Analytical and Clinical Validation of an Automated Microfluidic Interleukin-6 Immunoassay.

IF 1.1 4区 医学 Q4 MEDICAL LABORATORY TECHNOLOGY Annals of clinical and laboratory science Pub Date : 2024-05-01
Jianbo Yang, Mustafa A Barbhuiya, Christopher Hamilton, Kimberly Robyak, Yusheng Zhu
{"title":"Analytical and Clinical Validation of an Automated Microfluidic Interleukin-6 Immunoassay.","authors":"Jianbo Yang, Mustafa A Barbhuiya, Christopher Hamilton, Kimberly Robyak, Yusheng Zhu","doi":"","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>We validated an automated microfluidic interleukin-6 (IL-6) immunoassay on the Ella platform for clinical use.</p><p><strong>Methods: </strong>The assay was validated for precision, lower limit of quantification, analytical measurement range, accuracy, specificity, interference of biotin, tocilizumab, GP130, IL-6R<i>α</i>, hemolysis, icterus, and lipemia, and establishing the reference value. The clinical performance was evaluated in 96 COVID-19 patients.</p><p><strong>Results: </strong>The within-run and between-run coefficient of variations (CV) ranged 1.8%-4.8%. This assay has an analytical measurement range (AMR) 0.7-2652 pg/ml. There was a moderate correlation between Ella IL-6 assay (y) and a Luminex Quantitative Multiplex Bead Assay in a reference lab (x): y=8.561x-475.38, R=0.5047, SEE=1592.8 pg/mL, N=48). Measurement of IL-6 in plasma samples from 45 healthy adults showed the upper limit of reference of 5.0 pg/mL. 95.83% (95% CI: 89.67%-98.85%) of COVID-19 patients had IL-6 >5.0 pg/mL. This assay is resistant to the interference from hemoglobin up to 1,144 mg/dL, triglyceride 1,699 mg/dL, bilirubin 35 mg/dL, biotin 1000 ng/mL, tocilizumab 240 μg/mL, IL-11 50,000 pg/mL, GP130 50,000 pg/mL, and IL-6R<i>α</i> 1,000 pg/mL.</p><p><strong>Conclusions: </strong>The IL-6 assay on the Ella platform is robust with minimum manual operation. The analytical and clinical performance characteristics meet the clinical needs.</p>","PeriodicalId":8228,"journal":{"name":"Annals of clinical and laboratory science","volume":"54 3","pages":"388-393"},"PeriodicalIF":1.1000,"publicationDate":"2024-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of clinical and laboratory science","FirstCategoryId":"3","ListUrlMain":"","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: We validated an automated microfluidic interleukin-6 (IL-6) immunoassay on the Ella platform for clinical use.

Methods: The assay was validated for precision, lower limit of quantification, analytical measurement range, accuracy, specificity, interference of biotin, tocilizumab, GP130, IL-6Rα, hemolysis, icterus, and lipemia, and establishing the reference value. The clinical performance was evaluated in 96 COVID-19 patients.

Results: The within-run and between-run coefficient of variations (CV) ranged 1.8%-4.8%. This assay has an analytical measurement range (AMR) 0.7-2652 pg/ml. There was a moderate correlation between Ella IL-6 assay (y) and a Luminex Quantitative Multiplex Bead Assay in a reference lab (x): y=8.561x-475.38, R=0.5047, SEE=1592.8 pg/mL, N=48). Measurement of IL-6 in plasma samples from 45 healthy adults showed the upper limit of reference of 5.0 pg/mL. 95.83% (95% CI: 89.67%-98.85%) of COVID-19 patients had IL-6 >5.0 pg/mL. This assay is resistant to the interference from hemoglobin up to 1,144 mg/dL, triglyceride 1,699 mg/dL, bilirubin 35 mg/dL, biotin 1000 ng/mL, tocilizumab 240 μg/mL, IL-11 50,000 pg/mL, GP130 50,000 pg/mL, and IL-6Rα 1,000 pg/mL.

Conclusions: The IL-6 assay on the Ella platform is robust with minimum manual operation. The analytical and clinical performance characteristics meet the clinical needs.

分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
自动微流控白细胞介素-6 免疫分析仪的分析和临床验证
目的我们对艾拉平台上用于临床的白细胞介素-6(IL-6)自动微流控免疫测定进行了验证:方法:对该测定的精密度、定量下限、分析测量范围、准确度、特异性、生物素、托珠单抗、GP130、IL-6Rα、溶血、黄疸和脂血的干扰进行了验证,并确定了参考值。对96名COVID-19患者的临床表现进行了评估:结果:运行内和运行间变异系数(CV)为 1.8%-4.8%。该测定的分析测量范围(AMR)为 0.7-2652 pg/ml。艾拉 IL-6 分析法(y)与参考实验室的 Luminex 定量多重微珠分析法(x)之间存在中等程度的相关性:y=8.561x-475.38,R=0.5047,SEE=1592.8 pg/ml,N=48)。对 45 名健康成人血浆样本中 IL-6 的测量结果显示,参考上限为 5.0 pg/mL。95.83%(95% CI:89.67%-98.85%)的 COVID-19 患者 IL-6 >5.0 pg/mL。该测定不受血红蛋白高达 1,144 mg/dL、甘油三酯 1,699 mg/dL、胆红素 35 mg/dL、生物素 1000 ng/mL、妥珠单抗 240 μg/mL、IL-11 50,000 pg/mL、GP130 50,000 pg/mL、IL-6Rα 1,000 pg/mL的干扰:艾拉平台上的IL-6测定功能强大,只需最少的人工操作。结论:艾拉平台上的IL-6测定功能强大,手动操作最少,分析和临床性能特点符合临床需求。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Annals of clinical and laboratory science
Annals of clinical and laboratory science 医学-医学实验技术
CiteScore
1.60
自引率
0.00%
发文量
112
审稿时长
6-12 weeks
期刊介绍: The Annals of Clinical & Laboratory Science welcomes manuscripts that report research in clinical science, including pathology, clinical chemistry, biotechnology, molecular biology, cytogenetics, microbiology, immunology, hematology, transfusion medicine, organ and tissue transplantation, therapeutics, toxicology, and clinical informatics.
期刊最新文献
Communication: 4-Hydroxynonenal and Fracture Risk in the Community-Dwelling Older People. Letter to the Editor. Evaluation of Swab: Direct Extraction Kit, Comparing with AdvanSure E3 System for Nucleic Acid Extraction in Identification of SARS-CoV-2. A Case of Hb Phnom Penh Showing Different HbA1c Levels Depending on the High-Performance Liquid Chromatography System. Application of IdyllaTM System for Rapid Evaluation of EGFR Mutation Status in Formalin-Fixed and Paraffin-Embedded Samples of Non-Small Cell Lung Cancer. Biomarkers Identification of Early Rheumatoid Arthritis via Bioinformatics Approach and Experimental Verification.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1