{"title":"Performance of the quadripolar CRT-D system: Five-year results from the Quadripolar Pacing Post-Approval Study","authors":"Raffaele Corbisiero MD , Michael Cao MD , David Muller , Kwangdeok Lee PhD , David Martin MD , QUAD-PAS investigators","doi":"10.1016/j.hrthm.2024.07.006","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Cardiac resynchronization therapy (CRT) is associated with challenges such as elevated capture thresholds, diaphragmatic stimulation, and lead instability.</div></div><div><h3>Objective</h3><div>This study aimed to assess the long-term safety and efficacy of the quadripolar CRT-defibrillator (CRT-D) device system with the Quartet 1458Q left ventricular (LV) lead in a CRT-indicated population observed for 5 years and to evaluate all-cause mortality and impact of baseline characteristics on survival through 5 years.</div></div><div><h3>Methods</h3><div>Patients indicated for a CRT-D system were observed every 6 months after implantation for 5 years, and device performance and adverse events were assessed at each visit. The 3 primary end points were freedom from quadripolar CRT-D system-related complications through 5 years, freedom from Quartet 1458Q LV lead-related complications through 5 years, and mean programmed pacing capture threshold at 5 years.</div></div><div><h3>Results</h3><div>The study enrolled 1970 participants at 71 sites. The quadripolar CRT-D system was successfully implanted in 97.2% of participants. Freedom from quadripolar CRT-D device system-related complications through 5 years was 89.7%. Freedom from Quartet 1458Q LV lead-related complications through 5 years was 95.7%; 3.49% of participants had LV lead-related complications, and an overall LV lead complication rate was 0.0122 event per patient-year. The mean LV pacing capture threshold was 1.52 ± 1.01 V at 5 years. The 5-year survival rate was 67.4%.</div></div><div><h3>Conclusion</h3><div>The quadripolar CRT-D system with the Quartet 1458Q LV lead exhibited low rates of complications and stable electrical performance through 5 years of follow-up and suggested a higher 5-year survival rate compared with traditional CRT systems.</div></div>","PeriodicalId":12886,"journal":{"name":"Heart rhythm","volume":"22 2","pages":"Pages 349-356"},"PeriodicalIF":5.6000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Heart rhythm","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1547527124028777","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Cardiac resynchronization therapy (CRT) is associated with challenges such as elevated capture thresholds, diaphragmatic stimulation, and lead instability.
Objective
This study aimed to assess the long-term safety and efficacy of the quadripolar CRT-defibrillator (CRT-D) device system with the Quartet 1458Q left ventricular (LV) lead in a CRT-indicated population observed for 5 years and to evaluate all-cause mortality and impact of baseline characteristics on survival through 5 years.
Methods
Patients indicated for a CRT-D system were observed every 6 months after implantation for 5 years, and device performance and adverse events were assessed at each visit. The 3 primary end points were freedom from quadripolar CRT-D system-related complications through 5 years, freedom from Quartet 1458Q LV lead-related complications through 5 years, and mean programmed pacing capture threshold at 5 years.
Results
The study enrolled 1970 participants at 71 sites. The quadripolar CRT-D system was successfully implanted in 97.2% of participants. Freedom from quadripolar CRT-D device system-related complications through 5 years was 89.7%. Freedom from Quartet 1458Q LV lead-related complications through 5 years was 95.7%; 3.49% of participants had LV lead-related complications, and an overall LV lead complication rate was 0.0122 event per patient-year. The mean LV pacing capture threshold was 1.52 ± 1.01 V at 5 years. The 5-year survival rate was 67.4%.
Conclusion
The quadripolar CRT-D system with the Quartet 1458Q LV lead exhibited low rates of complications and stable electrical performance through 5 years of follow-up and suggested a higher 5-year survival rate compared with traditional CRT systems.
期刊介绍:
HeartRhythm, the official Journal of the Heart Rhythm Society and the Cardiac Electrophysiology Society, is a unique journal for fundamental discovery and clinical applicability.
HeartRhythm integrates the entire cardiac electrophysiology (EP) community from basic and clinical academic researchers, private practitioners, engineers, allied professionals, industry, and trainees, all of whom are vital and interdependent members of our EP community.
The Heart Rhythm Society is the international leader in science, education, and advocacy for cardiac arrhythmia professionals and patients, and the primary information resource on heart rhythm disorders. Its mission is to improve the care of patients by promoting research, education, and optimal health care policies and standards.