Jacob G Eide, William Mason, Hussein Mackie, Bernard Cook, Amrita Ray, Karam Asmaro, Adam Robin, Jack Rock, John R Craig
{"title":"Diagnostic Accuracy of Beta-2 Transferrin Gel Electrophoresis for Detecting Cerebrospinal Fluid Rhinorrhea.","authors":"Jacob G Eide, William Mason, Hussein Mackie, Bernard Cook, Amrita Ray, Karam Asmaro, Adam Robin, Jack Rock, John R Craig","doi":"10.1002/lary.31845","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Unilateral thin clear rhinorrhea (UTCR) may represent a variety of pathologies including cerebrospinal fluid (CSF) rhinorrhea. Beta-2 transferrin (B2Tf) gel electrophoresis (GE) has become the preferred testing modality due to reportedly high sensitivity (87%-100%) and specificity (71%-100%). However, there have been relatively few studies assessing its diagnostic accuracy. The purpose of this single-institution study was to determine the accuracy of B2Tf GE in detecting CSF rhinorrhea.</p><p><strong>Methods: </strong>A single-center retrospective review was conducted from 2016 and 2024 for all patients who presented with UTCR and underwent B2Tf GE. Institutional review board approval was obtained. The gold standard for diagnostic confirmation of true and false positives (TP, FP) as well as false negatives (FN) was endoscopic exploration. The gold standard for true negative (TN) was response to medical therapy.</p><p><strong>Results: </strong>A total of 105 patients underwent 149 B2Tf GE tests. 40 (38.1%) patients were diagnosed with CSF rhinorrhea. Of the 149 B2-Tf GE tests, there were 51 TPs, 72 TNs, 20 FPs, and 6 FNs yielding 89.5% sensitivity, 78.3% specificity, 71.8% positive predictive value, and 92.3% negative predictive value, respectively. Of the false results the most common causes for error were purulent sinusitis (n = 6, 23.1%), possible mucous contamination from nose-blowing during collection (n = 3, 11.5%), patient collection error (n = 3, 11.5%), and blood contamination (n = 1, 3.8%).</p><p><strong>Conclusion: </strong>Although these single-institutional data demonstrate test accuracy within ranges previously reported in the literature, they also demonstrate diagnostic limitations. Future studies should explore reasons for erroneous B2Tf GE results and how these may change clinical decision-making.</p><p><strong>Level of evidence: </strong>IV Laryngoscope, 2024.</p>","PeriodicalId":49921,"journal":{"name":"Laryngoscope","volume":" ","pages":""},"PeriodicalIF":2.2000,"publicationDate":"2024-10-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Laryngoscope","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/lary.31845","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"MEDICINE, RESEARCH & EXPERIMENTAL","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: Unilateral thin clear rhinorrhea (UTCR) may represent a variety of pathologies including cerebrospinal fluid (CSF) rhinorrhea. Beta-2 transferrin (B2Tf) gel electrophoresis (GE) has become the preferred testing modality due to reportedly high sensitivity (87%-100%) and specificity (71%-100%). However, there have been relatively few studies assessing its diagnostic accuracy. The purpose of this single-institution study was to determine the accuracy of B2Tf GE in detecting CSF rhinorrhea.
Methods: A single-center retrospective review was conducted from 2016 and 2024 for all patients who presented with UTCR and underwent B2Tf GE. Institutional review board approval was obtained. The gold standard for diagnostic confirmation of true and false positives (TP, FP) as well as false negatives (FN) was endoscopic exploration. The gold standard for true negative (TN) was response to medical therapy.
Results: A total of 105 patients underwent 149 B2Tf GE tests. 40 (38.1%) patients were diagnosed with CSF rhinorrhea. Of the 149 B2-Tf GE tests, there were 51 TPs, 72 TNs, 20 FPs, and 6 FNs yielding 89.5% sensitivity, 78.3% specificity, 71.8% positive predictive value, and 92.3% negative predictive value, respectively. Of the false results the most common causes for error were purulent sinusitis (n = 6, 23.1%), possible mucous contamination from nose-blowing during collection (n = 3, 11.5%), patient collection error (n = 3, 11.5%), and blood contamination (n = 1, 3.8%).
Conclusion: Although these single-institutional data demonstrate test accuracy within ranges previously reported in the literature, they also demonstrate diagnostic limitations. Future studies should explore reasons for erroneous B2Tf GE results and how these may change clinical decision-making.
期刊介绍:
The Laryngoscope has been the leading source of information on advances in the diagnosis and treatment of head and neck disorders since 1890. The Laryngoscope is the first choice among otolaryngologists for publication of their important findings and techniques. Each monthly issue of The Laryngoscope features peer-reviewed medical, clinical, and research contributions in general otolaryngology, allergy/rhinology, otology/neurotology, laryngology/bronchoesophagology, head and neck surgery, sleep medicine, pediatric otolaryngology, facial plastics and reconstructive surgery, oncology, and communicative disorders. Contributions include papers and posters presented at the Annual and Section Meetings of the Triological Society, as well as independent papers, "How I Do It", "Triological Best Practice" articles, and contemporary reviews. Theses authored by the Triological Society’s new Fellows as well as papers presented at meetings of the American Laryngological Association are published in The Laryngoscope.