Clinical audit of the image quality and customised contrast volume using P3T contrast injection software versus standard injection protocol in CT coronary angiography

IF 2.5 Q2 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Radiography Pub Date : 2024-06-01 DOI:10.1016/j.radi.2024.05.009
N. Jayamani , S. Pothiawala , H.B. Ong , A.S. Low Choon Seng , A. Mohamed Afif , Z. Arumugam , C.T. Sung , F.C. Teck , H.C. Liang
{"title":"Clinical audit of the image quality and customised contrast volume using P3T contrast injection software versus standard injection protocol in CT coronary angiography","authors":"N. Jayamani ,&nbsp;S. Pothiawala ,&nbsp;H.B. Ong ,&nbsp;A.S. Low Choon Seng ,&nbsp;A. Mohamed Afif ,&nbsp;Z. Arumugam ,&nbsp;C.T. Sung ,&nbsp;F.C. Teck ,&nbsp;H.C. Liang","doi":"10.1016/j.radi.2024.05.009","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p>The implications of shorter scan time and lower tube voltage in the dual-source CT coronary angiography (CTCA) scan protocol necessitate the adaptation of contrast media (CM) injection parameters. This audit evaluates the coronary arteries' vascular attenuation and image quality by comparing the personalised patient protocol technology (P3T) contrast injection software with standard injection protocol. The secondary aim is to determine the relationship between CM volume and the patient's weight.</p></div><div><h3>Methodology</h3><p>A Siemens Somatom Definition Force CT Unit was used to scan 30 sets of patients between August 2020 and October 2020. Patients were selected retrospectively and separated into Standard Injection and P3T injection protocols. An experienced radiologist blinded to the groups reviewed the coronary vessels' contrast enhancement and image quality.</p></div><div><h3>Results</h3><p>Overall, the mean HU of all the main coronary artery vessels obtained from P3T injection software reached above 350 HU and was diagnostically sufficient. The mean attenuation at the proximal region of RCA in the 80–99 kg weight category was significantly higher in the P3T injection software than the standard injection protocol (p &lt; 0.001). The CM volume proposed by P3T injection software for 40–59 kg was approximately 57 <span><math><mrow><mo>±</mo></mrow></math></span> 5 mls, while 75 ml was used for the standard injection protocol.</p></div><div><h3>Conclusion</h3><p>P3T injection software in CTCA resulted in an adequate diagnostic attenuation of coronary arteries (&gt;350HU) in all weight groups, most effectively in the higher weight group, while maintaining diagnostic image quality. Further, the P3T software reduces CM volumes in lower-weight patients.</p></div><div><h3>Implications</h3><p>P3T software enables reducing CM volume in lower-weight patients while improving vascular enhancement in CTCA scans in higher-weight patients.</p></div>","PeriodicalId":47416,"journal":{"name":"Radiography","volume":null,"pages":null},"PeriodicalIF":2.5000,"publicationDate":"2024-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiography","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1078817424001408","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction

The implications of shorter scan time and lower tube voltage in the dual-source CT coronary angiography (CTCA) scan protocol necessitate the adaptation of contrast media (CM) injection parameters. This audit evaluates the coronary arteries' vascular attenuation and image quality by comparing the personalised patient protocol technology (P3T) contrast injection software with standard injection protocol. The secondary aim is to determine the relationship between CM volume and the patient's weight.

Methodology

A Siemens Somatom Definition Force CT Unit was used to scan 30 sets of patients between August 2020 and October 2020. Patients were selected retrospectively and separated into Standard Injection and P3T injection protocols. An experienced radiologist blinded to the groups reviewed the coronary vessels' contrast enhancement and image quality.

Results

Overall, the mean HU of all the main coronary artery vessels obtained from P3T injection software reached above 350 HU and was diagnostically sufficient. The mean attenuation at the proximal region of RCA in the 80–99 kg weight category was significantly higher in the P3T injection software than the standard injection protocol (p < 0.001). The CM volume proposed by P3T injection software for 40–59 kg was approximately 57 ± 5 mls, while 75 ml was used for the standard injection protocol.

Conclusion

P3T injection software in CTCA resulted in an adequate diagnostic attenuation of coronary arteries (>350HU) in all weight groups, most effectively in the higher weight group, while maintaining diagnostic image quality. Further, the P3T software reduces CM volumes in lower-weight patients.

Implications

P3T software enables reducing CM volume in lower-weight patients while improving vascular enhancement in CTCA scans in higher-weight patients.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
在 CT 冠状动脉造影中使用 P3T 造影剂注射软件和标准注射方案,对图像质量和定制造影剂量进行临床审核。
导言:在双源 CT 冠状动脉造影术(CTCA)扫描方案中,由于扫描时间缩短和管电压降低,造影剂(CM)注射参数必须进行调整。本次审核通过比较个性化患者协议技术(P3T)造影剂注射软件和标准注射协议,评估冠状动脉的血管衰减和图像质量。次要目的是确定 CM 容量与患者体重之间的关系:在 2020 年 8 月至 2020 年 10 月期间,使用西门子 Somatom Definition Force CT 设备扫描了 30 组患者。对患者进行回顾性筛选,并将其分为标准注射方案和 P3T 注射方案。一位经验丰富的放射科医生对各组患者的冠状动脉血管造影剂增强情况和图像质量进行了盲法审查:总体而言,P3T 注射软件获得的所有主要冠状动脉血管的平均 HU 均达到 350 HU 以上,足以进行诊断。与标准注射方案相比,P3T 注射软件在体重为 80-99 公斤组的 RCA 近端区域的平均衰减明显更高(p 结论:P3T 注射软件在 CTC 中的应用可大大提高图像质量:在 CTCA 中使用 P3T 注射软件可在保持诊断图像质量的前提下,对所有体重组的冠状动脉进行适当的诊断衰减(>350HU),其中高体重组的效果最佳。此外,P3T 软件还能减少体重较轻患者的冠状动脉体积:意义:P3T 软件可减少体重较轻患者的 CM 容量,同时改善体重较重患者 CTCA 扫描的血管增强效果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Radiography
Radiography RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
4.70
自引率
34.60%
发文量
169
审稿时长
63 days
期刊介绍: Radiography is an International, English language, peer-reviewed journal of diagnostic imaging and radiation therapy. Radiography is the official professional journal of the College of Radiographers and is published quarterly. Radiography aims to publish the highest quality material, both clinical and scientific, on all aspects of diagnostic imaging and radiation therapy and oncology.
期刊最新文献
CT metal artefact reduction for hip and shoulder implants using novel algorithms and machine learning: A systematic review with pairwise and network meta-analyses Clinical insights into cross-sectional imaging integration in radiography education Paediatric diagnostic reference levels for common computed tomography procedures: A systematic review “Making it work in the face of extreme adversity” - Exploring perceptions for the future of the imaging and oncology workforce using ‘soundbite’ interviews Radiographer Education and Learning in Artificial Intelligence (REAL-AI): A survey of radiographers, radiologists, and students’ knowledge of and attitude to education on AI
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1