{"title":"[Impact of Different Acquisition Modes on Image Quality and Quantitative Accuracy for Bone SPECT Using Ordered Subset Conjugate-gradient Minimizer].","authors":"Yui Orisaka, Takayuki Shibutani, Takahiro Konishi, Anji Yokouchi, Katsutoshi Fujiwara, Hiroto Yoneyama","doi":"10.6009/jjrt.2024-1455","DOIUrl":null,"url":null,"abstract":"<p><p>We evaluated the effect of different xSPECT Bone acquisition methods on image quality and quantitative accuracy. A mixed bone-equivalent solution set to 250 HU and <sup>99m</sup>Tc was filled with the SIM<sup>2</sup> Bone phantom. The xSPECT Bone acquisitions were performed with the step and shoot (SS) mode and two continuous modes (continuous<sub>view</sub>, continuous<sub>total</sub>) of the times per view, and the total acquisition times were the same as the SS mode. The radioactive concentration ratio of normal bone and tumor was set to 1 : 6. The reconstructed images were evaluated using uniformity, contrast, spatial resolution, recovery coefficient (RC) of the tumor part, and detectability. The detectability score (DS), an overall evaluation of detectability, was calculated from the results of uniformity and contrast. Uniformity of SS, continuous<sub>view</sub>, and continuous<sub>total</sub> methods was 29.4%, 17.4%, and 11.1%, respectively. Spatial resolution, RC, and detectability were similar among the SS mode and two continuous modes; however, the continuous modes showed good uniformity and contrast compared with the SS mode. In addition, the detectability of the continuous mode was equal to or superior to that of the SS mode.</p>","PeriodicalId":74309,"journal":{"name":"Nihon Hoshasen Gijutsu Gakkai zasshi","volume":" ","pages":"1191-1197"},"PeriodicalIF":0.0000,"publicationDate":"2024-11-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nihon Hoshasen Gijutsu Gakkai zasshi","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6009/jjrt.2024-1455","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/10/17 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
We evaluated the effect of different xSPECT Bone acquisition methods on image quality and quantitative accuracy. A mixed bone-equivalent solution set to 250 HU and 99mTc was filled with the SIM2 Bone phantom. The xSPECT Bone acquisitions were performed with the step and shoot (SS) mode and two continuous modes (continuousview, continuoustotal) of the times per view, and the total acquisition times were the same as the SS mode. The radioactive concentration ratio of normal bone and tumor was set to 1 : 6. The reconstructed images were evaluated using uniformity, contrast, spatial resolution, recovery coefficient (RC) of the tumor part, and detectability. The detectability score (DS), an overall evaluation of detectability, was calculated from the results of uniformity and contrast. Uniformity of SS, continuousview, and continuoustotal methods was 29.4%, 17.4%, and 11.1%, respectively. Spatial resolution, RC, and detectability were similar among the SS mode and two continuous modes; however, the continuous modes showed good uniformity and contrast compared with the SS mode. In addition, the detectability of the continuous mode was equal to or superior to that of the SS mode.
我们评估了不同的 xSPECT 骨采集方法对图像质量和定量准确性的影响。在 SIM2 骨模型中注入了设定为 250 HU 和 99mTc 的混合骨当量溶液。xSPECT Bone 采集采用步进拍摄(SS)模式和两种连续模式(continuousview、continuousoustotal),每个视图的采集时间和总采集时间与 SS 模式相同。正常骨骼和肿瘤的放射性浓度比设定为 1:6。用均匀度、对比度、空间分辨率、肿瘤部分的恢复系数(RC)和可探测性对重建图像进行评估。可探测性评分(DS)是对可探测性的总体评价,由均匀度和对比度的结果计算得出。SS、continuousview和continuousoustotal方法的均匀度分别为29.4%、17.4%和11.1%。SS 模式和两种连续模式的空间分辨率、RC 和可探测性相似;然而,与 SS 模式相比,连续模式显示出良好的均匀性和对比度。此外,连续模式的可探测性等于或优于 SS 模式。