{"title":"Normalization of increasing shocking coil impedance with full output synchronized shock","authors":"Takahide Kadosaka MD, Masaya Watanabe MD, PhD, Motoki Nakao MD, PhD, Taro Koya MD, PhD, Taro Temma MD, PhD, Toshihisa Anzai MD, PhD","doi":"10.1111/jce.16416","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Introduction</h3>\n \n <p>Impedance is a crucial parameter in cardiovascular implantable electronic devices (CIEDs). Clinically, most CIEDs measure impedance using low voltage sub-threshold measurement (LVSM). Although the LVSM of shock impedance (LVSM-SI) is generally comparable with high voltage shock impedance (HVSI), LVSM-SI might be inaccurate if peri-lead tissue degeneration occurs.</p>\n </section>\n \n <section>\n \n <h3> Methods and Results</h3>\n \n <p>We present a case of elevated LVSM-SI occurring 8 years post-lead implantation, possibly attributed to encapsulation of the right ventricular lead coil. After 0.1 J shock was delivered, a full output synchronized shock was administered to measure HVSI, revealing a normal value. Furthermore, LVSM-SI was normalized and maintained within the normal range during long-term follow-up.</p>\n </section>\n \n <section>\n \n <h3> Conclusion</h3>\n \n <p>Our findings suggest conducting a full-output synchronized shock test to assess HVSI when abnormal LVSM-SI is detected in the remote phase post-ICD implantation, which may be considered to help normalize LVSM shock impedance.</p>\n </section>\n </div>","PeriodicalId":15178,"journal":{"name":"Journal of Cardiovascular Electrophysiology","volume":"35 11","pages":"2251-2253"},"PeriodicalIF":2.6000,"publicationDate":"2024-08-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Cardiovascular Electrophysiology","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/jce.16416","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction
Impedance is a crucial parameter in cardiovascular implantable electronic devices (CIEDs). Clinically, most CIEDs measure impedance using low voltage sub-threshold measurement (LVSM). Although the LVSM of shock impedance (LVSM-SI) is generally comparable with high voltage shock impedance (HVSI), LVSM-SI might be inaccurate if peri-lead tissue degeneration occurs.
Methods and Results
We present a case of elevated LVSM-SI occurring 8 years post-lead implantation, possibly attributed to encapsulation of the right ventricular lead coil. After 0.1 J shock was delivered, a full output synchronized shock was administered to measure HVSI, revealing a normal value. Furthermore, LVSM-SI was normalized and maintained within the normal range during long-term follow-up.
Conclusion
Our findings suggest conducting a full-output synchronized shock test to assess HVSI when abnormal LVSM-SI is detected in the remote phase post-ICD implantation, which may be considered to help normalize LVSM shock impedance.
期刊介绍:
Journal of Cardiovascular Electrophysiology (JCE) keeps its readership well informed of the latest developments in the study and management of arrhythmic disorders. Edited by Bradley P. Knight, M.D., and a distinguished international editorial board, JCE is the leading journal devoted to the study of the electrophysiology of the heart.