Janaína de Oliveira Abrahão, Rodrigo Hermont Cançado, Esdras de Campos França, Leandro de Arruda Santos, Pedro Damas Resende, Felipe Weidenbach Degrazia, Leniana Santos Neves
{"title":"Influence of distal-end heat treatment in the properties of heat-activated NiTi archwires.","authors":"Janaína de Oliveira Abrahão, Rodrigo Hermont Cançado, Esdras de Campos França, Leandro de Arruda Santos, Pedro Damas Resende, Felipe Weidenbach Degrazia, Leniana Santos Neves","doi":"10.1007/s00056-024-00547-w","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>The aim of this study was to evaluate the extent of property changes caused by heating the distal portion of heat-activated nickel-titanium (NiTi) wires.</p><p><strong>Methods: </strong>Forty preformed heat-activated NiTi archwires (3M Unitek, Monrovia, CA, USA) with a nominal cross-section of 0.018″ were used in this study. The archwires were divided into a control group, not submitted to heat treatment and, thus, maintaining the as-received properties, and an experimental group, in which the archwires were submitted to heat treatment for distal bending at one end. Wire segments of control and experimental groups were submitted to differential scanning calorimetry (DSC) and Vickers microhardness measurements.</p><p><strong>Results: </strong>The DSC results suggest local recrystallization and precipitate dissolution at the heat-treated tip, which decreases as the distance to the wire's tip increases. Vickers microhardness tests revealed significant changes for distances between 6 and 8 mm from the wire's tip. Heating the distal portion of heat-activated NiTi archwires should be performed with care since this clinical procedure may compromise the performance of these wires to a distance of 8 mm from the archwire end.</p><p><strong>Conclusion: </strong>Heat treatment for distal bending in heat-activated NiTi archwires may be performed, with little impact on the areas adjacent to heat treatment. In cases presenting molars requiring significant orthodontic corrections, it should be preferred to apply other techniques to avoid archwire sliding, such as crimpable stops, or to have flame control to avoid placing a heat-treated section in the tubes of these molars.</p>","PeriodicalId":54776,"journal":{"name":"Journal of Orofacial Orthopedics-Fortschritte Der Kieferorthopadie","volume":" ","pages":""},"PeriodicalIF":1.3000,"publicationDate":"2024-08-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Orofacial Orthopedics-Fortschritte Der Kieferorthopadie","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00056-024-00547-w","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: The aim of this study was to evaluate the extent of property changes caused by heating the distal portion of heat-activated nickel-titanium (NiTi) wires.
Methods: Forty preformed heat-activated NiTi archwires (3M Unitek, Monrovia, CA, USA) with a nominal cross-section of 0.018″ were used in this study. The archwires were divided into a control group, not submitted to heat treatment and, thus, maintaining the as-received properties, and an experimental group, in which the archwires were submitted to heat treatment for distal bending at one end. Wire segments of control and experimental groups were submitted to differential scanning calorimetry (DSC) and Vickers microhardness measurements.
Results: The DSC results suggest local recrystallization and precipitate dissolution at the heat-treated tip, which decreases as the distance to the wire's tip increases. Vickers microhardness tests revealed significant changes for distances between 6 and 8 mm from the wire's tip. Heating the distal portion of heat-activated NiTi archwires should be performed with care since this clinical procedure may compromise the performance of these wires to a distance of 8 mm from the archwire end.
Conclusion: Heat treatment for distal bending in heat-activated NiTi archwires may be performed, with little impact on the areas adjacent to heat treatment. In cases presenting molars requiring significant orthodontic corrections, it should be preferred to apply other techniques to avoid archwire sliding, such as crimpable stops, or to have flame control to avoid placing a heat-treated section in the tubes of these molars.
期刊介绍:
The Journal of Orofacial Orthopedics provides orthodontists and dentists who are also actively interested in orthodontics, whether in university clinics or private practice, with highly authoritative and up-to-date information based on experimental and clinical research. The journal is one of the leading publications for the promulgation of the results of original work both in the areas of scientific and clinical orthodontics and related areas. All articles undergo peer review before publication. The German Society of Orthodontics (DGKFO) also publishes in the journal important communications, statements and announcements.