A self-regulated expiratory flow device for mechanical ventilation: a bench study.

IF 2.8 Q2 CRITICAL CARE MEDICINE Intensive Care Medicine Experimental Pub Date : 2024-10-16 DOI:10.1186/s40635-024-00681-0
Lianye Yang, Ubbo F Wiersema, Shailesh Bihari, Roy Broughton, Andy Roberts, Nigel Kelley, Mark McEwen
{"title":"A self-regulated expiratory flow device for mechanical ventilation: a bench study.","authors":"Lianye Yang, Ubbo F Wiersema, Shailesh Bihari, Roy Broughton, Andy Roberts, Nigel Kelley, Mark McEwen","doi":"10.1186/s40635-024-00681-0","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Unregulated expiratory flow may contribute to ventilator-induced lung injury. The amount of energy dissipated into the lungs with tidal mechanical ventilation may be used to quantify potentially injurious ventilation. Previously reported devices for variable expiratory flow regulation (FLEX) require, either computer-controlled feedback, or an initial expiratory flow trigger. In this bench study we present a novel passive expiratory flow regulation device.</p><p><strong>Methods: </strong>The device was tested using a commercially available mechanical ventilator with a range of settings (tidal volume 420 ml and 630 ml, max. inspiratory flow rate 30 L/min and 50 L/min, respiratory rate 10 min<sup>-1</sup>, positive end-expiratory pressure 5 cmH<sub>2</sub>O), and a test lung with six different combinations of compliance and resistance settings. The effectiveness of the device was evaluated for reduction in peak expiratory flow, expiratory time, mean airway pressure, and the reduction of tidal dissipated energy (measured as the area within the airway pressure-volume loop).</p><p><strong>Results: </strong>Maximal and minimal reduction in peak expiratory flow was from 97.18 ± 0.41 L/min to 25.82 ± 0.07 L/min (p < 0.001), and from 44.11 ± 0.42 L/min to 26.30 ± 0.06 L/min, respectively. Maximal prolongation in expiratory time was recorded from 1.53 ± 0.06 s to 3.64 ± 0.21 s (p < 0.001). As a result of the extended expiration, the maximal decrease in I:E ratio was from 1:1.15 ± 0.03 to 1:2.45 ± 0.01 (p < 0.001). The greatest increase in mean airway pressure was from 10.04 ± 0.03 cmH<sub>2</sub>O to 17.33 ± 0.03 cmH<sub>2</sub>O. Dissipated energy was significantly reduced with the device under all test conditions (p < 0.001). The greatest reduction in dissipated energy was from 1.74 ± 0.00 J to 0.84 ± 0.00 J per breath. The least reduction in dissipated energy was from 0.30 ± 0.00 J to 0.16 ± 0.00 J per breath. The greatest and least percentage reduction in dissipated energy was 68% and 33%, respectively.</p><p><strong>Conclusions: </strong>The device bench tested in this study demonstrated a significant reduction in peak expiratory flow rate and dissipated energy, compared to ventilation with unregulated expiratory flow. Application of the device warrants further experimental and clinical evaluation.</p>","PeriodicalId":13750,"journal":{"name":"Intensive Care Medicine Experimental","volume":"12 1","pages":"92"},"PeriodicalIF":2.8000,"publicationDate":"2024-10-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11484996/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Intensive Care Medicine Experimental","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1186/s40635-024-00681-0","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CRITICAL CARE MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Unregulated expiratory flow may contribute to ventilator-induced lung injury. The amount of energy dissipated into the lungs with tidal mechanical ventilation may be used to quantify potentially injurious ventilation. Previously reported devices for variable expiratory flow regulation (FLEX) require, either computer-controlled feedback, or an initial expiratory flow trigger. In this bench study we present a novel passive expiratory flow regulation device.

Methods: The device was tested using a commercially available mechanical ventilator with a range of settings (tidal volume 420 ml and 630 ml, max. inspiratory flow rate 30 L/min and 50 L/min, respiratory rate 10 min-1, positive end-expiratory pressure 5 cmH2O), and a test lung with six different combinations of compliance and resistance settings. The effectiveness of the device was evaluated for reduction in peak expiratory flow, expiratory time, mean airway pressure, and the reduction of tidal dissipated energy (measured as the area within the airway pressure-volume loop).

Results: Maximal and minimal reduction in peak expiratory flow was from 97.18 ± 0.41 L/min to 25.82 ± 0.07 L/min (p < 0.001), and from 44.11 ± 0.42 L/min to 26.30 ± 0.06 L/min, respectively. Maximal prolongation in expiratory time was recorded from 1.53 ± 0.06 s to 3.64 ± 0.21 s (p < 0.001). As a result of the extended expiration, the maximal decrease in I:E ratio was from 1:1.15 ± 0.03 to 1:2.45 ± 0.01 (p < 0.001). The greatest increase in mean airway pressure was from 10.04 ± 0.03 cmH2O to 17.33 ± 0.03 cmH2O. Dissipated energy was significantly reduced with the device under all test conditions (p < 0.001). The greatest reduction in dissipated energy was from 1.74 ± 0.00 J to 0.84 ± 0.00 J per breath. The least reduction in dissipated energy was from 0.30 ± 0.00 J to 0.16 ± 0.00 J per breath. The greatest and least percentage reduction in dissipated energy was 68% and 33%, respectively.

Conclusions: The device bench tested in this study demonstrated a significant reduction in peak expiratory flow rate and dissipated energy, compared to ventilation with unregulated expiratory flow. Application of the device warrants further experimental and clinical evaluation.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
用于机械通气的自动调节呼气流量装置:一项实验研究。
简介不规则的呼气流量可能会导致呼吸机诱发的肺损伤。潮气量机械通气时耗散到肺部的能量可用来量化潜在的损伤性通气。之前报道的可变呼气流量调节(FLEX)装置需要计算机控制的反馈或初始呼气流量触发。在这项工作台研究中,我们介绍了一种新型被动呼气流量调节装置:方法:使用市售机械呼吸机对该装置进行了测试,设置范围包括潮气量 420 毫升和 630 毫升、最大吸气流速 30 升/分钟和 50 升/分钟、呼吸频率 10 分钟-1、呼气末正压 5 厘米水银柱,以及测试肺的顺应性和阻力设置的六种不同组合。根据呼气流量峰值、呼气时间、平均气道压力和潮气耗散能量减少量(以气道压力-容积环路内的面积测量)对该装置的效果进行了评估:峰值呼气流量的最大和最小降幅分别为 97.18 ± 0.41 升/分钟到 25.82 ± 0.07 升/分钟(p 2O 到 17.33 ± 0.03 cmH2O)。在所有测试条件下,该装置的耗散能量都明显降低(p 结论):与未调节呼气流量的通气方式相比,本研究中测试的装置可显著降低峰值呼气流量和耗散能量。该装置的应用需要进一步的实验和临床评估。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Intensive Care Medicine Experimental
Intensive Care Medicine Experimental CRITICAL CARE MEDICINE-
CiteScore
5.10
自引率
2.90%
发文量
48
审稿时长
13 weeks
期刊最新文献
Predictors of intradialytic hypotension in critically ill patients undergoing kidney replacement therapy: a systematic review. Is passive leg raising clinically useful in predicting intradialytic hypotension? Largely ignored-but pathogenetically significant: ambient temperature in rodent sepsis models. The development of a C5.0 machine learning model in a limited data set to predict early mortality in patients with ARDS undergoing an initial session of prone positioning. A new method to predict return of spontaneous circulation by peripheral intravenous analysis during cardiopulmonary resuscitation: a rat model pilot study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1