Hemodynamic effect of supra-annular implantation of SAPIEN 3 balloon expandable valve.

IF 3.1 Q2 CARDIAC & CARDIOVASCULAR SYSTEMS Cardiovascular Intervention and Therapeutics Pub Date : 2024-09-06 DOI:10.1007/s12928-024-01040-2
Masakazu Yasuda, Kazuki Mizutani, Kyohei Onishi, Naoko Onishi, Kosuke Fujita, Masafumi Ueno, Atsushi Okamura, Yoshitaka Iwanaga, Genichi Sakaguchi, Gaku Nakazawa
{"title":"Hemodynamic effect of supra-annular implantation of SAPIEN 3 balloon expandable valve.","authors":"Masakazu Yasuda, Kazuki Mizutani, Kyohei Onishi, Naoko Onishi, Kosuke Fujita, Masafumi Ueno, Atsushi Okamura, Yoshitaka Iwanaga, Genichi Sakaguchi, Gaku Nakazawa","doi":"10.1007/s12928-024-01040-2","DOIUrl":null,"url":null,"abstract":"<p><p>The hemodynamic impact of the implantation depth for balloon-expandable valves is under-investigated, especially with higher implantation techniques. We assessed the hemodynamic performance of supra-annular SAPIEN 3 valve implantation. This retrospective study involved consecutive patients who underwent transcatheter aortic valve replacement (TAVR) using the SAPIEN 3. The device implantation depth and transcatheter heart valve (THV) leaflet-nadir position were angiographically analyzed, and supra-annular implantation was defined as a higher leaflet-nadir position than the original annular line. The Doppler hemodynamic status was evaluated at patient discharge. Among 184 patients, 120 (65%) underwent supra-annular implantation, and their mean implantation depth was significantly lower than that of intra-annular implantation (1.1 vs. 5.2 mm, p < 0.001). No patients developed valve embolization or coronary occlusion, and none required the TAV-in-TAV procedure. Two (1.6%) patients in the supra-annular implantation group had a mild or greater paravalvular leak. Echocardiography demonstrated that supra-annular implantation had better hemodynamic performance, showing a larger indexed effective orifice area (iEOA) compared with intra-annular implantation (1.09 vs. 0.97 cm<sup>2</sup>/m<sup>2</sup>, p < 0.01). There was a weak but negative correlation between the implantation depth and iEOA (r =  - 0.27, p < 0.01). Moderate or severe prosthesis-patient mismatch (PPM) was found in 35.9% of the intra-annular group and 9.2% of supra-annular of the supra-annular group (p < 0.01). In the multivariable analysis, supra-annular implantation was an independent predictor of better THV function (iEOA > 0.85). Supra-annular SAPIEN 3 implantation provides beneficial hemodynamic effects and reduces the PPM risk.</p>","PeriodicalId":9439,"journal":{"name":"Cardiovascular Intervention and Therapeutics","volume":" ","pages":""},"PeriodicalIF":3.1000,"publicationDate":"2024-09-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cardiovascular Intervention and Therapeutics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s12928-024-01040-2","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

The hemodynamic impact of the implantation depth for balloon-expandable valves is under-investigated, especially with higher implantation techniques. We assessed the hemodynamic performance of supra-annular SAPIEN 3 valve implantation. This retrospective study involved consecutive patients who underwent transcatheter aortic valve replacement (TAVR) using the SAPIEN 3. The device implantation depth and transcatheter heart valve (THV) leaflet-nadir position were angiographically analyzed, and supra-annular implantation was defined as a higher leaflet-nadir position than the original annular line. The Doppler hemodynamic status was evaluated at patient discharge. Among 184 patients, 120 (65%) underwent supra-annular implantation, and their mean implantation depth was significantly lower than that of intra-annular implantation (1.1 vs. 5.2 mm, p < 0.001). No patients developed valve embolization or coronary occlusion, and none required the TAV-in-TAV procedure. Two (1.6%) patients in the supra-annular implantation group had a mild or greater paravalvular leak. Echocardiography demonstrated that supra-annular implantation had better hemodynamic performance, showing a larger indexed effective orifice area (iEOA) compared with intra-annular implantation (1.09 vs. 0.97 cm2/m2, p < 0.01). There was a weak but negative correlation between the implantation depth and iEOA (r =  - 0.27, p < 0.01). Moderate or severe prosthesis-patient mismatch (PPM) was found in 35.9% of the intra-annular group and 9.2% of supra-annular of the supra-annular group (p < 0.01). In the multivariable analysis, supra-annular implantation was an independent predictor of better THV function (iEOA > 0.85). Supra-annular SAPIEN 3 implantation provides beneficial hemodynamic effects and reduces the PPM risk.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
环上植入 SAPIEN 3 球囊扩张瓣的血液动力学效应。
对球囊扩张瓣膜植入深度对血流动力学的影响研究不足,尤其是对较高植入技术的影响。我们评估了环上 SAPIEN 3 瓣膜植入术的血液动力学表现。这项回顾性研究涉及使用 SAPIEN 3 进行经导管主动脉瓣置换术(TAVR)的连续患者。通过血管造影分析了装置植入深度和经导管心脏瓣膜(THV)的瓣叶脐位置,瓣上植入的定义是瓣叶脐位置高于原始瓣环线。患者出院时对多普勒血流动力学状态进行了评估。在184名患者中,120人(65%)进行了瓣上植入,其平均植入深度明显低于瓣内植入(1.1 mm vs. 5.2 mm, p 2/m2, p 0.85)。环上 SAPIEN 3 植入术可提供有益的血流动力学效应并降低 PPM 风险。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Cardiovascular Intervention and Therapeutics
Cardiovascular Intervention and Therapeutics CARDIAC & CARDIOVASCULAR SYSTEMS-
CiteScore
6.30
自引率
12.50%
发文量
68
期刊介绍: Cardiovascular Intervention and Therapeutics (CVIT) is an international journal covering the field of cardiovascular disease and includes cardiac (coronary and noncoronary) and peripheral interventions and therapeutics. Articles are subject to peer review and complete editorial evaluation prior to any decision regarding acceptability. CVIT is an official journal of The Japanese Association of Cardiovascular Intervention and Therapeutics.
期刊最新文献
3D printing-enhanced transcatheter closure of residual shunts post-ventricular septal perforation: multimodal imaging's crucial role. Physician attending cardiac magnetic resonance imaging for early aortic structural valve deterioration: non-invasive method to identify the regurgitate location. Efficacy of R-N cusp-overlap view as a complementary approach for optimal neocommissural alignment with Navitor valve. Assessment of fractal ratio of vessel diameters in coronary bifurcation lesions by angiography and intravascular ultrasound. Uncrossable aortic stenosis during transcatheter aortic valve replacement: application of balloon-assisted tracking technique.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1