{"title":"Validation of Noninvasive Methemoglobin and Carboxyhemoglobin Measurements Using Pulse Co-Oximeter in Healthy Dogs.","authors":"Jiwoong Her, Justin Roh, Deborah A Keys","doi":"10.2147/VMRR.S465945","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To assess the agreement between pulse co-oximeter and blood co-oximeter in measuring methemoglobin (MetHb) and carboxyhemoglobin (COHb) in healthy, awake dogs.</p><p><strong>Patients and methods: </strong>Forty-five healthy dogs were enrolled in the study. Pulse co-oximetry was performed using the Masimo Radical 7 Pulse Co-Oximeter with a Rainbow<sup>®</sup> adhesive sensor. Simultaneously, venous blood samples were collected, and MetHb and COHb were immediately measured using a Stat Profile Prime Plus VET Critical Care Analyzer. Paired measurements of MetHb and COHb were evaluated via Spearman correlation, intra-class correlation (ICC), and Bland-Altman plots to evaluate the degree of agreement between the pulse co-oximeter and the blood co-oximeter.</p><p><strong>Results: </strong>A total of 45 paired MetHb and COHb measurements were collected. There was a weak correlation between the pulse co-oximeter and the blood co-oximeter readings. The correlation coefficients for MetHb and COHb were 0.0 (95% CI, -0.3 to 0.3) and 0.03 (95% CI, -0.27 to 0.32), respectively. The ICC indicated poor agreement between the pulse and blood co-oximeter for MetHb (ICC = 0.00, 95% CI: -0.12 to 0.15) and COHb (ICC = 0.03, 95% CI: -0.27 to 0.33). Bland-Altman plots revealed low mean bias but wide limits of agreement, indicating that the pulse co-oximeter overestimated MetHb by on average of 0.7% (P < 0.0001) (95% LoA: -0.5 to 2.0) and COHb by on average 0.2% (P = 0.59) (95% LoA: -4.6 to 5.0).</p><p><strong>Conclusion: </strong>Obtaining MetHb and COHb measurements with the Masimo Radical 7 Pulse Co-Oximeter is straightforward in healthy, awake dogs. However, the device does not provide accurate measurements compared to the blood co-oximeter, specifically in the range of MetHb and COHb in healthy dogs, based on the wide LoA.</p>","PeriodicalId":75300,"journal":{"name":"Veterinary medicine (Auckland, N.Z.)","volume":"15 ","pages":"197-203"},"PeriodicalIF":1.7000,"publicationDate":"2024-08-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11334929/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Veterinary medicine (Auckland, N.Z.)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2147/VMRR.S465945","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"VETERINARY SCIENCES","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: To assess the agreement between pulse co-oximeter and blood co-oximeter in measuring methemoglobin (MetHb) and carboxyhemoglobin (COHb) in healthy, awake dogs.
Patients and methods: Forty-five healthy dogs were enrolled in the study. Pulse co-oximetry was performed using the Masimo Radical 7 Pulse Co-Oximeter with a Rainbow® adhesive sensor. Simultaneously, venous blood samples were collected, and MetHb and COHb were immediately measured using a Stat Profile Prime Plus VET Critical Care Analyzer. Paired measurements of MetHb and COHb were evaluated via Spearman correlation, intra-class correlation (ICC), and Bland-Altman plots to evaluate the degree of agreement between the pulse co-oximeter and the blood co-oximeter.
Results: A total of 45 paired MetHb and COHb measurements were collected. There was a weak correlation between the pulse co-oximeter and the blood co-oximeter readings. The correlation coefficients for MetHb and COHb were 0.0 (95% CI, -0.3 to 0.3) and 0.03 (95% CI, -0.27 to 0.32), respectively. The ICC indicated poor agreement between the pulse and blood co-oximeter for MetHb (ICC = 0.00, 95% CI: -0.12 to 0.15) and COHb (ICC = 0.03, 95% CI: -0.27 to 0.33). Bland-Altman plots revealed low mean bias but wide limits of agreement, indicating that the pulse co-oximeter overestimated MetHb by on average of 0.7% (P < 0.0001) (95% LoA: -0.5 to 2.0) and COHb by on average 0.2% (P = 0.59) (95% LoA: -4.6 to 5.0).
Conclusion: Obtaining MetHb and COHb measurements with the Masimo Radical 7 Pulse Co-Oximeter is straightforward in healthy, awake dogs. However, the device does not provide accurate measurements compared to the blood co-oximeter, specifically in the range of MetHb and COHb in healthy dogs, based on the wide LoA.