{"title":"A new portable negative pressure wound therapy device: a prospective study investigating clinical outcomes.","authors":"Victoria Rose, Nadine Hachah Haram, Sarah Gallala","doi":"10.12968/jowc.2024.0033","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Closed surgical incision sites at high risk of complications, and with exudate or leakage, are increasingly being managed with closed incision negative pressure wound therapy (ciNPWT) to reduce tissue stress and increase the force necessary to disrupt the incision. This study was undertaken to investigate the performance and safety of a canister-based, single-use NPWT (suNPWT) system when used on closed surgical incision sites.</p><p><strong>Method: </strong>The investigation was designed as a prospective, open, non-comparative, multicentre study aimed at confirming the safety and performance attributes of the suNPWT system when applied to low-to-moderately exuding closed surgical incisions. The primary performance measure was the wound remaining closed from baseline to the last follow-up visit on day 14. Secondary performance measures included: wound and periwound condition; wear time of the system; product consumption; adherence to therapy; and patients' pain progress. Details of adverse events were also collected.</p><p><strong>Results: </strong>Some 35 patients were recruited. The closed surgical incisions responded well to treatment with the tested suNPWT system. All wounds remained closed throughout the investigation. Consistent with other studies of ciNPWT reporting low infection rates, the current study observed either no or low exudation in 90.4% of wounds at the final visit, together with absence of surgical site infection. Pain severity levels were low, both at dressing change and during delivery of negative pressure. No serious adverse device events were reported.</p><p><strong>Conclusion: </strong>In this study, the suNPWT system supported the healing of closed surgical incisions with no safety concerns relating to its use.</p>","PeriodicalId":17590,"journal":{"name":"Journal of wound care","volume":"33 11","pages":"833-840"},"PeriodicalIF":1.5000,"publicationDate":"2024-11-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of wound care","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.12968/jowc.2024.0033","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"DERMATOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: Closed surgical incision sites at high risk of complications, and with exudate or leakage, are increasingly being managed with closed incision negative pressure wound therapy (ciNPWT) to reduce tissue stress and increase the force necessary to disrupt the incision. This study was undertaken to investigate the performance and safety of a canister-based, single-use NPWT (suNPWT) system when used on closed surgical incision sites.
Method: The investigation was designed as a prospective, open, non-comparative, multicentre study aimed at confirming the safety and performance attributes of the suNPWT system when applied to low-to-moderately exuding closed surgical incisions. The primary performance measure was the wound remaining closed from baseline to the last follow-up visit on day 14. Secondary performance measures included: wound and periwound condition; wear time of the system; product consumption; adherence to therapy; and patients' pain progress. Details of adverse events were also collected.
Results: Some 35 patients were recruited. The closed surgical incisions responded well to treatment with the tested suNPWT system. All wounds remained closed throughout the investigation. Consistent with other studies of ciNPWT reporting low infection rates, the current study observed either no or low exudation in 90.4% of wounds at the final visit, together with absence of surgical site infection. Pain severity levels were low, both at dressing change and during delivery of negative pressure. No serious adverse device events were reported.
Conclusion: In this study, the suNPWT system supported the healing of closed surgical incisions with no safety concerns relating to its use.
期刊介绍:
Journal of Wound Care (JWC) is the definitive wound-care journal and the leading source of up-to-date research and clinical information on everything related to tissue viability. The journal was first launched in 1992 and aimed at catering to the needs of the multidisciplinary team. Published monthly, the journal’s international audience includes nurses, doctors and researchers specialising in wound management and tissue viability, as well as generalists wishing to enhance their practice.
In addition to cutting edge and state-of-the-art research and practice articles, JWC also covers topics related to wound-care management, education and novel therapies, as well as JWC cases supplements, a supplement dedicated solely to case reports and case series in wound care. All articles are rigorously peer-reviewed by a panel of international experts, comprised of clinicians, nurses and researchers.
Specifically, JWC publishes:
High quality evidence on all aspects of wound care, including leg ulcers, pressure ulcers, the diabetic foot, burns, surgical wounds, wound infection and more
The latest developments and innovations in wound care through both preclinical and preliminary clinical trials of potential new treatments worldwide
In-depth prospective studies of new treatment applications, as well as high-level research evidence on existing treatments
Clinical case studies providing information on how to deal with complex wounds
Comprehensive literature reviews on current concepts and practice, including cost-effectiveness
Updates on the activities of wound care societies around the world.