Proof-of-principle of 3D-printed track-end detectors for dosimetry in proton therapy

IF 3.2 2区 医学 Q1 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Medical physics Pub Date : 2024-11-18 DOI:10.1002/mp.17515
Vicky Bietenbeck, Claus Maximilian Bäcker, Jörg Wulff, Beate Timmermann, Christian Bäumer
{"title":"Proof-of-principle of 3D-printed track-end detectors for dosimetry in proton therapy","authors":"Vicky Bietenbeck,&nbsp;Claus Maximilian Bäcker,&nbsp;Jörg Wulff,&nbsp;Beate Timmermann,&nbsp;Christian Bäumer","doi":"10.1002/mp.17515","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Dosimetric equipment in particle therapy (PT) is associated with high costs. There is a lack of versatile, tissue-equivalent detectors suitable for in-vivo dosimetry. Faraday-cup (FC) type detectors are sensitive to stopped protons, that is, to track-ends (TEs). They experience a renaissance in PT as they can cope with high dose rates. Owing to their simple functional principle, production of FC could benefit from the dynamic technological developments in additive manufacturing of sensors.</p>\n </section>\n \n <section>\n \n <h3> Purpose</h3>\n \n <p>To build FC-type detectors for PT by standard 3D-printing. This study seeks to build an integrating, single-channel (SC) FC for replacement of a traditional FC and a <span></span><math>\n <semantics>\n <mrow>\n <mn>2</mn>\n <mo>×</mo>\n <mn>2</mn>\n </mrow>\n <annotation>$2\\times 2$</annotation>\n </semantics></math> array of FC elements indicating the feasibility of a spatially resolving detector.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Samples of FCs were produced with a dual-extruder 3D-printer with polylactic-acid filaments, which contained graphite in the conductive parts of the detector. Production was optimized in terms of materials and printing temperature. Samples were characterized by electrical tests and non-destructive 3D x-ray imaging. Beam tests were conducted at a clinical PT machine.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Operational FC-type detectors for proton fields were printed. The detected charge of the SC FC corresponded qualitatively to the one of a traditional FC. A <span></span><math>\n <semantics>\n <mrow>\n <mn>2</mn>\n <mo>×</mo>\n <mn>2</mn>\n </mrow>\n <annotation>$2\\times 2$</annotation>\n </semantics></math> FC array was fabricated in a single run. There was a linear relationship between the response of the individual FC elements and the machine output.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>3D-printing is a viable method for producing low-cost, tissue-equivalent, FC-type detectors for PT. They could potentially be used as TE detectors in anthropomorphic phantoms.</p>\n </section>\n </div>","PeriodicalId":18384,"journal":{"name":"Medical physics","volume":"52 2","pages":"737-741"},"PeriodicalIF":3.2000,"publicationDate":"2024-11-18","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/mp.17515","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical physics","FirstCategoryId":"3","ListUrlMain":"https://aapm.onlinelibrary.wiley.com/doi/10.1002/mp.17515","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

Background

Dosimetric equipment in particle therapy (PT) is associated with high costs. There is a lack of versatile, tissue-equivalent detectors suitable for in-vivo dosimetry. Faraday-cup (FC) type detectors are sensitive to stopped protons, that is, to track-ends (TEs). They experience a renaissance in PT as they can cope with high dose rates. Owing to their simple functional principle, production of FC could benefit from the dynamic technological developments in additive manufacturing of sensors.

Purpose

To build FC-type detectors for PT by standard 3D-printing. This study seeks to build an integrating, single-channel (SC) FC for replacement of a traditional FC and a 2 × 2 $2\times 2$ array of FC elements indicating the feasibility of a spatially resolving detector.

Methods

Samples of FCs were produced with a dual-extruder 3D-printer with polylactic-acid filaments, which contained graphite in the conductive parts of the detector. Production was optimized in terms of materials and printing temperature. Samples were characterized by electrical tests and non-destructive 3D x-ray imaging. Beam tests were conducted at a clinical PT machine.

Results

Operational FC-type detectors for proton fields were printed. The detected charge of the SC FC corresponded qualitatively to the one of a traditional FC. A 2 × 2 $2\times 2$ FC array was fabricated in a single run. There was a linear relationship between the response of the individual FC elements and the machine output.

Conclusions

3D-printing is a viable method for producing low-cost, tissue-equivalent, FC-type detectors for PT. They could potentially be used as TE detectors in anthropomorphic phantoms.

Abstract Image

Abstract Image

Abstract Image

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
用于质子治疗剂量测定的 3D 打印轨迹端探测器的原理验证。
背景:粒子治疗(PT)的剂量测定设备成本高昂。目前缺乏适用于体内剂量测定的多功能组织等效探测器。法拉第杯(FC)型探测器对停止质子(即轨迹末端(TE))敏感。由于它们可以承受高剂量率,因此在 PT 领域得到了复兴。由于其功能原理简单,FC 型探测器的生产可受益于传感器快速成型技术的蓬勃发展。本研究旨在建造一个集成的单通道(SC)FC,用于替代传统的FC和一个2 × 2 (2次)FC元件阵列,表明空间分辨率探测器的可行性:FC样品是用双挤出机3D打印机用聚乳酸丝制作的,探测器的导电部分含有石墨。生产过程中对材料和打印温度进行了优化。样品通过电气测试和非破坏性 3D X 射线成像进行表征。在临床 PT 机上进行了光束测试:结果:打印出了用于质子场的可运行 FC 型探测器。SC FC 检测到的电荷与传统 FC 检测到的电荷基本一致。一次运行就制造出了一个 2 × 2 $2\times 2$ 的 FC 阵列。单个 FC 元件的响应与机器输出之间存在线性关系:三维打印是生产用于 PT 的低成本、组织等效的 FC 型探测器的可行方法。它们有可能被用作拟人模型中的 TE 探测器。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Medical physics
Medical physics 医学-核医学
CiteScore
6.80
自引率
15.80%
发文量
660
审稿时长
1.7 months
期刊介绍: Medical Physics publishes original, high impact physics, imaging science, and engineering research that advances patient diagnosis and therapy through contributions in 1) Basic science developments with high potential for clinical translation 2) Clinical applications of cutting edge engineering and physics innovations 3) Broadly applicable and innovative clinical physics developments Medical Physics is a journal of global scope and reach. By publishing in Medical Physics your research will reach an international, multidisciplinary audience including practicing medical physicists as well as physics- and engineering based translational scientists. We work closely with authors of promising articles to improve their quality.
期刊最新文献
Evaluating the feasibility of diffusion models for diagnosis of osteoporosis in women: A clinical diagnostic analysis of DWI, CTRW, and FROC diffusion models Iodine-enhanced x-ray phase-contrast CT for three-dimensional virtual histopathology evaluation of human cirrhosis A deep-learning model for one-shot transcranial ultrasound simulation and phase aberration correction A generalizable dose prediction model for automatic radiotherapy planning based on physics-informed priors and large-kernel convolutions Convolutional recurrent U-net for cardiac cine MRI reconstruction via effective spatio-temporal feature exploitation
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1