Is bicortical femoral pin insertion safe for image-based robotic knee arthroplasty surgery ? A comparative complications analysis in 970 consecutive cases
{"title":"Is bicortical femoral pin insertion safe for image-based robotic knee arthroplasty surgery ? A comparative complications analysis in 970 consecutive cases","authors":"Christos Koutserimpas , Clément Favroul , Cécile Batailler , Elvire Servien , Sébastien Lustig","doi":"10.1016/j.jisako.2024.100317","DOIUrl":null,"url":null,"abstract":"<div><h3>Objectives</h3><div>Limited data exists on complications associated with robotic image-based system in knee arthroplasty. This study aims to document complications in robotic arm-assisted knee arthroplasties and evaluate the system's safety by comparing two femoral pin insertion methods: bicortical diaphyseal with additional stab wounds, and unicortical metaphyseal placement through the main incision.</div></div><div><h3>Methods</h3><div>All patients undergoing primary knee arthroplasty with the image-based robotic system (Mako, Stryker, Mako Surgical Corp., Fort Lauderdale, FL, USA) from 1st March 2021 to 31st January 2024 with a minimum follow-up of 2 months were included. Demographics, system and non-system-related complications, as well as outcomes were recorded. Complications were categorized as either major (requiring a second surgical intervention) or minor.</div></div><div><h3>Results</h3><div>A total of 970 consecutive cases (median age 69.3 years) were analyzed. The unicortical group comprised 651 cases, while the bicortical group 319. The incidence of non-system-related complications was 2.37%, with the most common being joint stiffness (10 cases; 1.03%), followed by lateral femoral condyle fracture (4; 0.41%). The overall incidence of system-specific complications was 1.03%. Pin-related femoral fractures occurred in 0.2% of cases, all postoperatively and in the unicortical group. There was no statistically significant difference between the femoral pin insertion-related complication rates among the two groups (0.3% in the unicortical, compared to 0% in the bicortical group; p-value = 0.3). Complications included tibia fracture (0.1%), delayed wound healing (0.2%), superficial wound infection (0.1%), tibia osteomyelitis (0.1%), and “exostosis” (0.2%). The major complications rate was 0.3% and minor 0.7%.</div></div><div><h3>Conclusions</h3><div>Minimal system-specific overall complications indicate that robotic arm-assisted surgery is safe. The bicortical diaphyseal femoral pin insertion method does not increase the complication rates compared to the unicortical metaphyseal method.</div></div><div><h3>Level of evidence</h3><div>III.</div></div>","PeriodicalId":36847,"journal":{"name":"Journal of ISAKOS Joint Disorders & Orthopaedic Sports Medicine","volume":"10 ","pages":"Article 100317"},"PeriodicalIF":2.7000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of ISAKOS Joint Disorders & Orthopaedic Sports Medicine","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2059775424001640","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives
Limited data exists on complications associated with robotic image-based system in knee arthroplasty. This study aims to document complications in robotic arm-assisted knee arthroplasties and evaluate the system's safety by comparing two femoral pin insertion methods: bicortical diaphyseal with additional stab wounds, and unicortical metaphyseal placement through the main incision.
Methods
All patients undergoing primary knee arthroplasty with the image-based robotic system (Mako, Stryker, Mako Surgical Corp., Fort Lauderdale, FL, USA) from 1st March 2021 to 31st January 2024 with a minimum follow-up of 2 months were included. Demographics, system and non-system-related complications, as well as outcomes were recorded. Complications were categorized as either major (requiring a second surgical intervention) or minor.
Results
A total of 970 consecutive cases (median age 69.3 years) were analyzed. The unicortical group comprised 651 cases, while the bicortical group 319. The incidence of non-system-related complications was 2.37%, with the most common being joint stiffness (10 cases; 1.03%), followed by lateral femoral condyle fracture (4; 0.41%). The overall incidence of system-specific complications was 1.03%. Pin-related femoral fractures occurred in 0.2% of cases, all postoperatively and in the unicortical group. There was no statistically significant difference between the femoral pin insertion-related complication rates among the two groups (0.3% in the unicortical, compared to 0% in the bicortical group; p-value = 0.3). Complications included tibia fracture (0.1%), delayed wound healing (0.2%), superficial wound infection (0.1%), tibia osteomyelitis (0.1%), and “exostosis” (0.2%). The major complications rate was 0.3% and minor 0.7%.
Conclusions
Minimal system-specific overall complications indicate that robotic arm-assisted surgery is safe. The bicortical diaphyseal femoral pin insertion method does not increase the complication rates compared to the unicortical metaphyseal method.