{"title":"Technical Challenges and Morbidity Associated With Removal of an IlluminOss Implant From a Humeral Shaft: A Case Report.","authors":"Tony Huynh, Niloofar Dehghan","doi":"10.2106/JBJS.CC.24.00231","DOIUrl":null,"url":null,"abstract":"<p><strong>Case: </strong>A 77-year-old man experienced acute failure of fixation of his humeral shaft fracture after fixation with IlluminOss photodynamic system stabilization (Photodynamic Bone Stabilization System [PBSS]). Owing to the well-fixed IlluminOss PBSS implant to the humeral intramedullary canal, complete removal was deemed not indicated. Partial implant removal and revision open reduction internal fixation with a proximal humerus plate was performed. However, this was complicated by recalcitrant deep hardware infection that failed multiple surgical debridements, requiring eventual humeral osteotomy to remove the remainder of the infected IlluminOss PBSS implant.</p><p><strong>Conclusion: </strong>This is one of the first articles highlighting the technical challenges and morbidity associated with removal of an IlluminOss PBSS.</p>","PeriodicalId":14748,"journal":{"name":"JBJS case connector","volume":"14 4","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-10-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"JBJS case connector","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2106/JBJS.CC.24.00231","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/10/1 0:00:00","PubModel":"eCollection","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Case: A 77-year-old man experienced acute failure of fixation of his humeral shaft fracture after fixation with IlluminOss photodynamic system stabilization (Photodynamic Bone Stabilization System [PBSS]). Owing to the well-fixed IlluminOss PBSS implant to the humeral intramedullary canal, complete removal was deemed not indicated. Partial implant removal and revision open reduction internal fixation with a proximal humerus plate was performed. However, this was complicated by recalcitrant deep hardware infection that failed multiple surgical debridements, requiring eventual humeral osteotomy to remove the remainder of the infected IlluminOss PBSS implant.
Conclusion: This is one of the first articles highlighting the technical challenges and morbidity associated with removal of an IlluminOss PBSS.
期刊介绍:
JBJS Case Connector helps improve patient care by providing the medical community with a journal that harnesses technology to provide information tools for discovery and reporting of unusual musculoskeletal problems, findings, treatment, and outcomes. Co-edited by Thomas W. Bauer, MD, PhD, and Ronald W. Lindsey, MD, JBJS Case Connector assists orthopaedic surgeons in the search for precedents, connections, and trends in their efforts to improve patient care. Using this unique journal, surgeons can find the commonalities between cases, benefit from the experience of their peers, and filter case information by many important variables in order to provide the best possible care for orthopaedic patients. This cross-referenced online journal includes thousands of orthopaedic case reports. It compiles symptoms, conditions, and demographic details to empower surgeons to find cases similar to theirs. Surgeons can mine the database to reveal emerging trends and identify patterns, distinguishing between truly rare cases and repeated, related single instances of a larger problem. The JBJS Case Connector Image Quiz feature provides interactive quizzes based on images from content published by JBJS and includes a discussion area for further exploration of ideas and concepts. The JBJS Image Quiz app for iPad and iPhone is available in the App Store. Contributions to JBJS Case Connector are welcomed from anywhere in the world and are considered on their merits. Articles must be written in English and should be submitted as outlined in the Instructions to Authors. All authors must abide by the JBJS ethics policies and all submissions to JBJS Case Connector are covered by the JBJS embargo policy.