Clinical audit of the image quality and customised contrast volume using P3T contrast injection software versus standard injection protocol in CT coronary angiography
N. Jayamani , S. Pothiawala , H.B. Ong , A.S. Low Choon Seng , A. Mohamed Afif , Z. Arumugam , C.T. Sung , F.C. Teck , H.C. Liang
{"title":"Clinical audit of the image quality and customised contrast volume using P3T contrast injection software versus standard injection protocol in CT coronary angiography","authors":"N. Jayamani , S. Pothiawala , H.B. Ong , A.S. Low Choon Seng , A. Mohamed Afif , Z. Arumugam , C.T. Sung , F.C. Teck , H.C. Liang","doi":"10.1016/j.radi.2024.05.009","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p>The implications of shorter scan time and lower tube voltage in the dual-source CT coronary angiography (CTCA) scan protocol necessitate the adaptation of contrast media (CM) injection parameters. This audit evaluates the coronary arteries' vascular attenuation and image quality by comparing the personalised patient protocol technology (P3T) contrast injection software with standard injection protocol. The secondary aim is to determine the relationship between CM volume and the patient's weight.</p></div><div><h3>Methodology</h3><p>A Siemens Somatom Definition Force CT Unit was used to scan 30 sets of patients between August 2020 and October 2020. Patients were selected retrospectively and separated into Standard Injection and P3T injection protocols. An experienced radiologist blinded to the groups reviewed the coronary vessels' contrast enhancement and image quality.</p></div><div><h3>Results</h3><p>Overall, the mean HU of all the main coronary artery vessels obtained from P3T injection software reached above 350 HU and was diagnostically sufficient. The mean attenuation at the proximal region of RCA in the 80–99 kg weight category was significantly higher in the P3T injection software than the standard injection protocol (p < 0.001). The CM volume proposed by P3T injection software for 40–59 kg was approximately 57 <span><math><mrow><mo>±</mo></mrow></math></span> 5 mls, while 75 ml was used for the standard injection protocol.</p></div><div><h3>Conclusion</h3><p>P3T injection software in CTCA resulted in an adequate diagnostic attenuation of coronary arteries (>350HU) in all weight groups, most effectively in the higher weight group, while maintaining diagnostic image quality. Further, the P3T software reduces CM volumes in lower-weight patients.</p></div><div><h3>Implications</h3><p>P3T software enables reducing CM volume in lower-weight patients while improving vascular enhancement in CTCA scans in higher-weight patients.</p></div>","PeriodicalId":47416,"journal":{"name":"Radiography","volume":null,"pages":null},"PeriodicalIF":2.5000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiography","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1078817424001408","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction
The implications of shorter scan time and lower tube voltage in the dual-source CT coronary angiography (CTCA) scan protocol necessitate the adaptation of contrast media (CM) injection parameters. This audit evaluates the coronary arteries' vascular attenuation and image quality by comparing the personalised patient protocol technology (P3T) contrast injection software with standard injection protocol. The secondary aim is to determine the relationship between CM volume and the patient's weight.
Methodology
A Siemens Somatom Definition Force CT Unit was used to scan 30 sets of patients between August 2020 and October 2020. Patients were selected retrospectively and separated into Standard Injection and P3T injection protocols. An experienced radiologist blinded to the groups reviewed the coronary vessels' contrast enhancement and image quality.
Results
Overall, the mean HU of all the main coronary artery vessels obtained from P3T injection software reached above 350 HU and was diagnostically sufficient. The mean attenuation at the proximal region of RCA in the 80–99 kg weight category was significantly higher in the P3T injection software than the standard injection protocol (p < 0.001). The CM volume proposed by P3T injection software for 40–59 kg was approximately 57 5 mls, while 75 ml was used for the standard injection protocol.
Conclusion
P3T injection software in CTCA resulted in an adequate diagnostic attenuation of coronary arteries (>350HU) in all weight groups, most effectively in the higher weight group, while maintaining diagnostic image quality. Further, the P3T software reduces CM volumes in lower-weight patients.
Implications
P3T software enables reducing CM volume in lower-weight patients while improving vascular enhancement in CTCA scans in higher-weight patients.
RadiographyRADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
4.70
自引率
34.60%
发文量
169
审稿时长
63 days
期刊介绍:
Radiography is an International, English language, peer-reviewed journal of diagnostic imaging and radiation therapy. Radiography is the official professional journal of the College of Radiographers and is published quarterly. Radiography aims to publish the highest quality material, both clinical and scientific, on all aspects of diagnostic imaging and radiation therapy and oncology.