How does intelligent noise reduction software influence the image quality in pelvic digital radiography; a phantom study.

E D Hussner, S Sundby, C B Outzen, J Jensen, A Tingberg, H Precht
{"title":"How does intelligent noise reduction software influence the image quality in pelvic digital radiography; a phantom study.","authors":"E D Hussner, S Sundby, C B Outzen, J Jensen, A Tingberg, H Precht","doi":"10.1016/j.jmir.2024.101814","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>This study aims to evaluate the effects of a novel noise reduction software (INR, Canon Europe, Amsterdam, NL) on image quality (IQ) in Digital Radiography (DR) pelvic phantom images.</p><p><strong>Methods: </strong>In total, 53 pelvic phantom images and 360 technical images of a Contrast Detail Radiography phantom (CDRAD) were collected, including 8 different exposure levels (between 0.8 and 40 mAs at 70 kV) and 6 intensities of INR (ranging from settings 0-10). The pelvic images were evaluated by three reporting radiographers using absolute Visual Grading Analysis (VGA). The CDRAD images were analyzed with a CDRAD computer software.</p><p><strong>Results: </strong>The VGA showed that the images with the INR software had higher IQ than the images with no INR. The observers gave a high VGA score to the images with INR of 3.2 mAs and higher. There was a tendency for an INR level of 5 or 7 to give the highest VGA scores. In addition, the CDRAD study showed a significant improvement in IQ with increasing INR levels at the lower exposure levels.</p><p><strong>Conclusion: </strong>An improvement in overall IQ was seen at lower exposure levels when the INR software was used both for the VGA and the CDRAD study. Clinical tests including patient images need to be performed before implementing INR in practice to verify accurate diagnostic performance.</p>","PeriodicalId":94092,"journal":{"name":"Journal of medical imaging and radiation sciences","volume":"56 2","pages":"101814"},"PeriodicalIF":0.0000,"publicationDate":"2024-12-06","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of medical imaging and radiation sciences","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1016/j.jmir.2024.101814","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: This study aims to evaluate the effects of a novel noise reduction software (INR, Canon Europe, Amsterdam, NL) on image quality (IQ) in Digital Radiography (DR) pelvic phantom images.

Methods: In total, 53 pelvic phantom images and 360 technical images of a Contrast Detail Radiography phantom (CDRAD) were collected, including 8 different exposure levels (between 0.8 and 40 mAs at 70 kV) and 6 intensities of INR (ranging from settings 0-10). The pelvic images were evaluated by three reporting radiographers using absolute Visual Grading Analysis (VGA). The CDRAD images were analyzed with a CDRAD computer software.

Results: The VGA showed that the images with the INR software had higher IQ than the images with no INR. The observers gave a high VGA score to the images with INR of 3.2 mAs and higher. There was a tendency for an INR level of 5 or 7 to give the highest VGA scores. In addition, the CDRAD study showed a significant improvement in IQ with increasing INR levels at the lower exposure levels.

Conclusion: An improvement in overall IQ was seen at lower exposure levels when the INR software was used both for the VGA and the CDRAD study. Clinical tests including patient images need to be performed before implementing INR in practice to verify accurate diagnostic performance.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
求助全文
约1分钟内获得全文 去求助
来源期刊
自引率
0.00%
发文量
0
期刊介绍:
期刊最新文献
How does intelligent noise reduction software influence the image quality in pelvic digital radiography; a phantom study. Time constraints and workload in the computed tomography department. Medical Image sharing: What do the public see when reviewing radiographs? A pilot study. Feasibility of low contrast volume and low injection flow rate in CT pulmonary angiography. Caffeine effect on myocardial perfusion scintigraphy.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1