomic factors affecting care of glaucoma patients.