Rafael Lima Rodrigues de Carvalho, Mariana Avendanha Victoriano, Camila Claúdia Campos, Paula Frizera Vassallo, Vandack Nobre, Flávia Falci Ercole
{"title":"Accuracy and precision of non-invasive thermometers compared with the pulmonary artery temperature: a cross-sectional study.","authors":"Rafael Lima Rodrigues de Carvalho, Mariana Avendanha Victoriano, Camila Claúdia Campos, Paula Frizera Vassallo, Vandack Nobre, Flávia Falci Ercole","doi":"10.1590/1516-3180.2023.0409.R1.05062024","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Temperature fluctuations are critical indicators of a patient's condition in intensive care units (ICUs). While invasive methods offer a more reliable measurement of core temperature, they carry greater risks of complications, limiting their use in most situations. This underscores the need for research evaluating the reliability of non-invasive temperature monitoring methods.</p><p><strong>Objectives: </strong>This study aimed to assess the accuracy and precision of four non-invasive temperature measurement techniques compared to pulmonary artery temperature, considered the gold standard.</p><p><strong>Design and setting: </strong>We conducted a cross-sectional clinical study with repeated measures in the ICUs at Hospital das Clínicas da Universidade Federal de Minas Gerais and Hospital Felício Rocho, Belo Horizonte, Brazil.</p><p><strong>Methods: </strong>All patients admitted with a pulmonary artery catheter were included. We simultaneously recorded temperatures from the pulmonary artery, axillary area, oral cavity, temporal artery, and tympanic membrane. Bland-Altman plots were employed to assess the agreement between the different temperature measurements.</p><p><strong>Results: </strong>A total of 48 patients participated, with a mean age of 54 years. Females comprised 66.67% of the sample. Compared to pulmonary artery temperature, the accuracy and precision (mean and standard deviation) of the non-invasive methods were: axillary (-0.42°C, 0.59°C), oral (-0.30°C, 0.37°C), tympanic membrane (-0.21°C, 0.44°C), and temporal artery (-0.25°C, 0.61°C). Notably, in patients with abnormal body temperature (non-normothermic), only oral and tympanic membrane methods maintained their accuracy and precision.</p><p><strong>Conclusions: </strong>The non-invasive thermometers evaluated in this study demonstrated acceptable accuracy and precision (within the clinically relevant threshold of 0.5°C) compared to pulmonary artery temperature. Among the non-invasive methods, the tympanic membrane measurement proved to be the most reliable, followed by the oral method.</p>","PeriodicalId":49574,"journal":{"name":"Sao Paulo Medical Journal","volume":"142 6","pages":"e2023409"},"PeriodicalIF":1.3000,"publicationDate":"2024-11-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Sao Paulo Medical Journal","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1590/1516-3180.2023.0409.R1.05062024","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"MEDICINE, GENERAL & INTERNAL","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Temperature fluctuations are critical indicators of a patient's condition in intensive care units (ICUs). While invasive methods offer a more reliable measurement of core temperature, they carry greater risks of complications, limiting their use in most situations. This underscores the need for research evaluating the reliability of non-invasive temperature monitoring methods.
Objectives: This study aimed to assess the accuracy and precision of four non-invasive temperature measurement techniques compared to pulmonary artery temperature, considered the gold standard.
Design and setting: We conducted a cross-sectional clinical study with repeated measures in the ICUs at Hospital das Clínicas da Universidade Federal de Minas Gerais and Hospital Felício Rocho, Belo Horizonte, Brazil.
Methods: All patients admitted with a pulmonary artery catheter were included. We simultaneously recorded temperatures from the pulmonary artery, axillary area, oral cavity, temporal artery, and tympanic membrane. Bland-Altman plots were employed to assess the agreement between the different temperature measurements.
Results: A total of 48 patients participated, with a mean age of 54 years. Females comprised 66.67% of the sample. Compared to pulmonary artery temperature, the accuracy and precision (mean and standard deviation) of the non-invasive methods were: axillary (-0.42°C, 0.59°C), oral (-0.30°C, 0.37°C), tympanic membrane (-0.21°C, 0.44°C), and temporal artery (-0.25°C, 0.61°C). Notably, in patients with abnormal body temperature (non-normothermic), only oral and tympanic membrane methods maintained their accuracy and precision.
Conclusions: The non-invasive thermometers evaluated in this study demonstrated acceptable accuracy and precision (within the clinically relevant threshold of 0.5°C) compared to pulmonary artery temperature. Among the non-invasive methods, the tympanic membrane measurement proved to be the most reliable, followed by the oral method.
期刊介绍:
Published bimonthly by the Associação Paulista de Medicina, the journal accepts articles in the fields of clinical health science (internal medicine, gynecology and obstetrics, mental health, surgery, pediatrics and public health). Articles will be accepted in the form of original articles (clinical trials, cohort, case-control, prevalence, incidence, accuracy and cost-effectiveness studies and systematic reviews with or without meta-analysis), narrative reviews of the literature, case reports, short communications and letters to the editor. Papers with a commercial objective will not be accepted.