{"title":"Improving Visualization of In-stent Lumen Using Prototype Photon-counting Detector Computed Tomography with High-resolution Plaque Kernel.","authors":"Yoshinori Funama, Seitaro Oda, Fuyuhiko Teramoto, Yuko Aoki, Isao Takahashi, Shinichi Kojima, Taiga Goto, Kana Tanaka, Masafumi Kidoh, Yasunori Nagayama, Takeshi Nakaura, Toshinori Hirai","doi":"10.4103/jmp.jmp_163_23","DOIUrl":null,"url":null,"abstract":"<p><p>The study aimed to compare the performance of photon-counting detector computed tomography (PCD CT) with high-resolution (HR)-plaque kernel with that of the energy-integrating detector CT (EID CT) in terms of the visualization of the lumen size and the in-stent stenotic portion at different coronary vessel angles. The lumen sizes in PCD CT and EID CT images were 2.13 and 1.80 mm at 0°, 2.20 and 1.77 mm at 45°, and 2.27 mm and 1.67 mm at 90°, respectively. The lumen sizes in PCD CT with HR-plaque kernel were wider than those in EID CT. The mean degree of the in-stent stenotic portion at 50% was 69.7% for PCD CT and 90.4% for EID CT. PCD CT images with HR-plaque kernel enable improved visualization of lumen size and accurate measurements of the in-stent stenotic portion compared to conventional EID CT images regardless of the stent direction.</p>","PeriodicalId":51719,"journal":{"name":"Journal of Medical Physics","volume":"49 1","pages":"127-132"},"PeriodicalIF":0.7000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11141743/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jmp.jmp_163_23","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/3/30 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
The study aimed to compare the performance of photon-counting detector computed tomography (PCD CT) with high-resolution (HR)-plaque kernel with that of the energy-integrating detector CT (EID CT) in terms of the visualization of the lumen size and the in-stent stenotic portion at different coronary vessel angles. The lumen sizes in PCD CT and EID CT images were 2.13 and 1.80 mm at 0°, 2.20 and 1.77 mm at 45°, and 2.27 mm and 1.67 mm at 90°, respectively. The lumen sizes in PCD CT with HR-plaque kernel were wider than those in EID CT. The mean degree of the in-stent stenotic portion at 50% was 69.7% for PCD CT and 90.4% for EID CT. PCD CT images with HR-plaque kernel enable improved visualization of lumen size and accurate measurements of the in-stent stenotic portion compared to conventional EID CT images regardless of the stent direction.
期刊介绍:
JOURNAL OF MEDICAL PHYSICS is the official journal of Association of Medical Physicists of India (AMPI). The association has been bringing out a quarterly publication since 1976. Till the end of 1993, it was known as Medical Physics Bulletin, which then became Journal of Medical Physics. The main objective of the Journal is to serve as a vehicle of communication to highlight all aspects of the practice of medical radiation physics. The areas covered include all aspects of the application of radiation physics to biological sciences, radiotherapy, radiodiagnosis, nuclear medicine, dosimetry and radiation protection. Papers / manuscripts dealing with the aspects of physics related to cancer therapy / radiobiology also fall within the scope of the journal.