Kristian Kjærgaard,Sibel Yilmaz,Bart Kaptein,Søren Overgaard,Ming Ding
{"title":"Measuring total hip arthroplasty liner wear using the EOS Imaging System: experimental and clinical results.","authors":"Kristian Kjærgaard,Sibel Yilmaz,Bart Kaptein,Søren Overgaard,Ming Ding","doi":"10.2340/17453674.2024.41912","DOIUrl":null,"url":null,"abstract":"BACKGROUND AND PURPOSE\r\nThe low-dose EOS Imaging System is an emerging tool for 3-dimensional measurements in orthopedics. The clinical feasibility for measuring total hip arthroplasty (THA) liner wear has not yet been investigated. We aimed to evaluate the feasibility of using EOS to measure THA liner wear by examining the experimental accuracy using a THA phantom and clinical precision of patients with THA, considering a clinically relevant precision at the 95% repeatability limit to be 0.2 mm.\r\n\r\nMETHODS\r\nAn experimental THA phantom with movable stem and a fixed cup with a plastic liner was constructed to simulate progressive 3D liner wear. Series of 11 pairs of radiographs with 50 μm femoral movement in between were obtained for each 3D axis in EOS. 30 patients with a THA were scanned twice using EOS to assess precision. Model-based radiostereometric analysis (RSA) was used for wear measurement.\r\n\r\nRESULTS\r\nThe mean difference (true minus simulated wear) with standard deviation (SD) and 95% limits of agreement for experimental THA wear were 0.005 (0.037) and [-0.069 to 0.079] mm for the vertical (y) axis. The mean (SD) and 95% repeatability limit for precision for clinical measurement were -0.029 (0.105) and 0.218 mm.\r\n\r\nCONCLUSION\r\nExperimental THA liner wear using EOS was within clinically relevant tolerances and without bias. The clinical precision was just outside our defined clinically relevant precision. Compared with conventional RSA, EOS is less accurate and precise but may still be of value for certain clinical applications, provided larger sample size or longer follow-up are available.","PeriodicalId":6916,"journal":{"name":"Acta Orthopaedica","volume":"39 1","pages":"530-535"},"PeriodicalIF":2.5000,"publicationDate":"2024-09-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta Orthopaedica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2340/17453674.2024.41912","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0
Abstract
BACKGROUND AND PURPOSE
The low-dose EOS Imaging System is an emerging tool for 3-dimensional measurements in orthopedics. The clinical feasibility for measuring total hip arthroplasty (THA) liner wear has not yet been investigated. We aimed to evaluate the feasibility of using EOS to measure THA liner wear by examining the experimental accuracy using a THA phantom and clinical precision of patients with THA, considering a clinically relevant precision at the 95% repeatability limit to be 0.2 mm.
METHODS
An experimental THA phantom with movable stem and a fixed cup with a plastic liner was constructed to simulate progressive 3D liner wear. Series of 11 pairs of radiographs with 50 μm femoral movement in between were obtained for each 3D axis in EOS. 30 patients with a THA were scanned twice using EOS to assess precision. Model-based radiostereometric analysis (RSA) was used for wear measurement.
RESULTS
The mean difference (true minus simulated wear) with standard deviation (SD) and 95% limits of agreement for experimental THA wear were 0.005 (0.037) and [-0.069 to 0.079] mm for the vertical (y) axis. The mean (SD) and 95% repeatability limit for precision for clinical measurement were -0.029 (0.105) and 0.218 mm.
CONCLUSION
Experimental THA liner wear using EOS was within clinically relevant tolerances and without bias. The clinical precision was just outside our defined clinically relevant precision. Compared with conventional RSA, EOS is less accurate and precise but may still be of value for certain clinical applications, provided larger sample size or longer follow-up are available.
期刊介绍:
Acta Orthopaedica (previously Acta Orthopaedica Scandinavica) presents original articles of basic research interest, as well as clinical studies in the field of orthopedics and related sub disciplines. Ever since the journal was founded in 1930, by a group of Scandinavian orthopedic surgeons, the journal has been published for an international audience. Acta Orthopaedica is owned by the Nordic Orthopaedic Federation and is the official publication of this federation.