Repeatability and agreement of two ocular biometers with single and dual Scheimpflug cameras in keratoconus eyes.

IF 2.4 3区 医学 Q2 OPHTHALMOLOGY Graefe’s Archive for Clinical and Experimental Ophthalmology Pub Date : 2024-11-01 Epub Date: 2024-07-24 DOI:10.1007/s00417-024-06511-z
Christoph Lwowski, Daniel Krüger, Thomas Kohnen
{"title":"Repeatability and agreement of two ocular biometers with single and dual Scheimpflug cameras in keratoconus eyes.","authors":"Christoph Lwowski, Daniel Krüger, Thomas Kohnen","doi":"10.1007/s00417-024-06511-z","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>To evaluate the repeatability and agreement of two different ocular biometers and Scheimpflug devices in keratoconus eyes.</p><p><strong>Methods: </strong>This prospective, comparative trial took place at the University hospital, Goethe University, Frankfurt am Main, Germany. We included eyes with keratoconus, one eye per patient, randomly selected. Measurements were taken with Galilei G6 (Ziemer, Switzerland) and Pentacam AXL (Oculus, Germany), three consecutive measurements each. Repeatability and agreement were evaluated for simulated keratometry (simK), astigmatism (simA), maximum keratometry (KMax) and its axis, total keratometry (TCP), axial length (AL), anterior chamber depth (ACD), and thinnest pachymetry (TCT).</p><p><strong>Results: </strong>Both devices showed an excellent repeatability with intra class correlation (ICC) of > 0.97 for all parameters. The 95% limits of repeatability (LoR95%) and agreement (LoA95%) were narrow for all parameters. The Galilei G6 had a narrower LoAR95% for TCT (2.1 μm vs. 4.6 μm), but a wider LoR95% for KMax (0.52D vs. 1.18D). No relevant difference was found for the other parameters. Agreement between the devices was good to moderate, especially for simK and TCP.</p><p><strong>Conclusions: </strong>Both devices show excellent repeatability with narrow LoR95% and high ICC for all parameters. The only relevant difference was found for KMax and TCT in favor of Pentacam AXL and Galilei G6, respectively. Agreement was good to moderate, and most parameters should not be considered interchangeable.</p>","PeriodicalId":12795,"journal":{"name":"Graefe’s Archive for Clinical and Experimental Ophthalmology","volume":" ","pages":"3701-3709"},"PeriodicalIF":2.4000,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Graefe’s Archive for Clinical and Experimental Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00417-024-06511-z","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/7/24 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background: To evaluate the repeatability and agreement of two different ocular biometers and Scheimpflug devices in keratoconus eyes.

Methods: This prospective, comparative trial took place at the University hospital, Goethe University, Frankfurt am Main, Germany. We included eyes with keratoconus, one eye per patient, randomly selected. Measurements were taken with Galilei G6 (Ziemer, Switzerland) and Pentacam AXL (Oculus, Germany), three consecutive measurements each. Repeatability and agreement were evaluated for simulated keratometry (simK), astigmatism (simA), maximum keratometry (KMax) and its axis, total keratometry (TCP), axial length (AL), anterior chamber depth (ACD), and thinnest pachymetry (TCT).

Results: Both devices showed an excellent repeatability with intra class correlation (ICC) of > 0.97 for all parameters. The 95% limits of repeatability (LoR95%) and agreement (LoA95%) were narrow for all parameters. The Galilei G6 had a narrower LoAR95% for TCT (2.1 μm vs. 4.6 μm), but a wider LoR95% for KMax (0.52D vs. 1.18D). No relevant difference was found for the other parameters. Agreement between the devices was good to moderate, especially for simK and TCP.

Conclusions: Both devices show excellent repeatability with narrow LoR95% and high ICC for all parameters. The only relevant difference was found for KMax and TCT in favor of Pentacam AXL and Galilei G6, respectively. Agreement was good to moderate, and most parameters should not be considered interchangeable.

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
在角膜炎患者眼中使用单双 Scheimpflug 相机的两种眼球生物测量仪的重复性和一致性。
背景:评估两种不同的眼球生物测量仪和 Scheimpflug 设备在角膜炎患者眼中的重复性和一致性:评估两种不同的眼球生物测量仪和 Scheimpflug 设备在角膜炎患者眼中的重复性和一致性:这项前瞻性比较试验在德国美因河畔法兰克福歌德大学附属医院进行。我们随机抽取了每位患者的一只角膜炎眼。使用 Galilei G6(瑞士 Ziemer 公司)和 Pentacam AXL(德国 Oculus 公司)进行测量,每次连续测量三次。对模拟角膜度量(simK)、散光(simA)、最大角膜度量(KMax)及其轴、总角膜度量(TCP)、轴向长度(AL)、前房深度(ACD)和最薄厚度(TCT)的重复性和一致性进行了评估:两种设备都显示出极佳的重复性,所有参数的类内相关性(ICC)均大于 0.97。所有参数的重复性 95% 限值(LoR95%)和一致性 95% 限值(LoA95%)都很窄。Galilei G6 的 TCT LoAR95% 较窄(2.1 μm 对 4.6 μm),但 KMax LoR95% 较宽(0.52D 对 1.18D)。其他参数没有发现相关差异。设备之间的一致性良好至中等,尤其是 simK 和 TCP:结论:两种仪器都显示出极佳的重复性,所有参数的 LoR95% 都很窄,ICC 都很高。唯一的相关差异是 KMax 和 TCT,分别有利于 Pentacam AXL 和 Galilei G6。一致性良好至中等,大多数参数不应被视为可以互换。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
5.40
自引率
7.40%
发文量
398
审稿时长
3 months
期刊介绍: Graefe''s Archive for Clinical and Experimental Ophthalmology is a distinguished international journal that presents original clinical reports and clini-cally relevant experimental studies. Founded in 1854 by Albrecht von Graefe to serve as a source of useful clinical information and a stimulus for discussion, the journal has published articles by leading ophthalmologists and vision research scientists for more than a century. With peer review by an international Editorial Board and prompt English-language publication, Graefe''s Archive provides rapid dissemination of clinical and clinically related experimental information.
期刊最新文献
Evaluating anterior segment stability and corneal endothelium after prolene gonioscopy assisted transluminal trabeculotomy (GATT) in open-angle glaucoma. Effects of ambient atmospheric pressure on intraocular pressure measured using a Goldman applanation tonometer in normal eyes under ordinary conditions. Enhancing diabetic retinopathy and macular edema detection through multi scale feature fusion using deep learning model. Primary angle-closed diseases recognition through artificial intelligence-based anterior segment-optical coherence tomography imaging. Anterior segment OCT for imaging PAUL® glaucoma implant patch grafts: a useful method for follow-up and risk management.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1