{"title":"Influence of connection design and material properties on stress distribution and fatigue lifetime of zygomatic implants: A finite element analysis","authors":"Laura H.J. Alberto , Saadman Alamgir , Jason A. Griggs , Michael D. Roach , Randal S. Williamson , Ravi Chandran , Matthew W. Priddy , Yuefeng J. Lu , Patrick F. Bergin , Yuanyuan Duan","doi":"10.1016/j.jmbbm.2024.106723","DOIUrl":null,"url":null,"abstract":"<div><p>Zygomatic implants (ZIs) were developed as a graftless alternative to rehabilitate severely reabsorbed maxillae. This study aims to employ three-dimensional finite element analysis (FEA) to simulate the impact of external hexagonal implant connection (EHC) and internal hexagonal implant connection (IHC) on the stress distribution and fatigue lifetime within the ZI systems using parameters defined in ISO 14801:2016. Two ZI assemblies (Nobel Biocare and Noris Medical) were scanned in a micro-CT scanner and reconstructed using Nrecon software. Three-dimensional models were generated by Simpleware ScanIP Medical software. All models were exported to FEA software (ABAQUS) and subsequently to a fatigue analysis software (Fe-safe). A compressive 150 N load was applied at a 40° angle on the cap surface. A 15 Hz frequency was applied in the in silico cyclic test. The implant components had material properties of commercially pure grade 4 titanium (CPTi) and Titanium-6Aluminum-4Vanadium alloy (Ti64). Von Mises stress data, contour plots, and fatigue limits were collected and analyzed. EHC models exhibited higher peak stresses in implant components for both materials compared to IHC models. However, simulated bone support results showed the opposite trend, with higher stresses on IHCthan EHC models. The fatigue analysis revealed that assemblies with both designs exceeded ISO 14801:2016 number of cycles limits using Ti64, while CPTi groups exhibited comparatively lower worst life-repeats. In conclusion, ZIs with IHC were found to have a more homogeneous and advantageous stress distribution within both materials tested. Ti64 demonstrates a prolonged service life for both design connections.</p></div>","PeriodicalId":380,"journal":{"name":"Journal of the Mechanical Behavior of Biomedical Materials","volume":"160 ","pages":"Article 106723"},"PeriodicalIF":3.3000,"publicationDate":"2024-09-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of the Mechanical Behavior of Biomedical Materials","FirstCategoryId":"5","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1751616124003552","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0
Abstract
Zygomatic implants (ZIs) were developed as a graftless alternative to rehabilitate severely reabsorbed maxillae. This study aims to employ three-dimensional finite element analysis (FEA) to simulate the impact of external hexagonal implant connection (EHC) and internal hexagonal implant connection (IHC) on the stress distribution and fatigue lifetime within the ZI systems using parameters defined in ISO 14801:2016. Two ZI assemblies (Nobel Biocare and Noris Medical) were scanned in a micro-CT scanner and reconstructed using Nrecon software. Three-dimensional models were generated by Simpleware ScanIP Medical software. All models were exported to FEA software (ABAQUS) and subsequently to a fatigue analysis software (Fe-safe). A compressive 150 N load was applied at a 40° angle on the cap surface. A 15 Hz frequency was applied in the in silico cyclic test. The implant components had material properties of commercially pure grade 4 titanium (CPTi) and Titanium-6Aluminum-4Vanadium alloy (Ti64). Von Mises stress data, contour plots, and fatigue limits were collected and analyzed. EHC models exhibited higher peak stresses in implant components for both materials compared to IHC models. However, simulated bone support results showed the opposite trend, with higher stresses on IHCthan EHC models. The fatigue analysis revealed that assemblies with both designs exceeded ISO 14801:2016 number of cycles limits using Ti64, while CPTi groups exhibited comparatively lower worst life-repeats. In conclusion, ZIs with IHC were found to have a more homogeneous and advantageous stress distribution within both materials tested. Ti64 demonstrates a prolonged service life for both design connections.
期刊介绍:
The Journal of the Mechanical Behavior of Biomedical Materials is concerned with the mechanical deformation, damage and failure under applied forces, of biological material (at the tissue, cellular and molecular levels) and of biomaterials, i.e. those materials which are designed to mimic or replace biological materials.
The primary focus of the journal is the synthesis of materials science, biology, and medical and dental science. Reports of fundamental scientific investigations are welcome, as are articles concerned with the practical application of materials in medical devices. Both experimental and theoretical work is of interest; theoretical papers will normally include comparison of predictions with experimental data, though we recognize that this may not always be appropriate. The journal also publishes technical notes concerned with emerging experimental or theoretical techniques, letters to the editor and, by invitation, review articles and papers describing existing techniques for the benefit of an interdisciplinary readership.