Feasibility and safety of endomyocardial biopsy by lumenless pacing lead sheath–directed method during cardiac implantable electronic device implantation
Yang Ye MD, PhD, FHRS , Youyou Chen MD , Hao Jiang MD , Jiling Zeng MD , Xingchen Wang MD , Ying Yang MD, PhD , Xiang Lan Liu MD, PhD , XuLin Hong MD, PhD , Bei Wang MD , Ya Xun Sun MD, PhD , Dongwu Lai MD, PhD , Xi-qi Xu MD, PhD , Guo-sheng Fu MD, PhD
{"title":"Feasibility and safety of endomyocardial biopsy by lumenless pacing lead sheath–directed method during cardiac implantable electronic device implantation","authors":"Yang Ye MD, PhD, FHRS , Youyou Chen MD , Hao Jiang MD , Jiling Zeng MD , Xingchen Wang MD , Ying Yang MD, PhD , Xiang Lan Liu MD, PhD , XuLin Hong MD, PhD , Bei Wang MD , Ya Xun Sun MD, PhD , Dongwu Lai MD, PhD , Xi-qi Xu MD, PhD , Guo-sheng Fu MD, PhD","doi":"10.1016/j.hrthm.2024.11.018","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Patients with an indication for a cardiovascular implantable electronic device (CIED) are complicated with special cardiomyopathy or other unspecified cardiac abnormalities and may need endomyocardial biopsy (EMB). However, EMB by a bioptome is usually avoided to reduce the risk of lead displacement in the CIED periprocedural period.</div></div><div><h3>Objective</h3><div>We aimed to assess the safety and feasibility of a novel approach for transvenous right ventricular (RV) EMB using the lead sheath method (L-S-M) during CIED implantation and compared it with the traditional bioptome method (T-B-M).</div></div><div><h3>Methods</h3><div>A consecutive 80 patients referred for EMB were enrolled. In the L-S-M group, a sheath with a lumenless pacing lead was positioned toward the middle to apical interventricular septum under fluoroscopy (n = 60). The CIED implantation was performed through the same venous access site. In the T-B-M group, a bioptome was used (n = 20). The clinical characteristics, procedural details, success rate, and complications were evaluated.</div></div><div><h3>Results</h3><div>There were 380 RV EMB procedures performed with comparable numbers of samples per patient, 4.1 ± 0.8 in the L-S-M group and 3.8 ± 0.8 in the T-B-M group. In the L-S-M group, 7 (11.7%) patients experienced minor complications with 3 transient right bundle branch blocks, 2 transient atrioventricular blocks, and 2 regional minor device pocket hematomas. In the T-B-M group, 3 (15%) experienced 1 temporary pacing for transient atrioventricular block, 1 chest discomfort, and 1 regional hematoma. No cardiac tamponade was detected.</div></div><div><h3>Conclusion</h3><div>RV EMB by the innovative L-S-M is technically feasible and safe. It can yield valuable and early diagnostic insights for patients who are candidates for CIEDs.</div></div>","PeriodicalId":12886,"journal":{"name":"Heart rhythm","volume":"22 4","pages":"Pages 989-997"},"PeriodicalIF":5.7000,"publicationDate":"2025-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Heart rhythm","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1547527124035719","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/11/17 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Background
Patients with an indication for a cardiovascular implantable electronic device (CIED) are complicated with special cardiomyopathy or other unspecified cardiac abnormalities and may need endomyocardial biopsy (EMB). However, EMB by a bioptome is usually avoided to reduce the risk of lead displacement in the CIED periprocedural period.
Objective
We aimed to assess the safety and feasibility of a novel approach for transvenous right ventricular (RV) EMB using the lead sheath method (L-S-M) during CIED implantation and compared it with the traditional bioptome method (T-B-M).
Methods
A consecutive 80 patients referred for EMB were enrolled. In the L-S-M group, a sheath with a lumenless pacing lead was positioned toward the middle to apical interventricular septum under fluoroscopy (n = 60). The CIED implantation was performed through the same venous access site. In the T-B-M group, a bioptome was used (n = 20). The clinical characteristics, procedural details, success rate, and complications were evaluated.
Results
There were 380 RV EMB procedures performed with comparable numbers of samples per patient, 4.1 ± 0.8 in the L-S-M group and 3.8 ± 0.8 in the T-B-M group. In the L-S-M group, 7 (11.7%) patients experienced minor complications with 3 transient right bundle branch blocks, 2 transient atrioventricular blocks, and 2 regional minor device pocket hematomas. In the T-B-M group, 3 (15%) experienced 1 temporary pacing for transient atrioventricular block, 1 chest discomfort, and 1 regional hematoma. No cardiac tamponade was detected.
Conclusion
RV EMB by the innovative L-S-M is technically feasible and safe. It can yield valuable and early diagnostic insights for patients who are candidates for CIEDs.
期刊介绍:
HeartRhythm, the official Journal of the Heart Rhythm Society and the Cardiac Electrophysiology Society, is a unique journal for fundamental discovery and clinical applicability.
HeartRhythm integrates the entire cardiac electrophysiology (EP) community from basic and clinical academic researchers, private practitioners, engineers, allied professionals, industry, and trainees, all of whom are vital and interdependent members of our EP community.
The Heart Rhythm Society is the international leader in science, education, and advocacy for cardiac arrhythmia professionals and patients, and the primary information resource on heart rhythm disorders. Its mission is to improve the care of patients by promoting research, education, and optimal health care policies and standards.