Effective prism diopter for strabismus measurement: Conversion from anterior Prentice position to posterior parallel position and from glass prism to acrylic prism.

IF 2.1 4区 医学 Q2 OPHTHALMOLOGY Indian Journal of Ophthalmology Pub Date : 2024-10-25 DOI:10.4103/IJO.IJO_1113_24
Leilei Zou, Hong Liu, Shu Wang, Tian Tian, Cheng Fang, Gang Luo, Rui Liu
{"title":"Effective prism diopter for strabismus measurement: Conversion from anterior Prentice position to posterior parallel position and from glass prism to acrylic prism.","authors":"Leilei Zou, Hong Liu, Shu Wang, Tian Tian, Cheng Fang, Gang Luo, Rui Liu","doi":"10.4103/IJO.IJO_1113_24","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>In clinical ophthalmology, prisms, commonly made of glass, are labeled based on the prism diopter (PD) value measured in the anterior Prentice position. However, they are often used in the posterior parallel position for practicality, causing a discrepancy between labeled and effective PD. This study proposes a conversion tool to address this issue.</p><p><strong>Methods: </strong>Geometric optical analysis was employed to calculate the effective PD of glass prisms in the posterior parallel position. A lookup table was created to convert the labeled PD of the anterior Prentice position to the effective PD when the prism is used in the posterior parallel position. Clinical data from 162 patients with horizontal strabismus were collected to validate the method. Glass prism measurements, converted to effective PD, were compared with acrylic prism results.</p><p><strong>Results: </strong>The variations were significant in glass prism PD with rotation in the anterior Prentice position but stability in the posterior parallel position. Clinical deviations were larger with glass prisms, especially exceeding 30 PD, compared with acrylic prisms. Converted PD from glass prisms correlated well with acrylic prisms (R2 = 0.94, P = 0.002). The half-width of the 95% limit of agreement was ±12.32 PD.</p><p><strong>Conclusion: </strong>The proposed conversion table is valid, facilitating consistency between clinical and literature-based prism use in different positions and materials.</p>","PeriodicalId":13329,"journal":{"name":"Indian Journal of Ophthalmology","volume":null,"pages":null},"PeriodicalIF":2.1000,"publicationDate":"2024-10-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Indian Journal of Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.4103/IJO.IJO_1113_24","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: In clinical ophthalmology, prisms, commonly made of glass, are labeled based on the prism diopter (PD) value measured in the anterior Prentice position. However, they are often used in the posterior parallel position for practicality, causing a discrepancy between labeled and effective PD. This study proposes a conversion tool to address this issue.

Methods: Geometric optical analysis was employed to calculate the effective PD of glass prisms in the posterior parallel position. A lookup table was created to convert the labeled PD of the anterior Prentice position to the effective PD when the prism is used in the posterior parallel position. Clinical data from 162 patients with horizontal strabismus were collected to validate the method. Glass prism measurements, converted to effective PD, were compared with acrylic prism results.

Results: The variations were significant in glass prism PD with rotation in the anterior Prentice position but stability in the posterior parallel position. Clinical deviations were larger with glass prisms, especially exceeding 30 PD, compared with acrylic prisms. Converted PD from glass prisms correlated well with acrylic prisms (R2 = 0.94, P = 0.002). The half-width of the 95% limit of agreement was ±12.32 PD.

Conclusion: The proposed conversion table is valid, facilitating consistency between clinical and literature-based prism use in different positions and materials.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
用于斜视测量的有效棱镜屈光度:从普伦蒂斯前位转换到平行后位,从玻璃棱镜转换到丙烯酸棱镜。
目的:在临床眼科中,棱镜通常由玻璃制成,根据在前普伦蒂斯位置测量的棱镜屈光度 (PD) 值进行标注。然而,出于实用性考虑,它们通常在后平行位置上使用,这就造成了标注的 PD 值与有效 PD 值之间的差异。本研究提出了一种转换工具来解决这一问题:方法:采用几何光学分析法计算玻璃棱镜在后平行位置上的有效 PD。创建了一个查找表,用于将普伦蒂斯前位的标注 PD 转换为棱镜在后平行位使用时的有效 PD。收集了 162 名水平斜视患者的临床数据来验证该方法。将玻璃棱镜测量结果转换为有效PD,并与丙烯酸棱镜测量结果进行比较:结果:玻璃棱镜 PD 在普伦蒂斯前位旋转时变化显著,而在平行后位则保持稳定。与丙烯酸棱镜相比,玻璃棱镜的临床偏差更大,尤其是超过 30 PD。玻璃棱镜与丙烯酸棱镜的换算PD相关性良好(R2 = 0.94,P = 0.002)。95% 的一致性极限的半宽度为 ±12.32 PD:建议的换算表是有效的,有助于在不同位置和材料上使用棱镜的临床和文献依据的一致性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
3.80
自引率
19.40%
发文量
1963
审稿时长
38 weeks
期刊介绍: Indian Journal of Ophthalmology covers clinical, experimental, basic science research and translational research studies related to medical, ethical and social issues in field of ophthalmology and vision science. Articles with clinical interest and implications will be given preference.
期刊最新文献
Association between gut microbiota and central retinal artery occlusion: A two-sample Mendelian randomization study. Clinical outcomes of topography-guided versus wavefront-optimized LASIK for correction of myopia and compound myopic astigmatism. Accuracy of Intraocular lens power calculation in pediatric traumatic cataract. Alterations in the tear film and ocular surface in pediatric migraine patients. Guidelines for setting up low-vision and rehabilitation services in India.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1