Normalization of increasing shocking coil impedance with full output synchronized shock.

IF 2.3 3区 医学 Q2 CARDIAC & CARDIOVASCULAR SYSTEMS Journal of Cardiovascular Electrophysiology Pub Date : 2024-08-21 DOI:10.1111/jce.16416
Takahide Kadosaka, Masaya Watanabe, Motoki Nakao, Taro Koya, Taro Temma, Toshihisa Anzai
{"title":"Normalization of increasing shocking coil impedance with full output synchronized shock.","authors":"Takahide Kadosaka, Masaya Watanabe, Motoki Nakao, Taro Koya, Taro Temma, Toshihisa Anzai","doi":"10.1111/jce.16416","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Impedance is a crucial parameter in cardiovascular implantable electronic devices (CIEDs). Clinically, most CIEDs measure impedance using low voltage sub-threshold measurement (LVSM). Although the LVSM of shock impedance (LVSM-SI) is generally comparable with high voltage shock impedance (HVSI), LVSM-SI might be inaccurate if peri-lead tissue degeneration occurs.</p><p><strong>Methods and results: </strong>We present a case of elevated LVSM-SI occurring 8 years post-lead implantation, possibly attributed to encapsulation of the right ventricular lead coil. After 0.1 J shock was delivered, a full output synchronized shock was administered to measure HVSI, revealing a normal value. Furthermore, LVSM-SI was normalized and maintained within the normal range during long-term follow-up.</p><p><strong>Conclusion: </strong>Our findings suggest conducting a full-output synchronized shock test to assess HVSI when abnormal LVSM-SI is detected in the remote phase post-ICD implantation, which may be considered to help normalize LVSM shock impedance.</p>","PeriodicalId":15178,"journal":{"name":"Journal of Cardiovascular Electrophysiology","volume":null,"pages":null},"PeriodicalIF":2.3000,"publicationDate":"2024-08-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Cardiovascular Electrophysiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1111/jce.16416","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Impedance is a crucial parameter in cardiovascular implantable electronic devices (CIEDs). Clinically, most CIEDs measure impedance using low voltage sub-threshold measurement (LVSM). Although the LVSM of shock impedance (LVSM-SI) is generally comparable with high voltage shock impedance (HVSI), LVSM-SI might be inaccurate if peri-lead tissue degeneration occurs.

Methods and results: We present a case of elevated LVSM-SI occurring 8 years post-lead implantation, possibly attributed to encapsulation of the right ventricular lead coil. After 0.1 J shock was delivered, a full output synchronized shock was administered to measure HVSI, revealing a normal value. Furthermore, LVSM-SI was normalized and maintained within the normal range during long-term follow-up.

Conclusion: Our findings suggest conducting a full-output synchronized shock test to assess HVSI when abnormal LVSM-SI is detected in the remote phase post-ICD implantation, which may be considered to help normalize LVSM shock impedance.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
通过全输出同步电击,使不断增大的电击线圈阻抗正常化。
简介阻抗是心血管植入式电子设备(CIED)的一个重要参数。临床上,大多数 CIED 采用低压阈下测量法(LVSM)测量阻抗。虽然冲击阻抗(LVSM-SI)通常与高压冲击阻抗(HVSI)相当,但如果发生导联周围组织变性,LVSM-SI 可能会不准确:我们介绍了一例导联植入 8 年后出现 LVSM-SI 升高的病例,其原因可能是右心室导联线圈被包裹。在进行 0.1 J 电击后,进行全输出同步电击以测量 HVSI,结果显示数值正常。此外,LVSM-SI 恢复正常,并在长期随访中保持在正常范围内:我们的研究结果表明,当在植入 ICD 后的远期阶段发现 LVSM-SI 异常时,可考虑进行全输出同步电击测试以评估 HVSI,从而帮助 LVSM 冲击阻抗恢复正常。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
5.20
自引率
14.80%
发文量
433
审稿时长
3-6 weeks
期刊介绍: Journal of Cardiovascular Electrophysiology (JCE) keeps its readership well informed of the latest developments in the study and management of arrhythmic disorders. Edited by Bradley P. Knight, M.D., and a distinguished international editorial board, JCE is the leading journal devoted to the study of the electrophysiology of the heart.
期刊最新文献
Assessment of palpitations in patients with frequent premature ventricular contractions. Body mass index as a determinant of scar formation post-AF ablation: Insights from DECAAF II. Efficacy and limitation of nonparoxysmal atrial fibrillation ablation in patients with heart failure with preserved ejection fraction. Frequency and Genotype-Dependence of intrinsic chronotropic insufficiency among patients with congenital long QT syndrome. Endocardial ablation of ganglionated plexus for the treatment of carotid sinus syndrome.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1