Signal-to-noise trade-offs between magnet diameter and shield-to-coil distance for cylindrical Halbach-based portable MRI systems for neuroimaging.

IF 2 4区 医学 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Magnetic Resonance Materials in Physics, Biology and Medicine Pub Date : 2025-02-01 Epub Date: 2024-10-17 DOI:10.1007/s10334-024-01210-6
Javad Parsa, Andrew Webb
{"title":"Signal-to-noise trade-offs between magnet diameter and shield-to-coil distance for cylindrical Halbach-based portable MRI systems for neuroimaging.","authors":"Javad Parsa, Andrew Webb","doi":"10.1007/s10334-024-01210-6","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To investigate the trade-off between magnet bore diameter and the distance between the conductive Faraday shield and RF head coil for low-field point-of-care neuroimaging systems.</p><p><strong>Methods: </strong>Electromagnetic simulations were performed for three different Faraday shield geometries and two commonly used RF coil designs (spiral and solenoid) to assess the effects of a close-fitting shield on the RF coil's transmit and receive efficiencies. Experimental measurements were performed to confirm the accuracy of the simulations. Parallel simulations were performed to assess the static magnet ( <math><msub><mi>B</mi> <mn>0</mn></msub> </math> ) field as a function of the magnet bore diameter. The obtainable SNR was then calculated as a function of these two related variables.</p><p><strong>Results: </strong>Simulations of the RF coil characteristics and <math><msubsup><mi>B</mi> <mrow><mn>1</mn></mrow> <mo>+</mo></msubsup> </math> transmit efficiencies agreed well with corresponding experimentally determined parameters. Overall, the RF coil transmit efficiency was, as expected, higher when the gap between the shield and coil increased. The calculated intrinsic SNR showed that maximum SNR would be obtained for a cylindrical shield of diameter 310 mm with an inner diameter of the magnet of 320 mm (assuming 10 mm for the gradient coils).</p><p><strong>Conclusion: </strong>This work presents an overview of the trade-offs in transmit efficiencies for RF coils used for POC MRI neuroimaging as a function of coil-to-shield distance and inner diameter of the Halbach magnet. Results show that there is a relatively shallow optimum between a magnet diameter of 290 and 330 mm, with values falling more than 10% if either smaller or larger magnets are used.</p>","PeriodicalId":18067,"journal":{"name":"Magnetic Resonance Materials in Physics, Biology and Medicine","volume":" ","pages":"97-105"},"PeriodicalIF":2.0000,"publicationDate":"2025-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Magnetic Resonance Materials in Physics, Biology and Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s10334-024-01210-6","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/10/17 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: To investigate the trade-off between magnet bore diameter and the distance between the conductive Faraday shield and RF head coil for low-field point-of-care neuroimaging systems.

Methods: Electromagnetic simulations were performed for three different Faraday shield geometries and two commonly used RF coil designs (spiral and solenoid) to assess the effects of a close-fitting shield on the RF coil's transmit and receive efficiencies. Experimental measurements were performed to confirm the accuracy of the simulations. Parallel simulations were performed to assess the static magnet ( B 0 ) field as a function of the magnet bore diameter. The obtainable SNR was then calculated as a function of these two related variables.

Results: Simulations of the RF coil characteristics and B 1 + transmit efficiencies agreed well with corresponding experimentally determined parameters. Overall, the RF coil transmit efficiency was, as expected, higher when the gap between the shield and coil increased. The calculated intrinsic SNR showed that maximum SNR would be obtained for a cylindrical shield of diameter 310 mm with an inner diameter of the magnet of 320 mm (assuming 10 mm for the gradient coils).

Conclusion: This work presents an overview of the trade-offs in transmit efficiencies for RF coils used for POC MRI neuroimaging as a function of coil-to-shield distance and inner diameter of the Halbach magnet. Results show that there is a relatively shallow optimum between a magnet diameter of 290 and 330 mm, with values falling more than 10% if either smaller or larger magnets are used.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
用于神经成像的基于哈尔巴赫技术的圆柱形便携式磁共振成像系统在磁体直径和屏蔽线圈间距之间的信噪比权衡。
目的研究低场强护理点神经成像系统中磁体孔直径与导电法拉第屏蔽罩和射频头线圈之间距离的权衡:针对三种不同的法拉第屏蔽几何形状和两种常用的射频线圈设计(螺旋线圈和电磁线圈)进行了电磁模拟,以评估紧密贴合的屏蔽对射频线圈发射和接收效率的影响。实验测量证实了模拟的准确性。同时还进行了模拟,以评估静态磁铁 ( B 0 ) 场与磁铁孔直径的函数关系。然后将可获得的信噪比作为这两个相关变量的函数进行计算:结果:射频线圈特性和 B 1 + 发射效率的模拟结果与实验确定的相应参数非常吻合。总体而言,当屏蔽罩和线圈之间的间隙增大时,射频线圈的发射效率会更高。计算的固有信噪比显示,直径为 310 毫米的圆柱形屏蔽罩和内径为 320 毫米的磁体(假设梯度线圈为 10 毫米)可获得最大信噪比:本研究综述了用于 POC MRI 神经成像的射频线圈的传输效率权衡与线圈到屏蔽罩的距离和哈尔巴赫磁体内径的函数关系。结果表明,在磁体直径为 290 毫米和 330 毫米之间存在一个相对较浅的最佳值,如果使用较小或较大的磁体,该值会下降 10%以上。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
4.60
自引率
0.00%
发文量
58
审稿时长
>12 weeks
期刊介绍: MAGMA is a multidisciplinary international journal devoted to the publication of articles on all aspects of magnetic resonance techniques and their applications in medicine and biology. MAGMA currently publishes research papers, reviews, letters to the editor, and commentaries, six times a year. The subject areas covered by MAGMA include: advances in materials, hardware and software in magnetic resonance technology, new developments and results in research and practical applications of magnetic resonance imaging and spectroscopy related to biology and medicine, study of animal models and intact cells using magnetic resonance, reports of clinical trials on humans and clinical validation of magnetic resonance protocols.
期刊最新文献
Deep learning for efficient reconstruction of highly accelerated 3D FLAIR MRI in neurological deficits. Brain tumor detection and segmentation using deep learning. Accelerating multi-coil MR image reconstruction using weak supervision. Concentric-object and equiangular-object methods to perform standardized regional analysis in renal mpMRI. Free-breathing qRF-MRF with pilot tone respiratory motion navigator for T1, T2, T2*, and off-resonance mapping of the human body at 3 T.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1