Axial rod slip at the end-of-construct screw in scoliosis surgery: relevance, occurrence and prevention.

IF 1.6 Q3 CLINICAL NEUROLOGY Spine deformity Pub Date : 2024-11-01 Epub Date: 2024-08-20 DOI:10.1007/s43390-024-00925-9
T P Schlösser, I Blaauw, M R van der Valk, Guido van Solinge, C Faber, M C Kruyt
{"title":"Axial rod slip at the end-of-construct screw in scoliosis surgery: relevance, occurrence and prevention.","authors":"T P Schlösser, I Blaauw, M R van der Valk, Guido van Solinge, C Faber, M C Kruyt","doi":"10.1007/s43390-024-00925-9","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Despite standardized biomechanical tests for spinal implants, we recently recognized pedicle screw failure to maintain the rod fixated as a clinical concern in scoliosis surgery. This occurrence study investigates the risk and magnitude of axial rod slip (ARS), its relation with technique and preventive measures.</p><p><strong>Methods: </strong>Retrospective multicenter review of all primary scoliosis cases (2018-2020) with > 1 year FU from three centers, instrumented with uniplanar screws and 5.5 mm CoCr rods (Mesa 2, Stryker Corporation, Kalamazoo, MI, USA). ARS was defined as > 1 mm change in residual distal rod length from the screw in the lowest instrumented vertebra (LIV) and assessed by two independent observers. Slip distance, direction, relation to distal screw density and time of observation were recorded, as well as the effect of ARS on caudal curve increase. To prevent slip, more recent patients were instrumented with a different end-of-construct screw (Reline, NuVasive Inc. San Diego, CA, USA) and analyzed for comparison.</p><p><strong>Results: </strong>ARS risk was 27% (56/205) with a distance of 3.6 ± 2.2 mm, predominantly convex. 42% occurred before 4 months, the rest before 1 year. The caudal curve substantially increased three times more often in patients with ARS. Interobserver reliability was high and slip was in the expected direction. ARS was unrelated to distal screw density. Remarkable variation in ARS rates (53%, 31%, 13%) existed between the centers, while there was no difference in mean screw density (≈1.3 screws/level) or curve correction (≈60%). Revision surgery for ARS was required in 2.9% (6/207). Using the different end-of-construct screw, ARS risk was only 2% (1/56) and no revisions were required.</p><p><strong>Conclusion: </strong>This study demonstrates the prevalence of axial rod slip at the end of construct in scoliosis surgery and its clinical relevance. While minimal ARS can be subclinical, ARS should not be mistaken for adding on. The most severe ARS predominantly occurred convex at the high-loaded distal screw when L3 was the LIV. Longer constructs (LIV L3 or L4) have a higher risk of ARS. The minimal risk of ARS with another end-of-construct screw underscores the influence of screw type on ARS occurrence in our series. Further research is essential to refine techniques and enhance patient outcomes.</p>","PeriodicalId":21796,"journal":{"name":"Spine deformity","volume":" ","pages":"1699-1707"},"PeriodicalIF":1.6000,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11499526/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Spine deformity","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s43390-024-00925-9","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/8/20 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: Despite standardized biomechanical tests for spinal implants, we recently recognized pedicle screw failure to maintain the rod fixated as a clinical concern in scoliosis surgery. This occurrence study investigates the risk and magnitude of axial rod slip (ARS), its relation with technique and preventive measures.

Methods: Retrospective multicenter review of all primary scoliosis cases (2018-2020) with > 1 year FU from three centers, instrumented with uniplanar screws and 5.5 mm CoCr rods (Mesa 2, Stryker Corporation, Kalamazoo, MI, USA). ARS was defined as > 1 mm change in residual distal rod length from the screw in the lowest instrumented vertebra (LIV) and assessed by two independent observers. Slip distance, direction, relation to distal screw density and time of observation were recorded, as well as the effect of ARS on caudal curve increase. To prevent slip, more recent patients were instrumented with a different end-of-construct screw (Reline, NuVasive Inc. San Diego, CA, USA) and analyzed for comparison.

Results: ARS risk was 27% (56/205) with a distance of 3.6 ± 2.2 mm, predominantly convex. 42% occurred before 4 months, the rest before 1 year. The caudal curve substantially increased three times more often in patients with ARS. Interobserver reliability was high and slip was in the expected direction. ARS was unrelated to distal screw density. Remarkable variation in ARS rates (53%, 31%, 13%) existed between the centers, while there was no difference in mean screw density (≈1.3 screws/level) or curve correction (≈60%). Revision surgery for ARS was required in 2.9% (6/207). Using the different end-of-construct screw, ARS risk was only 2% (1/56) and no revisions were required.

Conclusion: This study demonstrates the prevalence of axial rod slip at the end of construct in scoliosis surgery and its clinical relevance. While minimal ARS can be subclinical, ARS should not be mistaken for adding on. The most severe ARS predominantly occurred convex at the high-loaded distal screw when L3 was the LIV. Longer constructs (LIV L3 or L4) have a higher risk of ARS. The minimal risk of ARS with another end-of-construct screw underscores the influence of screw type on ARS occurrence in our series. Further research is essential to refine techniques and enhance patient outcomes.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
脊柱侧弯手术中结构末端螺钉处的轴向杆滑脱:相关性、发生率和预防。
目的:尽管对脊柱植入物进行了标准化的生物力学测试,但我们最近发现椎弓根螺钉无法保持杆固定是脊柱侧弯手术中的一个临床问题。本研究调查了轴向杆滑脱(ARS)的风险和程度、其与技术和预防措施的关系:回顾性多中心审查三个中心所有FU>1年的原发性脊柱侧凸病例(2018-2020年),使用单平面螺钉和5.5 mm CoCr杆(Mesa 2,史赛克公司,美国密歇根州卡拉马祖)进行器械固定。ARS的定义是:从最低器械椎体(LIV)的螺钉算起,残余远端杆长度变化> 1 mm,并由两名独立观察者进行评估。记录滑脱距离、方向、与远端螺钉密度的关系和观察时间,以及ARS对尾曲线增加的影响。为防止滑脱,用不同的结构末端螺钉(Reline,NuVasive Inc:ARS风险为27%(56/205),距离为3.6 ± 2.2 mm,主要为凸面。42%发生在 4 个月前,其余发生在 1 年前。ARS患者的尾部曲线大幅增加的频率是正常人的三倍。观察者之间的可靠性很高,滑移方向符合预期。ARS与远端螺钉密度无关。各中心之间的ARS发生率存在显著差异(53%、31%、13%),而平均螺钉密度(≈1.3枚螺钉/水平)或曲线矫正率(≈60%)则没有差异。2.9%的患者(6/207)需要进行ARS翻修手术。使用不同的结构末端螺钉,ARS风险仅为2%(1/56),且无需进行翻修手术:这项研究证明了脊柱侧弯手术中结构末端轴向杆滑脱的发生率及其临床意义。虽然轻微的ARS可能是亚临床症状,但不应将ARS误认为是附加症状。当L3为LIV时,最严重的ARS主要发生在高负荷远端螺钉的凸面。较长的结构(LIV L3或L4)发生ARS的风险较高。在我们的系列研究中,使用另一颗结构末端螺钉发生ARS的风险极低,这凸显了螺钉类型对ARS发生的影响。进一步的研究对完善技术和提高患者预后至关重要。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
3.20
自引率
18.80%
发文量
167
期刊介绍: Spine Deformity the official journal of the?Scoliosis Research Society is a peer-refereed publication to disseminate knowledge on basic science and clinical research into the?etiology?biomechanics?treatment?methods and outcomes of all types of?spinal deformities. The international members of the Editorial Board provide a worldwide perspective for the journal's area of interest.The?journal?will enhance the mission of the Society which is to foster the optimal care of all patients with?spine?deformities worldwide. Articles published in?Spine Deformity?are Medline indexed in PubMed.? The journal publishes original articles in the form of clinical and basic research. Spine Deformity will only publish studies that have institutional review board (IRB) or similar ethics committee approval for human and animal studies and have strictly observed these guidelines. The minimum follow-up period for follow-up clinical studies is 24 months.
期刊最新文献
Human spinal height growth: a description of normal spine growth patterns and adult spine height prediction from a longitudinal cohort. Making wrong site surgery a "never event" in spinal deformity surgery by use of a "landmark vertebra" to eliminate variability in identifying a target vertebral level. Magnetically controlled growing rods increase 3D true spine length in idiopathic early onset scoliosis patients: results from a multicenter study. Factors contributing to severe scoliosis after open chest surgery for congenital heart disease: a case-control analysis. Zones where reduced implant density leads to correction loss after scoliosis surgery for Lenke 1A adolescent idiopathic scoliosis: a multicenter study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1