HeartMate 3 driveline damage by gradual corrosion due to liquid infiltration: a case report.

IF 1.1 4区 医学 Q4 ENGINEERING, BIOMEDICAL Journal of Artificial Organs Pub Date : 2024-08-27 DOI:10.1007/s10047-024-01464-w
Tomo Yoshizumi, Hideki Ito, Sachie Terazawa, Yoshiyuki Tokuda, Hajime Sakurai, Yuji Narita, Masato Mutsuga
{"title":"HeartMate 3 driveline damage by gradual corrosion due to liquid infiltration: a case report.","authors":"Tomo Yoshizumi, Hideki Ito, Sachie Terazawa, Yoshiyuki Tokuda, Hajime Sakurai, Yuji Narita, Masato Mutsuga","doi":"10.1007/s10047-024-01464-w","DOIUrl":null,"url":null,"abstract":"<p><p>A 31-year-old man with a HeartMate 3 was admitted with a \"Driveline Communication Fault\" alarm due to liquid infiltration. Eleven months earlier, the connector between the modular and pump cables had gotten wet when he was taking a shower. The cable connector was dried immediately, and no alarm had been observed during follow-up. Subsequently, the modular cable and system controller were replaced, with corrosion found on the modular cable connector. The \"Communication Fault\" alarm recurred, and complete damage to the communication cables was discovered. The pump was replaced, and the removed pump cable connector showed corrosion as well. If the driveline connector gets wet, the multidisciplinary team should discuss if it should be immediately disconnected and dried, averting the need for future pump replacements due to corrosion.</p>","PeriodicalId":15177,"journal":{"name":"Journal of Artificial Organs","volume":null,"pages":null},"PeriodicalIF":1.1000,"publicationDate":"2024-08-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Artificial Organs","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1007/s10047-024-01464-w","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

A 31-year-old man with a HeartMate 3 was admitted with a "Driveline Communication Fault" alarm due to liquid infiltration. Eleven months earlier, the connector between the modular and pump cables had gotten wet when he was taking a shower. The cable connector was dried immediately, and no alarm had been observed during follow-up. Subsequently, the modular cable and system controller were replaced, with corrosion found on the modular cable connector. The "Communication Fault" alarm recurred, and complete damage to the communication cables was discovered. The pump was replaced, and the removed pump cable connector showed corrosion as well. If the driveline connector gets wet, the multidisciplinary team should discuss if it should be immediately disconnected and dried, averting the need for future pump replacements due to corrosion.

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
HeartMate 3 传动系统因液体渗入而逐渐腐蚀损坏:病例报告。
一名使用 HeartMate 3 的 31 岁男子因液体渗入导致 "传动系通信故障 "报警而入院。11 个月前,他在洗澡时弄湿了模块和泵电缆之间的连接器。电缆连接器立即被烘干,在后续处理过程中也未发现警报。随后,更换了模块电缆和系统控制器,并在模块电缆连接器上发现了腐蚀。通信故障 "警报再次出现,并发现通信电缆完全损坏。更换了泵,拆下的泵电缆连接器也出现了腐蚀。如果传动系统连接器受潮,多学科小组应讨论是否应立即断开并擦干,以避免今后因腐蚀而更换泵。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Journal of Artificial Organs
Journal of Artificial Organs 医学-工程:生物医学
CiteScore
2.80
自引率
15.40%
发文量
68
审稿时长
6-12 weeks
期刊介绍: The aim of the Journal of Artificial Organs is to introduce to colleagues worldwide a broad spectrum of important new achievements in the field of artificial organs, ranging from fundamental research to clinical applications. The scope of the Journal of Artificial Organs encompasses but is not restricted to blood purification, cardiovascular intervention, biomaterials, and artificial metabolic organs. Additionally, the journal will cover technical and industrial innovations. Membership in the Japanese Society for Artificial Organs is not a prerequisite for submission.
期刊最新文献
Exploring the role of plasmapheresis prior to thyroidectomy in managing thyrotoxicosis: a comprehensive scoping review. A compact machine perfusion device for whole blood perfusion in isolated rat liver. Advanced roll porous scaffold 3D bioprinting technology. Impact of the spleen size on short-term prognosis in patients with cardiogenic shock receiving Impella-incorporated temporary mechanical circulatory support Axial-flow polymer bridge pump with hydrodynamic bearings
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1