{"title":"An alternative colorimetric RT-LAMP assay for the rapid detection of SARS-CoV-2: development and validation in Thailand.","authors":"Wiphat Klayut, Thanee Wongchai, Nattagarn Chuenchom, Payu Bhakdeenuan, Supranee Bunchoo, Sopa Srisungngam, Sirilada Suphankong, Benjawan Phetsuksiri, Ballang Uppapong, Janisara Rudeeaneksin","doi":"10.3855/jidc.19442","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>COVID-19, an emerging infectious disease caused by SARS-CoV-2, continues to be a global public health threat. The development of a colorimetric reverse transcription loop-mediated isothermal amplification (RT-LAMP) can extend the availability of simple, reliable molecular tests for the rapid detection of COVID-19.</p><p><strong>Methodology: </strong>The RT-LAMP assay was developed using a new primer set targeting a portion of SARS-CoV-2 orf8. The method was validated at 63 ºC for 60 minutes with naked-eye visualization of the color change. The clinical performance was compared to a real-time reverse transcription-polymerase chain reaction (rtRT-PCR) using 273 RNA samples extracted from nasopharyngeal swab specimens.</p><p><strong>Results: </strong>The developed RT-LAMP was specific to SARS-CoV-2 with a limit of detection at 15 RNA copies per reaction. The assay demonstrated diagnostic accuracy, sensitivity, specificity, positive predictive value, and negative predictive value of 90.48% (95% CI: 86.36-93.68%), 87.00% (95% CI: 81.53-91.33%), 100% (95% CI: 95.07-100%), 100% (95% CI: not available), and 73.74% (95% CI: 66.22-80.07%), respectively, compared to the rtRT-PCR. The greatest sensitivity of 98.03% (95% CI: 94.34-99.59%) was demonstrated in samples with the cycle threshold (Ct) values < 30 cycles.</p><p><strong>Conclusions: </strong>The RT-LAMP method in this study showed good performance. The assay can increase the scope of laboratory testing for rapidly detecting SARS-CoV-2 in Thailand. Due to a decrease in COVID-19 cases, its application is beneficial when commercial alternatives are unavailable.</p>","PeriodicalId":49160,"journal":{"name":"Journal of Infection in Developing Countries","volume":"18 8","pages":"1196-1203"},"PeriodicalIF":1.4000,"publicationDate":"2024-08-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Infection in Developing Countries","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3855/jidc.19442","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"INFECTIOUS DISEASES","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: COVID-19, an emerging infectious disease caused by SARS-CoV-2, continues to be a global public health threat. The development of a colorimetric reverse transcription loop-mediated isothermal amplification (RT-LAMP) can extend the availability of simple, reliable molecular tests for the rapid detection of COVID-19.
Methodology: The RT-LAMP assay was developed using a new primer set targeting a portion of SARS-CoV-2 orf8. The method was validated at 63 ºC for 60 minutes with naked-eye visualization of the color change. The clinical performance was compared to a real-time reverse transcription-polymerase chain reaction (rtRT-PCR) using 273 RNA samples extracted from nasopharyngeal swab specimens.
Results: The developed RT-LAMP was specific to SARS-CoV-2 with a limit of detection at 15 RNA copies per reaction. The assay demonstrated diagnostic accuracy, sensitivity, specificity, positive predictive value, and negative predictive value of 90.48% (95% CI: 86.36-93.68%), 87.00% (95% CI: 81.53-91.33%), 100% (95% CI: 95.07-100%), 100% (95% CI: not available), and 73.74% (95% CI: 66.22-80.07%), respectively, compared to the rtRT-PCR. The greatest sensitivity of 98.03% (95% CI: 94.34-99.59%) was demonstrated in samples with the cycle threshold (Ct) values < 30 cycles.
Conclusions: The RT-LAMP method in this study showed good performance. The assay can increase the scope of laboratory testing for rapidly detecting SARS-CoV-2 in Thailand. Due to a decrease in COVID-19 cases, its application is beneficial when commercial alternatives are unavailable.