{"title":"Investigation of very high radiation hardness of 21 \\(\\varvec{\\mu }\\)m silicon self-biased detectors","authors":"Andrzej J. Kordyasz, Monika Paluch-Ferszt, Zygmunt Szefliński, Katarzyna Z. Krutul-Bitowska, Michał Kowalczyk, Andrzej Bednarek, Paweł J. Napiorkowski, Ł.ukasz Kordyasz, Grzegorz Gawlik, Konrad Krzyżak, Michał Gajewski","doi":"10.1140/epja/s10050-024-01454-9","DOIUrl":null,"url":null,"abstract":"<div><p>The radiation damage of 21 <span>\\(\\upmu \\)</span>m thick self-biased epitaxial <span>\\(\\varDelta E\\)</span> detectors were tested as a function of fluence of 90 MeV <span>\\(^{14}\\)</span>N ions. Technology of production and technique of measurements of <span>\\(\\varDelta E\\)</span> detectors were described. A new technique of soldering contact to thin detector is shown. In the present work the 21 <span>\\(\\upmu \\)</span>m thick self-biased detectors marked as d4 and d5 show proper operation with the fluence about 4<span>\\(\\cdot \\)</span>10<span>\\(^{15}\\)</span> <span>\\(\\hbox {ions/cm}^{2}\\)</span> and the fluence about 8<span>\\(\\cdot \\)</span>10<span>\\(^{15}\\)</span> <span>\\(\\hbox {ions/cm}^{2}\\)</span>, respectively. The charge collection efficiency of thin d5 <span>\\(\\varDelta E\\)</span> detector was increased about double at fluence about 8<span>\\(\\cdot \\)</span>10<span>\\(^{15}\\)</span> <span>\\(\\hbox {ions/cm}^{2}\\)</span>. The charge collection efficiency of thin d4 <span>\\(\\varDelta E\\)</span> detector was increased about 35<span>\\(\\%\\)</span> at fluence about 4<span>\\(\\cdot \\)</span>10<span>\\(^{15}\\)</span> <span>\\(\\hbox {ions/cm}^{2}\\)</span> followed decrease about 70<span>\\(\\%\\)</span> of detector counting rate registration from fluence 9.1<span>\\(\\cdot \\)</span>10<span>\\(^{15}\\)</span> <span>\\(\\hbox {ions/cm}^{2}\\)</span> to fluence about 5<span>\\(\\cdot \\)</span>10<span>\\(^{16}\\)</span> <span>\\(\\hbox {ions/cm}^{2}\\)</span> due to partially removing of Al evaporated contact from detector surface as an effect of heavy ion irradiation. Increase of charge collection efficiency of thin self-biased detectors manufactured by the low-temperature technique was probably produced by increasing of build-in potential as an effect of activate carrier concentration of boron ions in epitaxial layer by heavy ion irradiation.</p></div>","PeriodicalId":786,"journal":{"name":"The European Physical Journal A","volume":"60 11","pages":""},"PeriodicalIF":2.6000,"publicationDate":"2024-11-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The European Physical Journal A","FirstCategoryId":"4","ListUrlMain":"https://link.springer.com/article/10.1140/epja/s10050-024-01454-9","RegionNum":3,"RegionCategory":"物理与天体物理","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"PHYSICS, NUCLEAR","Score":null,"Total":0}
引用次数: 0
Abstract
The radiation damage of 21 \(\upmu \)m thick self-biased epitaxial \(\varDelta E\) detectors were tested as a function of fluence of 90 MeV \(^{14}\)N ions. Technology of production and technique of measurements of \(\varDelta E\) detectors were described. A new technique of soldering contact to thin detector is shown. In the present work the 21 \(\upmu \)m thick self-biased detectors marked as d4 and d5 show proper operation with the fluence about 4\(\cdot \)10\(^{15}\)\(\hbox {ions/cm}^{2}\) and the fluence about 8\(\cdot \)10\(^{15}\)\(\hbox {ions/cm}^{2}\), respectively. The charge collection efficiency of thin d5 \(\varDelta E\) detector was increased about double at fluence about 8\(\cdot \)10\(^{15}\)\(\hbox {ions/cm}^{2}\). The charge collection efficiency of thin d4 \(\varDelta E\) detector was increased about 35\(\%\) at fluence about 4\(\cdot \)10\(^{15}\)\(\hbox {ions/cm}^{2}\) followed decrease about 70\(\%\) of detector counting rate registration from fluence 9.1\(\cdot \)10\(^{15}\)\(\hbox {ions/cm}^{2}\) to fluence about 5\(\cdot \)10\(^{16}\)\(\hbox {ions/cm}^{2}\) due to partially removing of Al evaporated contact from detector surface as an effect of heavy ion irradiation. Increase of charge collection efficiency of thin self-biased detectors manufactured by the low-temperature technique was probably produced by increasing of build-in potential as an effect of activate carrier concentration of boron ions in epitaxial layer by heavy ion irradiation.
期刊介绍:
Hadron Physics
Hadron Structure
Hadron Spectroscopy
Hadronic and Electroweak Interactions of Hadrons
Nonperturbative Approaches to QCD
Phenomenological Approaches to Hadron Physics
Nuclear and Quark Matter
Heavy-Ion Collisions
Phase Diagram of the Strong Interaction
Hard Probes
Quark-Gluon Plasma and Hadronic Matter
Relativistic Transport and Hydrodynamics
Compact Stars
Nuclear Physics
Nuclear Structure and Reactions
Few-Body Systems
Radioactive Beams
Electroweak Interactions
Nuclear Astrophysics
Article Categories
Letters (Open Access)
Regular Articles
New Tools and Techniques
Reviews.