Evaluation of agreement of IOP measurements by Tono-Vera tonometer to Goldmann applanation tonometry.

Frontiers in ophthalmology Pub Date : 2024-08-16 eCollection Date: 2024-01-01 DOI:10.3389/fopht.2024.1441343
Charles R Niles, Alexander R Crinzi, Russell Bonaventura, David A Taylor
{"title":"Evaluation of agreement of IOP measurements by Tono-Vera tonometer to Goldmann applanation tonometry.","authors":"Charles R Niles, Alexander R Crinzi, Russell Bonaventura, David A Taylor","doi":"10.3389/fopht.2024.1441343","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the accuracy of the new Tono-Vera rebound tonometer (Reichert Inc, Buffalo, NY) compared to Goldmann Applanation Tonometry.</p><p><strong>Methods: </strong>This prospective, observational, cross-sectional study was designed in accordance with ANSI Z80.10-2014 and ISO 8612-2009 guidelines for tonometer comparison. Intraocular Pressure (IOP) was measured by Goldmann Applanation and Tono-Vera on 160 eyes of 160 subjects. Corneal Astigmatism and Central Corneal Thickness were also measured. A single investigator (CN) conducted all measurements. The average of two measurements from each tonometer was used in the analysis. Bland-Altman plots, total least squares regression analysis, and simple linear regression were used to evaluate agreement between the tonometers.</p><p><strong>Results: </strong>Average IOP values from Goldmann Applanation and Tono-Vera were not significantly different (19.17 and 19.03 respectively, p=0.40, paired t-test). The total least squares regression analysis indicated strong agreement between the two tonometers (slope +0.97, offset +0.49 mmHg, standard deviation 2.11 mmHg). There were 2 IOP measurement pairs that exceeded the ± 5 mmHg limits of agreement required in ANSI Z80.10-2014 and ISO 8612-2009, which is within the range of acceptability specified in the standards.</p><p><strong>Conclusion: </strong>We evaluated IOP measurements by Tono-Vera Rebound Tonometer vs Goldmann Applanation Tonometry for eyes with a wide range of IOP values and found no statistically significant differences in the results. Tono-Vera meets the requirements of ANSI Z80.10-2014 and ISO 8612-2009, demonstrating accuracy comparable to Goldmann tonometry.</p>","PeriodicalId":73096,"journal":{"name":"Frontiers in ophthalmology","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-08-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11362087/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Frontiers in ophthalmology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3389/fopht.2024.1441343","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To evaluate the accuracy of the new Tono-Vera rebound tonometer (Reichert Inc, Buffalo, NY) compared to Goldmann Applanation Tonometry.

Methods: This prospective, observational, cross-sectional study was designed in accordance with ANSI Z80.10-2014 and ISO 8612-2009 guidelines for tonometer comparison. Intraocular Pressure (IOP) was measured by Goldmann Applanation and Tono-Vera on 160 eyes of 160 subjects. Corneal Astigmatism and Central Corneal Thickness were also measured. A single investigator (CN) conducted all measurements. The average of two measurements from each tonometer was used in the analysis. Bland-Altman plots, total least squares regression analysis, and simple linear regression were used to evaluate agreement between the tonometers.

Results: Average IOP values from Goldmann Applanation and Tono-Vera were not significantly different (19.17 and 19.03 respectively, p=0.40, paired t-test). The total least squares regression analysis indicated strong agreement between the two tonometers (slope +0.97, offset +0.49 mmHg, standard deviation 2.11 mmHg). There were 2 IOP measurement pairs that exceeded the ± 5 mmHg limits of agreement required in ANSI Z80.10-2014 and ISO 8612-2009, which is within the range of acceptability specified in the standards.

Conclusion: We evaluated IOP measurements by Tono-Vera Rebound Tonometer vs Goldmann Applanation Tonometry for eyes with a wide range of IOP values and found no statistically significant differences in the results. Tono-Vera meets the requirements of ANSI Z80.10-2014 and ISO 8612-2009, demonstrating accuracy comparable to Goldmann tonometry.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
评估 Tono-Vera 眼压计与戈德曼眼压计测量眼压的一致性。
目的:评估新型 Tono-Vera 回弹式眼压计(Reichert Inc,Buffalo,NY)与戈德曼眼压计(Goldmann Applanation Tonometry)相比的准确性:这项前瞻性、观察性、横断面研究的设计符合 ANSI Z80.10-2014 和 ISO 8612-2009 关于眼压计比较的指南。通过戈德曼角膜塑形镜和 Tono-Vera 对 160 名受试者的 160 只眼睛进行眼压测量。此外,还测量了角膜散光和角膜中央厚度。所有测量均由一名研究人员(CN)进行。分析中使用了每个眼压计两次测量的平均值。使用Bland-Altman图、总最小二乘法回归分析和简单线性回归来评估眼压计之间的一致性:结果:Goldmann Applanation 和 Tono-Vera 的平均眼压值差异不大(分别为 19.17 和 19.03,p=0.40,配对 t 检验)。总最小二乘法回归分析表明,两种眼压计的测量结果非常一致(斜率 +0.97,偏移 +0.49mmHg,标准偏差 2.11mmHg)。有 2 对眼压测量值超过了 ANSI Z80.10-2014 和 ISO 8612-2009 所要求的 ± 5 mmHg 的一致性限值,这在标准规定的可接受范围之内:我们评估了 Tono-Vera 回弹式眼压计与戈德曼角膜屈光计对眼压值范围较宽的眼睛进行的眼压测量,结果没有发现统计学上的显著差异。Tono-Vera 符合 ANSI Z80.10-2014 和 ISO 8612-2009 的要求,其准确性可与戈德曼眼压计媲美。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
0.50
自引率
0.00%
发文量
0
期刊最新文献
Comparison of the PlusoptiX A16 and vision screener V100. UV light and the ocular lens: a review of exposure models and resulting biomolecular changes. Diagnostic accuracy of a modularized, virtual-reality-based automated pupillometer for detection of relative afferent pupillary defect in unilateral optic neuropathies. Advances in the management of intraocular foreign bodies. Large animal model species in pluripotent stem cell therapy research and development for retinal diseases: a systematic review.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1