Methods and validation of velacur determined fat fraction in patients with MASLD

Mohammad Honarvar , Julio Lobo , Caitlin Schneider , Samuel Klein , Gordon I. Smith , Rohit Loomba , Alnoor Ramji , Tarek Hassanein , Eric M. Yoshida , Emily Pang , Michael P. Curry , Nezam H. Afdhal
{"title":"Methods and validation of velacur determined fat fraction in patients with MASLD","authors":"Mohammad Honarvar ,&nbsp;Julio Lobo ,&nbsp;Caitlin Schneider ,&nbsp;Samuel Klein ,&nbsp;Gordon I. Smith ,&nbsp;Rohit Loomba ,&nbsp;Alnoor Ramji ,&nbsp;Tarek Hassanein ,&nbsp;Eric M. Yoshida ,&nbsp;Emily Pang ,&nbsp;Michael P. Curry ,&nbsp;Nezam H. Afdhal","doi":"10.1016/j.wfumbo.2024.100061","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p>As prevalence of patients with steatotic liver diseases increases throughout the world, it is necessary to have accurate and accessible methods to estimate liver fat content. Using quantitative ultrasound parameters, such as attenuation and backscatter, it is possible to estimate liver fat with MRI proton density fat fraction as the reference standard. Velacur determined fat fraction (VDFF) is a new output measurement on Velacur (Sonic Incytes Medical Corp, Vancouver, BC).</p></div><div><h3>Methods</h3><p>This study described the results of parameter fitting and validation of VDFF, which is a combination of quantitative ultrasound parameters. Patients were recruited from sites within the US and Canada. All patients had contemporaneous Velacur and MRI proton density fat fraction scans. The quantitative ultrasound parameter fitting was completed using linear regression on a random sub-sample approach, and a separate cohort was used for validation. The AUC for detection of 5 % liver fat based on MRI-PDFF and the correlation between MRI-PDFF and VDFF was measured in both cohorts.</p></div><div><h3>Results</h3><p>VDFF had an AUROC of 0.97 for the detection of MRI-PDFF &gt;5 % in the parameter fitting cohort, and 0.99 in the validation cohort. The correlation [95 % CI] between MRI-PDFF and VDFF was r = 0.84 [0.78–0.89] for the parameter fitting cohort and r = 0.90 [0.82–0.95] for the validation cohort.</p></div><div><h3>Conclusion</h3><p>The Velacur Determined Fat Fraction (VDFF) is an accurate and accessible way to estimate steatosis as measured by MRI-PDFF. Velacur VDFF can fill the unmet need of an accurate means to diagnosis hepatic steatosis and serve as a potential alternative to biopsy or MRI-PDFF.</p></div>","PeriodicalId":101281,"journal":{"name":"WFUMB Ultrasound Open","volume":"2 2","pages":"Article 100061"},"PeriodicalIF":0.0000,"publicationDate":"2024-07-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2949668324000296/pdfft?md5=b2ba02eeedafea6bdb5468e19fe2dbc6&pid=1-s2.0-S2949668324000296-main.pdf","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"WFUMB Ultrasound Open","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2949668324000296","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction

As prevalence of patients with steatotic liver diseases increases throughout the world, it is necessary to have accurate and accessible methods to estimate liver fat content. Using quantitative ultrasound parameters, such as attenuation and backscatter, it is possible to estimate liver fat with MRI proton density fat fraction as the reference standard. Velacur determined fat fraction (VDFF) is a new output measurement on Velacur (Sonic Incytes Medical Corp, Vancouver, BC).

Methods

This study described the results of parameter fitting and validation of VDFF, which is a combination of quantitative ultrasound parameters. Patients were recruited from sites within the US and Canada. All patients had contemporaneous Velacur and MRI proton density fat fraction scans. The quantitative ultrasound parameter fitting was completed using linear regression on a random sub-sample approach, and a separate cohort was used for validation. The AUC for detection of 5 % liver fat based on MRI-PDFF and the correlation between MRI-PDFF and VDFF was measured in both cohorts.

Results

VDFF had an AUROC of 0.97 for the detection of MRI-PDFF >5 % in the parameter fitting cohort, and 0.99 in the validation cohort. The correlation [95 % CI] between MRI-PDFF and VDFF was r = 0.84 [0.78–0.89] for the parameter fitting cohort and r = 0.90 [0.82–0.95] for the validation cohort.

Conclusion

The Velacur Determined Fat Fraction (VDFF) is an accurate and accessible way to estimate steatosis as measured by MRI-PDFF. Velacur VDFF can fill the unmet need of an accurate means to diagnosis hepatic steatosis and serve as a potential alternative to biopsy or MRI-PDFF.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Velacur 测定 MASLD 患者脂肪分数的方法与验证
引言 随着脂肪性肝病患者的发病率在全球范围内不断上升,有必要采用准确、简便的方法来估算肝脏脂肪含量。利用定量超声参数(如衰减和反向散射),以核磁共振质子密度脂肪分数作为参考标准,可以估算肝脏脂肪含量。Velacur确定的脂肪分数(VDFF)是Velacur(Sonic Incytes Medical Corp,不列颠哥伦比亚省温哥华市)的一种新的输出测量方法。这项研究描述了VDFF的参数拟合和验证结果,VDFF是定量超声参数的组合。患者是从美国和加拿大的医疗机构招募的。所有患者都接受了同期的 Velacur 和 MRI 质子密度脂肪分数扫描。定量超声参数拟合采用随机子样本线性回归法完成,并使用单独的队列进行验证。结果 在参数拟合队列中,VDFF 检测 MRI-PDFF >5 % 的 AUROC 为 0.97,在验证队列中为 0.99。在参数拟合队列中,MRI-PDFF 和 VDFF 之间的相关性[95 % CI]为 r = 0.84 [0.78-0.89],在验证队列中为 r = 0.90 [0.82-0.95]。Velacur VDFF 可以满足人们对准确诊断肝脂肪变性的需求,是活检或 MRI-PDFF 的潜在替代方法。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Added clinical advantage of combining ultrasound with radiograph in assessing ankle injuries: Comparison with MRI Treatment of inoperable pancreatic adenocarcinoma with focused ultrasound and microbubbles in patients receiving chemotherapy Breast cancer ultrasound image segmentation using improved 3DUnet++ Utilization of evoked vibrational signatures under ultrasound examination as a novel method of tissue classification Protocol refinement and inter- and intra-rater reliability assessment of ultrasound-based measurements of hamstring architecture, and echo intensity, and intra-rater reliability of shear wave elastography
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1