Performance of the quadripolar CRT-D system: Five-year results from the Quadripolar Pacing Post-Approval Study

IF 5.6 2区 医学 Q1 CARDIAC & CARDIOVASCULAR SYSTEMS Heart rhythm Pub Date : 2025-02-01 DOI:10.1016/j.hrthm.2024.07.006
Raffaele Corbisiero MD , Michael Cao MD , David Muller , Kwangdeok Lee PhD , David Martin MD , QUAD-PAS investigators
{"title":"Performance of the quadripolar CRT-D system: Five-year results from the Quadripolar Pacing Post-Approval Study","authors":"Raffaele Corbisiero MD ,&nbsp;Michael Cao MD ,&nbsp;David Muller ,&nbsp;Kwangdeok Lee PhD ,&nbsp;David Martin MD ,&nbsp;QUAD-PAS investigators","doi":"10.1016/j.hrthm.2024.07.006","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Cardiac resynchronization therapy (CRT) is associated with challenges such as elevated capture thresholds, diaphragmatic stimulation, and lead instability.</div></div><div><h3>Objective</h3><div>This study aimed to assess the long-term safety and efficacy of the quadripolar CRT-defibrillator (CRT-D) device system with the Quartet 1458Q left ventricular (LV) lead in a CRT-indicated population observed for 5 years and to evaluate all-cause mortality and impact of baseline characteristics on survival through 5 years.</div></div><div><h3>Methods</h3><div>Patients indicated for a CRT-D system were observed every 6 months after implantation for 5 years, and device performance and adverse events were assessed at each visit. The 3 primary end points were freedom from quadripolar CRT-D system-related complications through 5 years, freedom from Quartet 1458Q LV lead-related complications through 5 years, and mean programmed pacing capture threshold at 5 years.</div></div><div><h3>Results</h3><div>The study enrolled 1970 participants at 71 sites. The quadripolar CRT-D system was successfully implanted in 97.2% of participants. Freedom from quadripolar CRT-D device system-related complications through 5 years was 89.7%. Freedom from Quartet 1458Q LV lead-related complications through 5 years was 95.7%; 3.49% of participants had LV lead-related complications, and an overall LV lead complication rate was 0.0122 event per patient-year. The mean LV pacing capture threshold was 1.52 ± 1.01 V at 5 years. The 5-year survival rate was 67.4%.</div></div><div><h3>Conclusion</h3><div>The quadripolar CRT-D system with the Quartet 1458Q LV lead exhibited low rates of complications and stable electrical performance through 5 years of follow-up and suggested a higher 5-year survival rate compared with traditional CRT systems.</div></div>","PeriodicalId":12886,"journal":{"name":"Heart rhythm","volume":"22 2","pages":"Pages 349-356"},"PeriodicalIF":5.6000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Heart rhythm","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1547527124028777","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Cardiac resynchronization therapy (CRT) is associated with challenges such as elevated capture thresholds, diaphragmatic stimulation, and lead instability.

Objective

This study aimed to assess the long-term safety and efficacy of the quadripolar CRT-defibrillator (CRT-D) device system with the Quartet 1458Q left ventricular (LV) lead in a CRT-indicated population observed for 5 years and to evaluate all-cause mortality and impact of baseline characteristics on survival through 5 years.

Methods

Patients indicated for a CRT-D system were observed every 6 months after implantation for 5 years, and device performance and adverse events were assessed at each visit. The 3 primary end points were freedom from quadripolar CRT-D system-related complications through 5 years, freedom from Quartet 1458Q LV lead-related complications through 5 years, and mean programmed pacing capture threshold at 5 years.

Results

The study enrolled 1970 participants at 71 sites. The quadripolar CRT-D system was successfully implanted in 97.2% of participants. Freedom from quadripolar CRT-D device system-related complications through 5 years was 89.7%. Freedom from Quartet 1458Q LV lead-related complications through 5 years was 95.7%; 3.49% of participants had LV lead-related complications, and an overall LV lead complication rate was 0.0122 event per patient-year. The mean LV pacing capture threshold was 1.52 ± 1.01 V at 5 years. The 5-year survival rate was 67.4%.

Conclusion

The quadripolar CRT-D system with the Quartet 1458Q LV lead exhibited low rates of complications and stable electrical performance through 5 years of follow-up and suggested a higher 5-year survival rate compared with traditional CRT systems.
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
四极 CRT-D 系统的性能:四极起搏批准后研究的 5 年结果。
背景:心脏再同步化疗法(CRT心脏再同步化疗法(CRT)面临着捕获阈值升高、膈肌刺激和导联不稳定等挑战:在 CRT 适应症人群中随访 5 年,评估带有 Quartet 1458Q 左室(LV)导联的四极 CRT-D 装置系统的长期安全性和有效性,并评估全因死亡率以及基线特征对 5 年生存期的影响:植入 CRT-D 系统后,每 6 个月对患者进行一次为期 5 年的随访,并在每次随访时评估设备性能和不良事件。三个主要终点是:5 年内无四极 CRT-D 系统相关并发症、5 年内无 Quartet 1458Q LV 导联相关并发症以及 5 年内平均编程起搏捕获阈值:该研究在 71 个地点招募了 1,970 名受试者。97.2%的受试者成功植入了四极 CRT-D 系统。5年内,89.7%的受试者未出现四极CRT-D设备系统相关并发症,95.7%的受试者未出现Quartet 1458Q左心室导联相关并发症。3.49%的受试者出现左心室导联相关并发症,总体左心室导联并发症发生率为每受试者年 0.0122 例。5 年时,左心室起搏捕获阈值的平均值为 1.52 ± 1.01 V。5年存活率为67.4%:结论:与传统 CRT 系统相比,使用 Quartet 1458Q LV 导联的四极 CRT-D 系统在 5 年随访期间并发症发生率低,电气性能稳定,5 年存活率更高。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Heart rhythm
Heart rhythm 医学-心血管系统
CiteScore
10.50
自引率
5.50%
发文量
1465
审稿时长
24 days
期刊介绍: HeartRhythm, the official Journal of the Heart Rhythm Society and the Cardiac Electrophysiology Society, is a unique journal for fundamental discovery and clinical applicability. HeartRhythm integrates the entire cardiac electrophysiology (EP) community from basic and clinical academic researchers, private practitioners, engineers, allied professionals, industry, and trainees, all of whom are vital and interdependent members of our EP community. The Heart Rhythm Society is the international leader in science, education, and advocacy for cardiac arrhythmia professionals and patients, and the primary information resource on heart rhythm disorders. Its mission is to improve the care of patients by promoting research, education, and optimal health care policies and standards.
期刊最新文献
Decreased Left Ventricular Systolic Function during the Late Phase after Response to Cardiac Resynchronization Therapy. Of course it's about scar. Operator Learning Curve with a Novel Dual-Energy Lattice-tip Ablation System. Predicting Pacemaker Implantation After TAVI In Patients With Right Bundle Branch Block: One Must Imagine Sisyphus Happy. Sex Differences in Atrial Remodeling and Atrial Fibrillation Recurrence After Catheter Ablation.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1