{"title":"Comparison of three implant systems under preload loss: A finite element analysis validated by digital image correlation methods.","authors":"Annikaer Anniwaer, Zhengrong Yin, Jiakang Zhu, Chunxiao Jin, Aihemaiti Muhetaer, Cui Huang","doi":"10.2186/jpr.JPR_D_24_00064","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study evaluated the effects of screw preload loss on three implant systems, both in silico and in vitro.</p><p><strong>Methods: </strong>Three finite element analysis (FEA) models of implant restorations were created using bone-level (BL, 4.8×12 mm; BLX, 4.5×12 mm) and tissue-level (TL, 4.8×12 mm) implant systems. The screws in each group were subjected to preloads of 100 N and 200 N, with an additional 130 N load applied to the crown tops. An in vitro study of the principal strain was conducted using digital image correlation (DIC) under the same conditions as for the FEA models. The results were evaluated for von Mises stress, principal strain, and sensitivity index.</p><p><strong>Results: </strong>During loading, the highest stress levels were observed in the implants and screws. In the BL group, the screws experienced the highest von Mises stress at 466.04 MPa and 795.26 MPa in the 100 N and 200 N groups, respectively. The BLX group showed the highest von Mises stress at 439.33 MPa and 780.88 MPa in the implants in the 100 N and 200 N groups. Sensitivity analysis revealed that the screws and abutments in the TL group were significantly more affected by the preload changes.</p><p><strong>Conclusions: </strong>The abutment in the TL group was particularly sensitive to preload changes compared with those in the BL and BLX groups. Variations in the preload significantly affect the stress distribution in implants and screws. Maintaining screw preload stability under loading is crucial in clinical practice to prevent mechanical failure.</p>","PeriodicalId":16887,"journal":{"name":"Journal of prosthodontic research","volume":" ","pages":""},"PeriodicalIF":3.2000,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of prosthodontic research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2186/jpr.JPR_D_24_00064","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: This study evaluated the effects of screw preload loss on three implant systems, both in silico and in vitro.
Methods: Three finite element analysis (FEA) models of implant restorations were created using bone-level (BL, 4.8×12 mm; BLX, 4.5×12 mm) and tissue-level (TL, 4.8×12 mm) implant systems. The screws in each group were subjected to preloads of 100 N and 200 N, with an additional 130 N load applied to the crown tops. An in vitro study of the principal strain was conducted using digital image correlation (DIC) under the same conditions as for the FEA models. The results were evaluated for von Mises stress, principal strain, and sensitivity index.
Results: During loading, the highest stress levels were observed in the implants and screws. In the BL group, the screws experienced the highest von Mises stress at 466.04 MPa and 795.26 MPa in the 100 N and 200 N groups, respectively. The BLX group showed the highest von Mises stress at 439.33 MPa and 780.88 MPa in the implants in the 100 N and 200 N groups. Sensitivity analysis revealed that the screws and abutments in the TL group were significantly more affected by the preload changes.
Conclusions: The abutment in the TL group was particularly sensitive to preload changes compared with those in the BL and BLX groups. Variations in the preload significantly affect the stress distribution in implants and screws. Maintaining screw preload stability under loading is crucial in clinical practice to prevent mechanical failure.
期刊介绍:
Journal of Prosthodontic Research is published 4 times annually, in January, April, July, and October, under supervision by the Editorial Board of Japan Prosthodontic Society, which selects all materials submitted for publication.
Journal of Prosthodontic Research originated as an official journal of Japan Prosthodontic Society. It has recently developed a long-range plan to become the most prestigious Asian journal of dental research regarding all aspects of oral and occlusal rehabilitation, fixed/removable prosthodontics, oral implantology and applied oral biology and physiology. The Journal will cover all diagnostic and clinical management aspects necessary to reestablish subjective and objective harmonious oral aesthetics and function.
The most-targeted topics:
1) Clinical Epidemiology and Prosthodontics
2) Fixed/Removable Prosthodontics
3) Oral Implantology
4) Prosthodontics-Related Biosciences (Regenerative Medicine, Bone Biology, Mechanobiology, Microbiology/Immunology)
5) Oral Physiology and Biomechanics (Masticating and Swallowing Function, Parafunction, e.g., bruxism)
6) Orofacial Pain and Temporomandibular Disorders (TMDs)
7) Adhesive Dentistry / Dental Materials / Aesthetic Dentistry
8) Maxillofacial Prosthodontics and Dysphagia Rehabilitation
9) Digital Dentistry
Prosthodontic treatment may become necessary as a result of developmental or acquired disturbances in the orofacial region, of orofacial trauma, or of a variety of dental and oral diseases and orofacial pain conditions.
Reviews, Original articles, technical procedure and case reports can be submitted. Letters to the Editor commenting on papers or any aspect of Journal of Prosthodontic Research are welcomed.