{"title":"Collapse and Occlusion of Balloon-Expandable Covered Stent in the Right Common Iliac Arteries Due to External Stress.","authors":"Seimei Go, Taira Kobayashi, Taiichi Takasaki, Shinya Takahashi","doi":"10.3400/avd.cr.24-00038","DOIUrl":null,"url":null,"abstract":"<p><p>The patient previously underwent surgery for cervical cancer. She reported intermittent claudication, and computed tomography (CT) revealed total occlusion of the left iliac artery. We conducted endovascular treatment (EVT) using balloon-expandable covered stents (BECS). The patient underwent colostomy closure after EVT. The patient reported intermittent claudication beginning the day after the procedure. The CT showed a collapse of the distal side of the stent implanted in the right common iliac artery, accompanied by localized thrombo-occlusion. During open surgery, metal hooks typically affect the iliac artery; force transmission through the tissue may indirectly compress the iliac artery, leading to BECS failure.</p>","PeriodicalId":7995,"journal":{"name":"Annals of vascular diseases","volume":"17 3","pages":"313-316"},"PeriodicalIF":0.6000,"publicationDate":"2024-09-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11444825/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of vascular diseases","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3400/avd.cr.24-00038","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/7/6 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
引用次数: 0
Abstract
The patient previously underwent surgery for cervical cancer. She reported intermittent claudication, and computed tomography (CT) revealed total occlusion of the left iliac artery. We conducted endovascular treatment (EVT) using balloon-expandable covered stents (BECS). The patient underwent colostomy closure after EVT. The patient reported intermittent claudication beginning the day after the procedure. The CT showed a collapse of the distal side of the stent implanted in the right common iliac artery, accompanied by localized thrombo-occlusion. During open surgery, metal hooks typically affect the iliac artery; force transmission through the tissue may indirectly compress the iliac artery, leading to BECS failure.