Automated visual acuity estimation by optokinetic nystagmus using a stepped sweep stimulus.

IF 2.8 3区 医学 Q1 OPHTHALMOLOGY Ophthalmic and Physiological Optics Pub Date : 2024-11-01 Epub Date: 2024-09-11 DOI:10.1111/opo.13391
Jason Turuwhenua, Zaw LinTun, Mohammad Norouzifard, Misty Edmonds, Rebecca Findlay, Joanna Black, Benjamin Thompson
{"title":"Automated visual acuity estimation by optokinetic nystagmus using a stepped sweep stimulus.","authors":"Jason Turuwhenua, Zaw LinTun, Mohammad Norouzifard, Misty Edmonds, Rebecca Findlay, Joanna Black, Benjamin Thompson","doi":"10.1111/opo.13391","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To describe an automatic system for objective measurement of visual acuity (VA) using optokinetic nystagmus (OKN). This pilot study tested the system's sensitivity and specificity for detecting reduced VA in healthy adults by comparing VA-OKN to VA with an Early Treatment of Diabetic Retinopathy Study (ETDRS) chart (VA-ETDRS).</p><p><strong>Methods: </strong>Adult participants (age 30 ± 12 years) with either reduced VA (n = 11, VA-ETDRS > 0.20 logMAR) or normal VA (n = 12, VA-ETDRS ≤ 0.20 logMAR) completed monocular VA-OKN measurements in each eye. The VA-OKN stimulus was an array of drifting (5°/s) vanishing discs presented in descending/ascending size order (0.00-1.00 logMAR in 0.10 steps). The stimulus was stepped every 2 s, and 10 sweeps were shown per eye (five ascending and five descending). Eye-tracking data determined when OKN activity ceased (descending sweep) or began (ascending sweep), which was used to determine VA-OKN for each sweep. The estimates were averaged across sweeps to produce an automated VA-OKN. The automated sweeps were then provided in randomised order to a reviewer blinded to the VA-ETDRS findings who determined a final VA-OKN for an eye.</p><p><strong>Results: </strong>A single randomly selected eye from each observer was used for analysis. The sensitivity and specificity of VA-OKN using the same 0.20 logMAR threshold as VA-ETDRS was 100%. Comparisons between the VA-OKN and VA-ETDRS measures were made for participants in the reduced VA group. There was no significant difference between VA-OKN and VA-ETDRS (p = 0.55) and the two measures produced comparable values (r<sup>2</sup> = 0.84, 95% limits of agreement = 0.19 logMAR, intra-class correlation coefficient = 0.90 [95% CI:0.68-0.97]).</p><p><strong>Conclusions: </strong>Visual acuity using optokinetic nystagmus correctly identified a VA deficit in adults and for those with a VA deficit, VA-OKN was strongly correlated with the gold-standard clinical measure of VA. OKN is a promising method which has the potential for use in cognitively impaired adults and pre-verbal children.</p>","PeriodicalId":19522,"journal":{"name":"Ophthalmic and Physiological Optics","volume":" ","pages":"1500-1512"},"PeriodicalIF":2.8000,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Ophthalmic and Physiological Optics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1111/opo.13391","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/9/11 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To describe an automatic system for objective measurement of visual acuity (VA) using optokinetic nystagmus (OKN). This pilot study tested the system's sensitivity and specificity for detecting reduced VA in healthy adults by comparing VA-OKN to VA with an Early Treatment of Diabetic Retinopathy Study (ETDRS) chart (VA-ETDRS).

Methods: Adult participants (age 30 ± 12 years) with either reduced VA (n = 11, VA-ETDRS > 0.20 logMAR) or normal VA (n = 12, VA-ETDRS ≤ 0.20 logMAR) completed monocular VA-OKN measurements in each eye. The VA-OKN stimulus was an array of drifting (5°/s) vanishing discs presented in descending/ascending size order (0.00-1.00 logMAR in 0.10 steps). The stimulus was stepped every 2 s, and 10 sweeps were shown per eye (five ascending and five descending). Eye-tracking data determined when OKN activity ceased (descending sweep) or began (ascending sweep), which was used to determine VA-OKN for each sweep. The estimates were averaged across sweeps to produce an automated VA-OKN. The automated sweeps were then provided in randomised order to a reviewer blinded to the VA-ETDRS findings who determined a final VA-OKN for an eye.

Results: A single randomly selected eye from each observer was used for analysis. The sensitivity and specificity of VA-OKN using the same 0.20 logMAR threshold as VA-ETDRS was 100%. Comparisons between the VA-OKN and VA-ETDRS measures were made for participants in the reduced VA group. There was no significant difference between VA-OKN and VA-ETDRS (p = 0.55) and the two measures produced comparable values (r2 = 0.84, 95% limits of agreement = 0.19 logMAR, intra-class correlation coefficient = 0.90 [95% CI:0.68-0.97]).

Conclusions: Visual acuity using optokinetic nystagmus correctly identified a VA deficit in adults and for those with a VA deficit, VA-OKN was strongly correlated with the gold-standard clinical measure of VA. OKN is a promising method which has the potential for use in cognitively impaired adults and pre-verbal children.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
使用阶梯扫频刺激,通过光动眼震自动估算视力。
目的:介绍一种利用光动眼震(OKN)客观测量视力(VA)的自动系统。这项试点研究通过比较 VA-OKN 与糖尿病视网膜病变早期治疗研究 (ETDRS) 图表(VA-ETDRS)的视力,测试了该系统检测健康成年人视力下降的灵敏度和特异性:成年参与者(年龄 30 ± 12 岁)中,视力下降者(n = 11,VA-ETDRS > 0.20 logMAR)或视力正常者(n = 12,VA-ETDRS ≤ 0.20 logMAR)分别完成了单眼 VA-OKN 测量。VA-OKN 刺激是一个以大小递减/递增顺序(0.00-1.00 logMAR,0.10 步)呈现的漂移(5°/s)消失圆盘阵列。刺激每 2 秒阶跃一次,每只眼睛扫描 10 次(5 次上升,5 次下降)。眼球跟踪数据可确定 OKN 活动何时停止(下降扫频)或开始(上升扫频),从而确定每次扫频的 VA-OKN。对每次扫描的估计值进行平均,得出自动 VA-OKN。然后将自动扫查结果按随机顺序提供给对 VA-ETDRS 结果视而不见的审查员,由其确定一只眼的最终 VA-OKN:结果:每个观察者随机抽取一只眼进行分析。使用与 VA-ETDRS 相同的 0.20 logMAR 临界值,VA-OKN 的灵敏度和特异性均为 100%。VA-OKN 和 VA-ETDRS 测量值之间的比较针对的是视力减退组的参与者。VA-OKN 和 VA-ETDRS 之间没有明显差异(p = 0.55),两种测量方法得出的数值相当(r2 = 0.84,95% 的一致性界限 = 0.19 logMAR,类内相关系数 = 0.90 [95% CI:0.68-0.97]):使用光动力眼震能正确识别成人的视力缺陷,对于视力缺陷者,VA-OKN 与视力的金标准临床测量值密切相关。OKN 是一种很有前途的方法,有可能用于认知障碍的成人和语言前儿童。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
5.10
自引率
13.80%
发文量
135
审稿时长
6-12 weeks
期刊介绍: Ophthalmic & Physiological Optics, first published in 1925, is a leading international interdisciplinary journal that addresses basic and applied questions pertinent to contemporary research in vision science and optometry. OPO publishes original research papers, technical notes, reviews and letters and will interest researchers, educators and clinicians concerned with the development, use and restoration of vision.
期刊最新文献
Factors associated with myopia in the Portuguese child population: An epidemiological study. Mechanism of optical treatments for myopia: Are lenslets joining the DOTs? Uptake of eye care services in South India: Retrospective mapping of self-reported barriers using the Theoretical Domains Framework. One hundred years and counting. Extracting full information from OCT scans-signs of early age-related macular degeneration within inner retinal layers by local neighbourhood statistics. Part I: Methodology.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1