Feasibility study of modularized pin ridge filter implementation in proton FLASH planning for liver stereotactic ablative body radiotherapy.

IF 3.3 3区 医学 Q2 ENGINEERING, BIOMEDICAL Physics in medicine and biology Pub Date : 2024-11-21 DOI:10.1088/1361-6560/ad95d6
Chaoqiong Ma, Xiaofeng Yang, Jufri Setianegara, Yinan Wang, Yuan Gao, David S Yu, Pretesh Patel, Jun Zhou
{"title":"Feasibility study of modularized pin ridge filter implementation in proton FLASH planning for liver stereotactic ablative body radiotherapy.","authors":"Chaoqiong Ma, Xiaofeng Yang, Jufri Setianegara, Yinan Wang, Yuan Gao, David S Yu, Pretesh Patel, Jun Zhou","doi":"10.1088/1361-6560/ad95d6","DOIUrl":null,"url":null,"abstract":"<p><p>We previously developed a FLASH planning framework for streamlined pin-ridge-filter (pin-RF) design, demonstrating its feasibility for single-energy proton FLASH planning. In this study, we refined the pin-RF design for easy assembly using reusable modules, focusing on its application in liver stereotactic ablative body radiotherapy (SABR). This framework generates an intermediate intensity-modulated proton therapy (IMPT) plan and translates it into step widths and thicknesses of pin-RFs for a single-energy FLASH plan. Parameters like energy spacing, monitor unit limit, and spot quantity were adjusted during IMPT planning, resulting in pin-RFs assembled using predefined modules with widths from 1 to 6 mm, each with a water-equivalent-thickness of 5 mm. This approach was validated on three liver SABR cases. FLASH doses, quantified using the FLASH effectiveness model at 1 to 5 Gy thresholds, were compared to conventional IMPT (IMPT-CONV) doses to assess clinical benefits. The highest demand for 6 mm width modules, moderate for 2-4 mm, and minimal for 1- and 5-mm modules were shown across all cases. At lower dose thresholds, the two-beam case reduced indicators including liver V21Gy and skin Dmax by >19.4%, while the three-beam cases showed reductions ≤11.4%, indicating the need for higher fractional beam doses for an enhanced FLASH effect. Positive clinical benefits were seen only in the two-beam case at the 5 Gy threshold. At the 1 Gy threshold, the two-beam FLASH plan outperformed the IMPT-CONV plan, reducing dose indicators for all relevant normal tissues by up to 31.2%. In contrast, the three-beam cases showed negative clinical benefits, with skin Dmax and liver V21Gy increasing by up to 17.4% due to lower fractional beam doses and closer beam arrangements. This study evaluated the feasibility of modularizing streamlined pin-RFs in single-energy proton FLASH planning for liver SABR, offering guidance on optimal module composition and strategies to enhance FLASH planning.</p>","PeriodicalId":20185,"journal":{"name":"Physics in medicine and biology","volume":" ","pages":""},"PeriodicalIF":3.3000,"publicationDate":"2024-11-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physics in medicine and biology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1088/1361-6560/ad95d6","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

Abstract

We previously developed a FLASH planning framework for streamlined pin-ridge-filter (pin-RF) design, demonstrating its feasibility for single-energy proton FLASH planning. In this study, we refined the pin-RF design for easy assembly using reusable modules, focusing on its application in liver stereotactic ablative body radiotherapy (SABR). This framework generates an intermediate intensity-modulated proton therapy (IMPT) plan and translates it into step widths and thicknesses of pin-RFs for a single-energy FLASH plan. Parameters like energy spacing, monitor unit limit, and spot quantity were adjusted during IMPT planning, resulting in pin-RFs assembled using predefined modules with widths from 1 to 6 mm, each with a water-equivalent-thickness of 5 mm. This approach was validated on three liver SABR cases. FLASH doses, quantified using the FLASH effectiveness model at 1 to 5 Gy thresholds, were compared to conventional IMPT (IMPT-CONV) doses to assess clinical benefits. The highest demand for 6 mm width modules, moderate for 2-4 mm, and minimal for 1- and 5-mm modules were shown across all cases. At lower dose thresholds, the two-beam case reduced indicators including liver V21Gy and skin Dmax by >19.4%, while the three-beam cases showed reductions ≤11.4%, indicating the need for higher fractional beam doses for an enhanced FLASH effect. Positive clinical benefits were seen only in the two-beam case at the 5 Gy threshold. At the 1 Gy threshold, the two-beam FLASH plan outperformed the IMPT-CONV plan, reducing dose indicators for all relevant normal tissues by up to 31.2%. In contrast, the three-beam cases showed negative clinical benefits, with skin Dmax and liver V21Gy increasing by up to 17.4% due to lower fractional beam doses and closer beam arrangements. This study evaluated the feasibility of modularizing streamlined pin-RFs in single-energy proton FLASH planning for liver SABR, offering guidance on optimal module composition and strategies to enhance FLASH planning.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
在质子 FLASH 计划中实施模块化针脊滤波器进行肝脏立体定向消融体放射治疗的可行性研究。
我们之前开发了一种用于简化针脊过滤器(pin-RF)设计的 FLASH 计划框架,证明了其在单能量质子 FLASH 计划中的可行性。在本研究中,我们改进了针脊过滤器的设计,使其能够使用可重复使用的模块进行简易组装,重点关注其在肝脏立体定向消融体放射治疗(SABR)中的应用。该框架可生成中间强度调制质子治疗(IMPT)计划,并将其转化为单能量 FLASH 计划中针-RF 的阶梯宽度和厚度。在 IMPT 计划期间,对能量间隔、监控单元限制和光斑数量等参数进行调整,从而使用宽度为 1 到 6 毫米的预定义模块组装出针-RF,每个模块的水等效厚度为 5 毫米。这种方法在三个肝脏 SABR 病例中得到了验证。在 1 到 5 Gy 的阈值下,使用 FLASH 效能模型量化的 FLASH 剂量与传统的 IMPT(IMPT-CONV)剂量进行了比较,以评估临床效益。在所有病例中,6 毫米宽模块的需求量最大,2-4 毫米模块的需求量适中,1 毫米和 5 毫米模块的需求量最小。在较低的剂量阈值下,双光束病例的肝脏V21Gy和皮肤Dmax等指标降低了19.4%以上,而三光束病例则降低了≤11.4%,这表明需要更高的部分光束剂量来增强FLASH效果。只有在 5 Gy 临界值下的双光束病例才有积极的临床疗效。在 1 Gy 临界值时,双光束 FLASH 方案优于 IMPT-CONV 方案,可将所有相关正常组织的剂量指标降低 31.2%。与此相反,三光束病例显示出负面的临床效益,皮肤Dmax和肝脏V21Gy增加了17.4%,原因是部分光束剂量更低,光束排列更紧密。这项研究评估了在肝脏SABR的单能质子FLASH计划中将流线型针式RF模块化的可行性,为最佳模块组成和增强FLASH计划的策略提供了指导。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
Physics in medicine and biology
Physics in medicine and biology 医学-工程:生物医学
CiteScore
6.50
自引率
14.30%
发文量
409
审稿时长
2 months
期刊介绍: The development and application of theoretical, computational and experimental physics to medicine, physiology and biology. Topics covered are: therapy physics (including ionizing and non-ionizing radiation); biomedical imaging (e.g. x-ray, magnetic resonance, ultrasound, optical and nuclear imaging); image-guided interventions; image reconstruction and analysis (including kinetic modelling); artificial intelligence in biomedical physics and analysis; nanoparticles in imaging and therapy; radiobiology; radiation protection and patient dose monitoring; radiation dosimetry
期刊最新文献
Automated planning of curved needle channels in 3D printed patient-tailored applicators for cervical cancer brachytherapy. Comparison of contrast-enhanced ultrasound imaging (CEUS) and super-resolution ultrasound (SRU) for the quantification of ischaemia flow redistribution: a theoretical study. Novel frequency selective B1focusing passive Lenz resonators for substantial MRI signal-to-noise ratio amplification. On the microdosimetric characterisation of the radiation quality of a carbon-ion beam and the effect of the target volume thickness. Automated treatment planning with deep reinforcement learning for head-and-neck (HN) cancer intensity modulated radiation therapy (IMRT).
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1