RayStation/GATE Monte Carlo simulation framework for verification of proton therapy based on the12N imaging.

IF 3.3 3区 医学 Q2 ENGINEERING, BIOMEDICAL Physics in medicine and biology Pub Date : 2024-09-27 DOI:10.1088/1361-6560/ad7d5c
Zahra Ahmadi Ganjeh, Brian Zapien-Campos, Erik Traneus, Stefan Both, Peter Dendooven
{"title":"RayStation/GATE Monte Carlo simulation framework for verification of proton therapy based on the<sup>12</sup>N imaging.","authors":"Zahra Ahmadi Ganjeh, Brian Zapien-Campos, Erik Traneus, Stefan Both, Peter Dendooven","doi":"10.1088/1361-6560/ad7d5c","DOIUrl":null,"url":null,"abstract":"<p><p><i>Objective</i>.<sup>12</sup>N, having a half-life of 11 ms, is a highly effective positron emitter that can potentially provide near real-time feedback in proton therapy. There is currently no framework for comparing and validating positron emission imaging of<sup>12</sup>N. This work describes the development and validation of a Monte Carlo (MC) framework to calculate the images of<sup>12</sup>N, as well as long-lived isotopes, originating from activation by protons.<i>Approach</i>. The available dual-panel Biograph mCT PET scanner was modeled in GATE and validated by comparing the simulated sensitivity map with the measured one. The distributions of<sup>12</sup>N and long-lived isotopes were calculated by RayStation and used as the input of GATE simulations. The RayStation/GATE combination was verified using proton beam irradiations of homogeneous phantoms. A 120 MeV pulsed pencil beam with 10<sup>8</sup>protons per pulse was used. Two-dimensional images were created from the GATE output and compared with the images based on the measurements and the 1D longitudinal projection of the full 2D image was used to calculate the<sup>12</sup>N activity range.<i>Main results</i>. The simulated sensitivity in the center of the FoV (5.44%) agrees well with the measured one (5.41%). The simulated and measured 2D sensitivity maps agree in good detail. The relative difference between the measured and simulated positron activity range for both<sup>12</sup>N and long-lived isotopes is less than 1%. The broadening of the<sup>12</sup>N images relative to those of the longer-lived isotopes can be understood in terms of the large positron range of<sup>12</sup>N.<i>Significance</i>. We developed and validated a MC framework based on RayStation/GATE to support the in-beam PET method for quality assurance of proton therapy. The inclusion of the very short-lived isotope<sup>12</sup>N makes the framework useful for developing near real-time verification. This represents a significant step towards translating<sup>12</sup>N real-time in vivo verification to the clinic.</p>","PeriodicalId":20185,"journal":{"name":"Physics in medicine and biology","volume":null,"pages":null},"PeriodicalIF":3.3000,"publicationDate":"2024-09-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physics in medicine and biology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1088/1361-6560/ad7d5c","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

Objective.12N, having a half-life of 11 ms, is a highly effective positron emitter that can potentially provide near real-time feedback in proton therapy. There is currently no framework for comparing and validating positron emission imaging of12N. This work describes the development and validation of a Monte Carlo (MC) framework to calculate the images of12N, as well as long-lived isotopes, originating from activation by protons.Approach. The available dual-panel Biograph mCT PET scanner was modeled in GATE and validated by comparing the simulated sensitivity map with the measured one. The distributions of12N and long-lived isotopes were calculated by RayStation and used as the input of GATE simulations. The RayStation/GATE combination was verified using proton beam irradiations of homogeneous phantoms. A 120 MeV pulsed pencil beam with 108protons per pulse was used. Two-dimensional images were created from the GATE output and compared with the images based on the measurements and the 1D longitudinal projection of the full 2D image was used to calculate the12N activity range.Main results. The simulated sensitivity in the center of the FoV (5.44%) agrees well with the measured one (5.41%). The simulated and measured 2D sensitivity maps agree in good detail. The relative difference between the measured and simulated positron activity range for both12N and long-lived isotopes is less than 1%. The broadening of the12N images relative to those of the longer-lived isotopes can be understood in terms of the large positron range of12N.Significance. We developed and validated a MC framework based on RayStation/GATE to support the in-beam PET method for quality assurance of proton therapy. The inclusion of the very short-lived isotope12N makes the framework useful for developing near real-time verification. This represents a significant step towards translating12N real-time in vivo verification to the clinic.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
RayStation/GATE 蒙特卡洛模拟框架,用于验证基于 12N 成像的质子疗法。
目的: 12N 的半衰期为 11 毫秒,是一种非常有前途的正电子发射体,可为质子治疗提供近乎实时的反馈。目前还没有对 12N 的正电子发射成像进行比较和验证的框架。这项工作介绍了蒙特卡洛框架的开发和验证情况,该框架用于计算质子活化产生的 12N 以及长寿命同位素的图像。12N 和长寿命同位素的分布由 RayStation 计算,并用作 GATE 模拟的输入。使用质子束照射均质模型对 RayStation/GATE 组合进行了验证。使用的是每脉冲 108 个质子的 120 MeV 脉冲铅笔束。根据 GATE 输出创建了二维图像,并与基于测量的图像进行了比较,完整二维图像的一维纵向投影被用来计算 12N 活动范围。模拟和测量的二维灵敏度图在细节上非常吻合。12N 和长寿命同位素的测量正电子活度范围与模拟正电子活度范围之间的相对差异都小于 1%。12N 图像相对于长寿命同位素图像的扩大可以从 12N 的大正电子范围来理解。由于新颖地加入了寿命极短的同位素 12N,该框架可用于开发近实时验证。这是将 12N 实时体内验证应用于临床的重要一步。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Physics in medicine and biology
Physics in medicine and biology 医学-工程:生物医学
CiteScore
6.50
自引率
14.30%
发文量
409
审稿时长
2 months
期刊介绍: The development and application of theoretical, computational and experimental physics to medicine, physiology and biology. Topics covered are: therapy physics (including ionizing and non-ionizing radiation); biomedical imaging (e.g. x-ray, magnetic resonance, ultrasound, optical and nuclear imaging); image-guided interventions; image reconstruction and analysis (including kinetic modelling); artificial intelligence in biomedical physics and analysis; nanoparticles in imaging and therapy; radiobiology; radiation protection and patient dose monitoring; radiation dosimetry
期刊最新文献
Understanding and modeling human-AI interaction of artificial intelligence tool in radiation oncology clinic using deep neural network: a feasibility study using three year prospective data. FLIP: a novel method for patient-specific dose quantification in circulating blood in large vessels during proton or photon external beam radiotherapy treatments. Noise & mottle suppression methods for cumulative Cherenkov images of radiation therapy delivery. Quantitative assessment of areal bone mineral density using multi-energy localizer radiographs from photon-counting detector CT. TMAA-net: tensor-domain multi-planal anti-aliasing network for sparse-view CT image reconstruction.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1