Kun Wang , Xinwei Zang , Wenjie Zhang , Xiangyu Cao , Huiru Zhao , Chunyan Li , Cuiying Liang , Jun Wu
{"title":"Unified calibration of D-dimer can improve the uniformity of different detection systems","authors":"Kun Wang , Xinwei Zang , Wenjie Zhang , Xiangyu Cao , Huiru Zhao , Chunyan Li , Cuiying Liang , Jun Wu","doi":"10.1016/j.plabm.2024.e00413","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>D-dimer at a low level is important evidence for excluding the onset and progression of thrombosis. It is readily detectable and yields rapid results, although significant variability exists among different detection systems. Our study aims to enhance the consistency across various detection systems.</p></div><div><h3>Methods</h3><p>Twelve detection systems were included in our study. We sought to address this inconsistency by using various calibrators (two supplied by manufacturers and two comprising pooled human plasma diluted with different diluents) to standardize D-dimer measurements. We categorized the data into three groups according to D-dimer concentration levels: low (≤0.5 mg/L), medium (>0.5 mg/L - <3 mg/L), and high (≥3 mg/L). We then analyzed the data focusing on range, consistency, comparability, negative coincidence rate, and false negative rate.</p></div><div><h3>Results</h3><p>Calibrating with pooled human plasma led to narrower result ranges in the low and medium groups (P < 0.05). In the low group, consistency improved from weak to strong (ICC 0.4–0.7, P﹤0.05), while it remained excellent in the other groups and overall (ICC﹥0.75, P﹤0.05). The percentage of pairwise comparability increased in both the low and high groups. Additionally, there was an increase in the negative coincidence rate.</p></div><div><h3>Conclusion</h3><p>These findings demonstrate that uniform calibration of D-dimer can significantly enhance the consistency of results across different detection systems.</p></div>","PeriodicalId":20421,"journal":{"name":"Practical Laboratory Medicine","volume":null,"pages":null},"PeriodicalIF":1.7000,"publicationDate":"2024-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2352551724000593/pdfft?md5=df6911df6a73af616b3526357832c179&pid=1-s2.0-S2352551724000593-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Practical Laboratory Medicine","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2352551724000593","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background
D-dimer at a low level is important evidence for excluding the onset and progression of thrombosis. It is readily detectable and yields rapid results, although significant variability exists among different detection systems. Our study aims to enhance the consistency across various detection systems.
Methods
Twelve detection systems were included in our study. We sought to address this inconsistency by using various calibrators (two supplied by manufacturers and two comprising pooled human plasma diluted with different diluents) to standardize D-dimer measurements. We categorized the data into three groups according to D-dimer concentration levels: low (≤0.5 mg/L), medium (>0.5 mg/L - <3 mg/L), and high (≥3 mg/L). We then analyzed the data focusing on range, consistency, comparability, negative coincidence rate, and false negative rate.
Results
Calibrating with pooled human plasma led to narrower result ranges in the low and medium groups (P < 0.05). In the low group, consistency improved from weak to strong (ICC 0.4–0.7, P﹤0.05), while it remained excellent in the other groups and overall (ICC﹥0.75, P﹤0.05). The percentage of pairwise comparability increased in both the low and high groups. Additionally, there was an increase in the negative coincidence rate.
Conclusion
These findings demonstrate that uniform calibration of D-dimer can significantly enhance the consistency of results across different detection systems.
期刊介绍:
Practical Laboratory Medicine is a high-quality, peer-reviewed, international open-access journal publishing original research, new methods and critical evaluations, case reports and short papers in the fields of clinical chemistry and laboratory medicine. The objective of the journal is to provide practical information of immediate relevance to workers in clinical laboratories. The primary scope of the journal covers clinical chemistry, hematology, molecular biology and genetics relevant to laboratory medicine, microbiology, immunology, therapeutic drug monitoring and toxicology, laboratory management and informatics. We welcome papers which describe critical evaluations of biomarkers and their role in the diagnosis and treatment of clinically significant disease, validation of commercial and in-house IVD methods, method comparisons, interference reports, the development of new reagents and reference materials, reference range studies and regulatory compliance reports. Manuscripts describing the development of new methods applicable to laboratory medicine (including point-of-care testing) are particularly encouraged, even if preliminary or small scale.