High predicted durability for the novel small-diameter OmniaSecure defibrillation lead

IF 5.6 2区 医学 Q1 CARDIAC & CARDIOVASCULAR SYSTEMS Heart rhythm Pub Date : 2025-02-01 DOI:10.1016/j.hrthm.2024.09.005
George H. Crossley III MD, FHRS, CCDS , Pamela K. Mason MD, FHRS , Bert Hansky MD , Paolo De Filippo MD , Maully J. Shah MBBS, FHRS, CCDS, CEPS-P , Francois Philippon MD , Darius Sholevar MD , Travis D. Richardson MD , Michael B. West MD , Jay Dinerman MD , James Dawson , Adam Himes , Leah Severseike , Amy E. Thompson , Prashanthan Sanders MBBS, PhD, FHRS
{"title":"High predicted durability for the novel small-diameter OmniaSecure defibrillation lead","authors":"George H. Crossley III MD, FHRS, CCDS ,&nbsp;Pamela K. Mason MD, FHRS ,&nbsp;Bert Hansky MD ,&nbsp;Paolo De Filippo MD ,&nbsp;Maully J. Shah MBBS, FHRS, CCDS, CEPS-P ,&nbsp;Francois Philippon MD ,&nbsp;Darius Sholevar MD ,&nbsp;Travis D. Richardson MD ,&nbsp;Michael B. West MD ,&nbsp;Jay Dinerman MD ,&nbsp;James Dawson ,&nbsp;Adam Himes ,&nbsp;Leah Severseike ,&nbsp;Amy E. Thompson ,&nbsp;Prashanthan Sanders MBBS, PhD, FHRS","doi":"10.1016/j.hrthm.2024.09.005","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Defibrillation leads remain the Achilles heel of implantable cardioverter-defibrillators. As patients with implantable cardioverter-defibrillators are living longer and battery longevity increases, more durable leads are needed. The LEADR trial evaluated the novel, lumenless, small-diameter, OmniaSecure defibrillation lead and demonstrated favorable safety and efficacy profile as well as zero fractures through 12.7 ± 4.8 months and remains in clinical follow-up. To augment the clinical trial, advanced cardiac lead reliability modeling was used to project long-term lead durability.</div></div><div><h3>Objective</h3><div>We aimed to project the 10-year fracture-free survival of the OmniaSecure defibrillation lead using reliability modeling.</div></div><div><h3>Methods</h3><div>The validated reliability model, which incorporates patient and bench test data, was used to project the 10-year fracture-free survival of the OmniaSecure lead. A subset of LEADR trial patients underwent biplane fluoroscopy imaging during cardiac and patient motion to evaluate the lead’s bending curvature in vivo. Bench tests then reproduced these use conditions with greater bending curvatures than observed in patients to exaggerate stress on the lead and to evaluate the lead fatigue strength.</div></div><div><h3>Results</h3><div>The reliability modeling projects a 98.2% fracture-free survival rate of the OmniaSecure lead at 10 years, including a 10-year fracture-free survival rate of 97.9% in adolescents, exceeding both the modeled and clinical 10-year performance of the highly reliable, larger diameter Sprint Quattro lead.</div></div><div><h3>Conclusion</h3><div>Consistent with early clinical trial experience, modeling projects highly durable 10-year performance of the OmniaSecure lead, including within the active adolescent pediatric population, which may uniquely benefit from a novel 4.7F defibrillation lead designed for reliability.</div></div><div><h3>ClinicalTrials.gov identifier</h3><div>NCT04863664</div></div>","PeriodicalId":12886,"journal":{"name":"Heart rhythm","volume":"22 2","pages":"Pages 302-310"},"PeriodicalIF":5.6000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Heart rhythm","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1547527124032958","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Defibrillation leads remain the Achilles heel of implantable cardioverter-defibrillators. As patients with implantable cardioverter-defibrillators are living longer and battery longevity increases, more durable leads are needed. The LEADR trial evaluated the novel, lumenless, small-diameter, OmniaSecure defibrillation lead and demonstrated favorable safety and efficacy profile as well as zero fractures through 12.7 ± 4.8 months and remains in clinical follow-up. To augment the clinical trial, advanced cardiac lead reliability modeling was used to project long-term lead durability.

Objective

We aimed to project the 10-year fracture-free survival of the OmniaSecure defibrillation lead using reliability modeling.

Methods

The validated reliability model, which incorporates patient and bench test data, was used to project the 10-year fracture-free survival of the OmniaSecure lead. A subset of LEADR trial patients underwent biplane fluoroscopy imaging during cardiac and patient motion to evaluate the lead’s bending curvature in vivo. Bench tests then reproduced these use conditions with greater bending curvatures than observed in patients to exaggerate stress on the lead and to evaluate the lead fatigue strength.

Results

The reliability modeling projects a 98.2% fracture-free survival rate of the OmniaSecure lead at 10 years, including a 10-year fracture-free survival rate of 97.9% in adolescents, exceeding both the modeled and clinical 10-year performance of the highly reliable, larger diameter Sprint Quattro lead.

Conclusion

Consistent with early clinical trial experience, modeling projects highly durable 10-year performance of the OmniaSecure lead, including within the active adolescent pediatric population, which may uniquely benefit from a novel 4.7F defibrillation lead designed for reliability.

ClinicalTrials.gov identifier

NCT04863664

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
新型小直径 OmniaSecure 除颤导线的预测耐用性很高。
背景:除颤导线仍然是植入式心律转复除颤器的致命弱点。随着植入式心律转复除颤器患者寿命的延长和电池寿命的增加,需要更耐用的导线。LEADR 试验对新型无腔小直径 OmniaSecure 除颤导线进行了评估,结果表明该导线具有良好的安全性和有效性,并且在 12.7 ± 4.8 个月的临床随访中没有发生骨折。为了加强临床试验,我们采用了先进的心脏导联可靠性建模来预测导联的长期耐用性:我们的目的是利用可靠性模型预测 OmniaSecure 除颤导联的 10 年无断裂存活率:方法:经过验证的可靠性模型结合了患者和工作台测试数据,用于预测 OmniaSecure 导联的 10 年无断裂存活率。一部分 LEADR 试验患者在心脏和患者运动过程中接受了双平面透视成像,以评估体内导联的弯曲曲率。然后,在台架试验中再现了这些使用条件,其弯曲曲率比在患者身上观察到的更大,以夸大导线上的应力,并评估导线的疲劳强度:结果:可靠性模型预测 OmniaSecure 导联 10 年无骨折存活率为 98.2%,其中青少年 10 年无骨折存活率为 97.9%,超过了可靠性高、直径更大的 Sprint Quattro 导联的模型和临床 10 年表现:结论:与早期临床试验经验一致,建模预测 OmniaSecure 导联的 10 年性能非常持久,包括在活跃的青少年儿童群体中,他们可能会独特地受益于专为可靠性设计的新型 4.7F 除颤导联:Gov 标识符:NCT04863664。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Heart rhythm
Heart rhythm 医学-心血管系统
CiteScore
10.50
自引率
5.50%
发文量
1465
审稿时长
24 days
期刊介绍: HeartRhythm, the official Journal of the Heart Rhythm Society and the Cardiac Electrophysiology Society, is a unique journal for fundamental discovery and clinical applicability. HeartRhythm integrates the entire cardiac electrophysiology (EP) community from basic and clinical academic researchers, private practitioners, engineers, allied professionals, industry, and trainees, all of whom are vital and interdependent members of our EP community. The Heart Rhythm Society is the international leader in science, education, and advocacy for cardiac arrhythmia professionals and patients, and the primary information resource on heart rhythm disorders. Its mission is to improve the care of patients by promoting research, education, and optimal health care policies and standards.
期刊最新文献
Evaluation of Non-Invasive Isochronal Late Activation Mapping in Scar Related VT with Electrocardiographic Imaging against Contact Mapping. Examining the effectiveness of psychosocial interventions for patients with a cardiac implantable electronic device: A systematic review and meta-analysis. In-hospital safety of cryoballoon and radiofrequency ablation in patients with atrial fibrillation - German-wide analysis of more than 300 000 procedures. It's Electrophysiology, Not Electrophotography: location and ablation of accessory pathways. Novel Pacing Maneuvers for Mapping the Upstream Insertions of Unidirectional Accessory Pathways.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1