{"title":"HeartMate 3 driveline damage by gradual corrosion due to liquid infiltration: a case report.","authors":"Tomo Yoshizumi, Hideki Ito, Sachie Terazawa, Yoshiyuki Tokuda, Hajime Sakurai, Yuji Narita, Masato Mutsuga","doi":"10.1007/s10047-024-01464-w","DOIUrl":null,"url":null,"abstract":"<p><p>A 31-year-old man with a HeartMate 3 was admitted with a \"Driveline Communication Fault\" alarm due to liquid infiltration. Eleven months earlier, the connector between the modular and pump cables had gotten wet when he was taking a shower. The cable connector was dried immediately, and no alarm had been observed during follow-up. Subsequently, the modular cable and system controller were replaced, with corrosion found on the modular cable connector. The \"Communication Fault\" alarm recurred, and complete damage to the communication cables was discovered. The pump was replaced, and the removed pump cable connector showed corrosion as well. If the driveline connector gets wet, the multidisciplinary team should discuss if it should be immediately disconnected and dried, averting the need for future pump replacements due to corrosion.</p>","PeriodicalId":15177,"journal":{"name":"Journal of Artificial Organs","volume":" ","pages":""},"PeriodicalIF":1.1000,"publicationDate":"2024-08-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Artificial Organs","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1007/s10047-024-01464-w","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0
Abstract
A 31-year-old man with a HeartMate 3 was admitted with a "Driveline Communication Fault" alarm due to liquid infiltration. Eleven months earlier, the connector between the modular and pump cables had gotten wet when he was taking a shower. The cable connector was dried immediately, and no alarm had been observed during follow-up. Subsequently, the modular cable and system controller were replaced, with corrosion found on the modular cable connector. The "Communication Fault" alarm recurred, and complete damage to the communication cables was discovered. The pump was replaced, and the removed pump cable connector showed corrosion as well. If the driveline connector gets wet, the multidisciplinary team should discuss if it should be immediately disconnected and dried, averting the need for future pump replacements due to corrosion.
期刊介绍:
The aim of the Journal of Artificial Organs is to introduce to colleagues worldwide a broad spectrum of important new achievements in the field of artificial organs, ranging from fundamental research to clinical applications. The scope of the Journal of Artificial Organs encompasses but is not restricted to blood purification, cardiovascular intervention, biomaterials, and artificial metabolic organs. Additionally, the journal will cover technical and industrial innovations. Membership in the Japanese Society for Artificial Organs is not a prerequisite for submission.