Photon-Counting Versus Dual-Source CT for Transcatheter Aortic Valve Implantation Planning

IF 3.8 2区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Academic Radiology Pub Date : 2024-12-01 DOI:10.1016/j.acra.2024.06.014
Timm Dirrichs MD , Jörg Schröder MD , Michael Frick MD , Marc Huppertz MD , Roman Iwa , Thomas Allmendinger , Ines Mecking , Christiane K. Kuhl MD
{"title":"Photon-Counting Versus Dual-Source CT for Transcatheter Aortic Valve Implantation Planning","authors":"Timm Dirrichs MD ,&nbsp;Jörg Schröder MD ,&nbsp;Michael Frick MD ,&nbsp;Marc Huppertz MD ,&nbsp;Roman Iwa ,&nbsp;Thomas Allmendinger ,&nbsp;Ines Mecking ,&nbsp;Christiane K. Kuhl MD","doi":"10.1016/j.acra.2024.06.014","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Cardiovascular CT is required for planning transcatheter aortic valve implantation (TAVI).</div></div><div><h3>Purpose</h3><div>To compare image quality, suitability for TAVI planning, and radiation dose of photon-counting CT (PCCT) with that of dual-source CT (DSCT).</div></div><div><h3>Material and Methods</h3><div>Retrospective study on consecutive TAVI candidates with aortic valve stenosis who underwent contrast-enhanced aorto-ilio-femoral PCCT and/or DSCT between 01/2022 and 07/2023. Signal-to-noise (SNR) and contrast-to-noise ratio (CNR) were calculated by standardized ROI analysis. Image quality and suitability for TAVI planning were assessed by four independent expert readers (two cardiac radiologists, two cardiologists) on a 5-point-scale. CT dose index (CTDI) and dose-length-product (DLP) were used to calculate effective radiation dose (eRD).</div></div><div><h3>Results</h3><div>300 patients (136 female, median age: 81 years, IQR: 76–84) underwent 302 CT examinations, with PCCT in 202, DSCT in 100; two patients underwent both. Although SNR and CNR were significantly lower in PCCT vs. DSCT images (33.0 ± 10.5 vs. 47.3 ± 16.4 and 47.3 ± 14.8 vs. 59.3 ± 21.9, P &lt; .001, respectively), visual image quality was higher in PCCT vs. DSCT (4.8 vs. 3.3, P &lt; .001), with moderate overall interreader agreement among radiologists and among cardiologists (κ = 0.60, respectively). Image quality was rated as “excellent” in 160/202 (79.2%) of PCCT vs. 5/100 (5%) of DSCT cases. Readers found images suitable to depict the aortic valve hinge points and to map the femoral access path in 99% of PCCT vs. 85% of DSCT (P &lt; 0.01), with suitability ranked significantly higher in PCCT vs. DSCT (4.8 vs. 3.3, P &lt; .001). Mean CTDI and DLP, and thus eRD, were significantly lower for PCCT vs. DSCT (22.4 vs. 62.9; 519.4 vs. 895.5, and 8.8 ± 4.5 mSv vs. 15.3 ± 5.8 mSv; all P &lt; .001).</div></div><div><h3>Conclusion</h3><div>PCCT improves image quality, effectively avoids non-diagnostic CT imaging for TAVI planning, and is associated with a lower radiation dose compared to state-of-the-art DSCT. Radiologists and cardiologists found PCCT images more suitable for TAVI planning.</div></div>","PeriodicalId":50928,"journal":{"name":"Academic Radiology","volume":"31 12","pages":"Pages 4780-4789"},"PeriodicalIF":3.8000,"publicationDate":"2024-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Academic Radiology","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1076633224003726","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Cardiovascular CT is required for planning transcatheter aortic valve implantation (TAVI).

Purpose

To compare image quality, suitability for TAVI planning, and radiation dose of photon-counting CT (PCCT) with that of dual-source CT (DSCT).

Material and Methods

Retrospective study on consecutive TAVI candidates with aortic valve stenosis who underwent contrast-enhanced aorto-ilio-femoral PCCT and/or DSCT between 01/2022 and 07/2023. Signal-to-noise (SNR) and contrast-to-noise ratio (CNR) were calculated by standardized ROI analysis. Image quality and suitability for TAVI planning were assessed by four independent expert readers (two cardiac radiologists, two cardiologists) on a 5-point-scale. CT dose index (CTDI) and dose-length-product (DLP) were used to calculate effective radiation dose (eRD).

Results

300 patients (136 female, median age: 81 years, IQR: 76–84) underwent 302 CT examinations, with PCCT in 202, DSCT in 100; two patients underwent both. Although SNR and CNR were significantly lower in PCCT vs. DSCT images (33.0 ± 10.5 vs. 47.3 ± 16.4 and 47.3 ± 14.8 vs. 59.3 ± 21.9, P < .001, respectively), visual image quality was higher in PCCT vs. DSCT (4.8 vs. 3.3, P < .001), with moderate overall interreader agreement among radiologists and among cardiologists (κ = 0.60, respectively). Image quality was rated as “excellent” in 160/202 (79.2%) of PCCT vs. 5/100 (5%) of DSCT cases. Readers found images suitable to depict the aortic valve hinge points and to map the femoral access path in 99% of PCCT vs. 85% of DSCT (P < 0.01), with suitability ranked significantly higher in PCCT vs. DSCT (4.8 vs. 3.3, P < .001). Mean CTDI and DLP, and thus eRD, were significantly lower for PCCT vs. DSCT (22.4 vs. 62.9; 519.4 vs. 895.5, and 8.8 ± 4.5 mSv vs. 15.3 ± 5.8 mSv; all P < .001).

Conclusion

PCCT improves image quality, effectively avoids non-diagnostic CT imaging for TAVI planning, and is associated with a lower radiation dose compared to state-of-the-art DSCT. Radiologists and cardiologists found PCCT images more suitable for TAVI planning.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
光子计数与双源 CT 在经导管主动脉瓣植入术规划中的对比。
背景:规划经导管主动脉瓣植入术(TAVI)时需要使用心血管 CT:目的:比较光子计数CT(PCCT)与双源CT(DSCT)的图像质量、TAVI计划的适用性和辐射剂量:对 2022 年 1 月至 2023 年 7 月期间接受造影剂增强主动脉-髂-股动脉 PCCT 和/或 DSCT 的主动脉瓣狭窄 TAVI 候选者进行回顾性研究。通过标准化 ROI 分析计算信噪比 (SNR) 和对比度-噪声比 (CNR)。图像质量和 TAVI 计划的适用性由四位独立专家读者(两位心脏放射专家和两位心脏病专家)按 5 分制进行评估。CT剂量指数(CTDI)和剂量-长度-乘积(DLP)用于计算有效辐射剂量(eRD):300 名患者(136 名女性,中位年龄:81 岁,IQR:76-84)接受了 302 次 CT 检查,其中 202 人接受了 PCCT 检查,100 人接受了 DSCT 检查;两名患者同时接受了这两种检查。虽然 PCCT 图像的 SNR 和 CNR 明显低于 DSCT 图像(33.0 ± 10.5 vs. 47.3 ± 16.4 和 47.3 ± 14.8 vs. 59.3 ± 21.9,P 结论:PCCT 提高了图像质量,但也降低了 CNR:与最先进的 DSCT 相比,PCCT 提高了图像质量,有效避免了 TAVI 计划中的非诊断 CT 成像,且辐射剂量更低。放射科医生和心脏病专家认为 PCCT 图像更适合用于 TAVI 计划。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Academic Radiology
Academic Radiology 医学-核医学
CiteScore
7.60
自引率
10.40%
发文量
432
审稿时长
18 days
期刊介绍: Academic Radiology publishes original reports of clinical and laboratory investigations in diagnostic imaging, the diagnostic use of radioactive isotopes, computed tomography, positron emission tomography, magnetic resonance imaging, ultrasound, digital subtraction angiography, image-guided interventions and related techniques. It also includes brief technical reports describing original observations, techniques, and instrumental developments; state-of-the-art reports on clinical issues, new technology and other topics of current medical importance; meta-analyses; scientific studies and opinions on radiologic education; and letters to the Editor.
期刊最新文献
Improving the Understanding of Iron and Myelin Changes in Patients with Multiple Sclerosis through χ-Separation Imaging. The Feasibility of Measuring Central Blood Pressure with Bedside Echocardiography: An Observational Cross-Sectional Study. Diagnostic Characteristics and Clinical Relevance of Incidental Hypermetabolic Breast Lesions Detected on 18F-FDG PET-CT: A Retrospective Evaluation. Diagnostic Efficacy of Ventilation-Perfusion Single Photo Emission Computed Tomography/Computed Tomography for Pulmonary Hypertension due to Fibrinous Mediastinitis. Predicting Intracerebral Hemorrhage Outcomes Using Deep Learning Models to Extract Head CT Imaging Features.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1