Francesco Vitali, Marco Zuin, Paul Charles, Javier Jiménez-Díaz, Seth H Sheldon, Ana Paula Tagliari, Federico Migliore, Michele Malagù, Mathieu Montoy, Felipe Higuera Sobrino, Alex M Courtney, Adriano Nunes Kochi, Samir Fareh, Matteo Bertini
{"title":"Ultrasound-guided vs. fluoro-guided axillary venous access for cardiac implantable electronic devices: a patient-based meta-analysis.","authors":"Francesco Vitali, Marco Zuin, Paul Charles, Javier Jiménez-Díaz, Seth H Sheldon, Ana Paula Tagliari, Federico Migliore, Michele Malagù, Mathieu Montoy, Felipe Higuera Sobrino, Alex M Courtney, Adriano Nunes Kochi, Samir Fareh, Matteo Bertini","doi":"10.1093/europace/euae274","DOIUrl":null,"url":null,"abstract":"<p><strong>Aims: </strong>The use of ultrasound (US)-guided venous puncture for cardiac pacing/defibrillation lead placement may minimize the risk of periprocedural complications and radiation exposure. However, none of the published studies have been sufficiently powered to recommend this approach as the standard of care. We compare the safety and efficacy of ultrasound-guided axillary venous puncture (US-AVP) vs. fluoroscopy-guided access for cardiac implantable electronic devices (CIEDs) by performing an individual patient data meta-analysis based on previously published studies.</p><p><strong>Methods and results: </strong>We conducted a thorough literature search encompassing longitudinal investigations (five randomized and one prospective studies) reporting data on X-ray-guided and US-AVP for CIED procedures. The primary endpoint was to compare the safety of the two techniques. Secondary endpoints included the success rate of each technique, the necessity of switching to alternative methods, the time needed to obtain venous access, X-ray exposure, and the occurrence of periprocedural complications. Six longitudinal eligible studies were identified including 700 patients (mean age 74.9 ± 12.1 years, 68.4% males). The two approaches for venous cannulation showed a similar success rate. The use of an X-ray-guided approach significantly increased the risk of inadvertent arterial punctures (OR: 2.15, 95% CI: 2.10-2.21, P = 0.003), after adjustment for potential confounders. Conversely, a US-AVP approach reduces time to vascular access, radiation exposure, and the number of attempts to vascular access.</p><p><strong>Conclusion: </strong>The US-AVP enhances safety by reducing radiation exposure and time to vascular access while maintaining a low rate of major complications compared to the X-ray-guided approach.</p><p><strong>Clinical trial registration: </strong>PROSPERO identifier: CRD42024539623.</p>","PeriodicalId":11981,"journal":{"name":"Europace","volume":" ","pages":""},"PeriodicalIF":7.9000,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Europace","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/europace/euae274","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Aims: The use of ultrasound (US)-guided venous puncture for cardiac pacing/defibrillation lead placement may minimize the risk of periprocedural complications and radiation exposure. However, none of the published studies have been sufficiently powered to recommend this approach as the standard of care. We compare the safety and efficacy of ultrasound-guided axillary venous puncture (US-AVP) vs. fluoroscopy-guided access for cardiac implantable electronic devices (CIEDs) by performing an individual patient data meta-analysis based on previously published studies.
Methods and results: We conducted a thorough literature search encompassing longitudinal investigations (five randomized and one prospective studies) reporting data on X-ray-guided and US-AVP for CIED procedures. The primary endpoint was to compare the safety of the two techniques. Secondary endpoints included the success rate of each technique, the necessity of switching to alternative methods, the time needed to obtain venous access, X-ray exposure, and the occurrence of periprocedural complications. Six longitudinal eligible studies were identified including 700 patients (mean age 74.9 ± 12.1 years, 68.4% males). The two approaches for venous cannulation showed a similar success rate. The use of an X-ray-guided approach significantly increased the risk of inadvertent arterial punctures (OR: 2.15, 95% CI: 2.10-2.21, P = 0.003), after adjustment for potential confounders. Conversely, a US-AVP approach reduces time to vascular access, radiation exposure, and the number of attempts to vascular access.
Conclusion: The US-AVP enhances safety by reducing radiation exposure and time to vascular access while maintaining a low rate of major complications compared to the X-ray-guided approach.
期刊介绍:
EP - Europace - European Journal of Pacing, Arrhythmias and Cardiac Electrophysiology of the European Heart Rhythm Association of the European Society of Cardiology. The journal aims to provide an avenue of communication of top quality European and international original scientific work and reviews in the fields of Arrhythmias, Pacing and Cellular Electrophysiology. The Journal offers the reader a collection of contemporary original peer-reviewed papers, invited papers and editorial comments together with book reviews and correspondence.