Concordance between platelet counts by impedance and optical technique during microcytic anemia: towards a threshold value of the mean corpuscular volume.

Q3 Medicine Tunisie Medicale Pub Date : 2024-06-05 DOI:10.62438/tunismed.v102i6.4834
Mayssa Ben Ayed, Bochra Gharbi, Eya Jaouachi, Hela Baccouche, Sonia Mahjoub, Aya Chakroun
{"title":"Concordance between platelet counts by impedance and optical technique during microcytic anemia: towards a threshold value of the mean corpuscular volume.","authors":"Mayssa Ben Ayed, Bochra Gharbi, Eya Jaouachi, Hela Baccouche, Sonia Mahjoub, Aya Chakroun","doi":"10.62438/tunismed.v102i6.4834","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Platelet count is crucial for clinical decision. In cases of microcytosis, platelet count based on impedance technique (PLT-I) may overestimate platelet count.</p><p><strong>Aim: </strong>To compare PLT-I with platelet count using the optical technique (PLT-O) and establish a Mean Corpuscular Volume (MCV) threshold for considering PLT-O.</p><p><strong>Methods: </strong>A prospective analytical study conducted over two months involved blood samples collected in standard K2 EDTA tubes for complete blood count analysis, revealing microcytosis (MCV<80 fL). PLT-O analysis in channel-Ret mode was performed using the Sysmex-XN1000 (Sysmex Corporation, Kobe, Japan). Percentage of fragmented red cells (FRC%) and percentage of microcytic red cells (Micro-R%) were recorded. Blood smears stained with May-Grünwald-Giemsa were examined for potential interfering particles.</p><p><strong>Results: </strong>A strong correlation was observed between the two techniques for all platelet values as well as for PLT <150 x 109/L (correlation coefficient r = 0.971, 95% CI: [0.956-0.982]; P<10-3 and r = 0.90, 95% CI: [0.79-0.95]; P< 10-3). The Bland-Altman plot revealed a bias of 16.53 x 109/L between the two methods, with agreement limits between -55.8 and 88.8 x 109/L. A threshold MCV value indicating the use of the optical method, with a cut-off at 72.9fL, demonstrated promising performance consistent with literature findings. However, less favorable performance was observed with Micro-R%.</p><p><strong>Conclusion: </strong>Impedance could be employed in routine practice. However, for MCV<72.9 fL or in the presence of schizocytes, the hemogram validation procedure may incorporate the use of PLT-O.</p>","PeriodicalId":38818,"journal":{"name":"Tunisie Medicale","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-06-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11358783/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Tunisie Medicale","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.62438/tunismed.v102i6.4834","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Platelet count is crucial for clinical decision. In cases of microcytosis, platelet count based on impedance technique (PLT-I) may overestimate platelet count.

Aim: To compare PLT-I with platelet count using the optical technique (PLT-O) and establish a Mean Corpuscular Volume (MCV) threshold for considering PLT-O.

Methods: A prospective analytical study conducted over two months involved blood samples collected in standard K2 EDTA tubes for complete blood count analysis, revealing microcytosis (MCV<80 fL). PLT-O analysis in channel-Ret mode was performed using the Sysmex-XN1000 (Sysmex Corporation, Kobe, Japan). Percentage of fragmented red cells (FRC%) and percentage of microcytic red cells (Micro-R%) were recorded. Blood smears stained with May-Grünwald-Giemsa were examined for potential interfering particles.

Results: A strong correlation was observed between the two techniques for all platelet values as well as for PLT <150 x 109/L (correlation coefficient r = 0.971, 95% CI: [0.956-0.982]; P<10-3 and r = 0.90, 95% CI: [0.79-0.95]; P< 10-3). The Bland-Altman plot revealed a bias of 16.53 x 109/L between the two methods, with agreement limits between -55.8 and 88.8 x 109/L. A threshold MCV value indicating the use of the optical method, with a cut-off at 72.9fL, demonstrated promising performance consistent with literature findings. However, less favorable performance was observed with Micro-R%.

Conclusion: Impedance could be employed in routine practice. However, for MCV<72.9 fL or in the presence of schizocytes, the hemogram validation procedure may incorporate the use of PLT-O.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
小细胞性贫血时通过阻抗和光学技术进行的血小板计数之间的一致性:走向平均血球容积的临界值。
导言血小板计数对临床决策至关重要。目的:比较血小板计数(PLT-I)和光学技术血小板计数(PLT-O),并确定考虑 PLT-O 的平均体细胞容积(MCV)阈值:一项为期两个月的前瞻性分析研究使用标准 K2 EDTA 管采集血样进行全血细胞计数分析,结果显示存在小红细胞症(MCVResults:两种技术在所有血小板值和 PLT 值上都有很强的相关性:阻抗法可用于常规操作。然而,对于 MCV
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Tunisie Medicale
Tunisie Medicale Medicine-Medicine (all)
CiteScore
1.00
自引率
0.00%
发文量
72
期刊最新文献
GAIA therapeutic farm: Prospective study on multidisciplinary care efficacy. Iatrogenic risk in elderly patients. Impact of virtual reality on the perception of procedural pain in paediatric oncology. Incidence and risk factors of venous thromboembolism in patients with amyotrophic lateral sclerosis: a systematic review and meta-analysis. Intravenous Leiomyomatosis of the Uterus: An Intriguing Case Revealed through Anatomopathological Examination.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1