Comparison of Ocular Biometry Measurements from Two Swept-Source OCT Devices: Eyestar 900 vs. Anterion.

IF 0.8 4区 医学 Q4 OPHTHALMOLOGY Klinische Monatsblatter fur Augenheilkunde Pub Date : 2024-12-10 DOI:10.1055/a-2463-4061
Julian V Kool van Langenberghe, Petra Schwarzer, Tobias Jahn, Christoph Tappeiner, Dimitrios Kyroudis, Eva Vrbka, David Goldblum
{"title":"Comparison of Ocular Biometry Measurements from Two Swept-Source OCT Devices: Eyestar 900 vs. Anterion.","authors":"Julian V Kool van Langenberghe, Petra Schwarzer, Tobias Jahn, Christoph Tappeiner, Dimitrios Kyroudis, Eva Vrbka, David Goldblum","doi":"10.1055/a-2463-4061","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Advanced swept-source optical coherence tomography (SS-OCT) devices are the current gold standard for the measurement of ocular biometric and keratometric parameters, which are essential for the calculation and selection of intraocular lenses (IOLs). This study compares the agreement of two SS-OCT devices, the Eyestar 900 (Haag-Streit, Köniz, Switzerland) and the Anterion (Heidelberg Engineering, Heidelberg, Germany).</p><p><strong>Materials and methods: </strong>All patients undergoing cataract surgery or seeking consultation for corneal abnormalities between January 2024 and May 2024 were eligible for inclusion. Both eyes were included in the final analysis.</p><p><strong>Results: </strong>A sample of 86 eyes from 43 patients was analysed. The mean differences (ES-AN) across all data between the Eyestar 900 (ES) and Anterion (AN) were as follows: anterior chamber depth (ACD) - 0.080 mm (ICC > 0.926), axial length (AL) 0.015 mm (ICC > 0.99), central corneal thickness (CCT) 0.914 µm (ICC > 0.921), corneal curvature along the flat meridian (K1) - 0.024 D (ICC > 0.904), and steep meridian (K2) - 0.210 D (ICC > 0.902), white-to-white corneal diameter (WTW) 0.215 mm (ICC > 0.81). There was excellent agreement for ACD, AL, CCT, K1, K2, and WTW. However, the corneal axis (AX) showed only moderate agreement (ICC > 0.389), with a mean difference of 5.97° and a notable standard deviation of 58.6°.</p><p><strong>Conclusions: </strong>The Eyestar 900 and Anterion demonstrated substantial agreement for all parameters except for the corneal meridian axis. It may be assumed that the inclusion of eyes with irregular corneal topographies may have contributed to this discrepancy.</p><p><strong>Summary: </strong>Overall, there was a high level of agreement between the Eyestar 900 and Anterion. It is reasonable to assume that the measurement outcomes provided by both devices are interchangeable for the calculation of target refractive outcomes for intraocular lenses, with no clinically relevant differences.</p>","PeriodicalId":17904,"journal":{"name":"Klinische Monatsblatter fur Augenheilkunde","volume":" ","pages":""},"PeriodicalIF":0.8000,"publicationDate":"2024-12-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Klinische Monatsblatter fur Augenheilkunde","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1055/a-2463-4061","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Advanced swept-source optical coherence tomography (SS-OCT) devices are the current gold standard for the measurement of ocular biometric and keratometric parameters, which are essential for the calculation and selection of intraocular lenses (IOLs). This study compares the agreement of two SS-OCT devices, the Eyestar 900 (Haag-Streit, Köniz, Switzerland) and the Anterion (Heidelberg Engineering, Heidelberg, Germany).

Materials and methods: All patients undergoing cataract surgery or seeking consultation for corneal abnormalities between January 2024 and May 2024 were eligible for inclusion. Both eyes were included in the final analysis.

Results: A sample of 86 eyes from 43 patients was analysed. The mean differences (ES-AN) across all data between the Eyestar 900 (ES) and Anterion (AN) were as follows: anterior chamber depth (ACD) - 0.080 mm (ICC > 0.926), axial length (AL) 0.015 mm (ICC > 0.99), central corneal thickness (CCT) 0.914 µm (ICC > 0.921), corneal curvature along the flat meridian (K1) - 0.024 D (ICC > 0.904), and steep meridian (K2) - 0.210 D (ICC > 0.902), white-to-white corneal diameter (WTW) 0.215 mm (ICC > 0.81). There was excellent agreement for ACD, AL, CCT, K1, K2, and WTW. However, the corneal axis (AX) showed only moderate agreement (ICC > 0.389), with a mean difference of 5.97° and a notable standard deviation of 58.6°.

Conclusions: The Eyestar 900 and Anterion demonstrated substantial agreement for all parameters except for the corneal meridian axis. It may be assumed that the inclusion of eyes with irregular corneal topographies may have contributed to this discrepancy.

Summary: Overall, there was a high level of agreement between the Eyestar 900 and Anterion. It is reasonable to assume that the measurement outcomes provided by both devices are interchangeable for the calculation of target refractive outcomes for intraocular lenses, with no clinically relevant differences.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
1.30
自引率
0.00%
发文量
235
审稿时长
4-8 weeks
期刊介绍: -Konzentriertes Fachwissen aus Klinik und Praxis: Die entscheidenden Ergebnisse der internationalen Forschung - für Sie auf den Punkt gebracht und kritisch kommentiert, Übersichtsarbeiten zu den maßgeblichen Themen der täglichen Praxis, Top informiert - breite klinische Berichterstattung. -CME-Punkte sammeln mit dem Refresher: Effiziente, CME-zertifizierte Fortbildung, mit dem Refresher, 3 CME-Punkte pro Ausgabe - bis zu 36 CME-Punkte im Jahr!. -Aktuelle Rubriken mit echtem Nutzwert: Kurzreferate zu den wichtigsten Artikeln internationaler Zeitschriften, Schwerpunktthema in jedem Heft: Ausführliche Übersichtsarbeiten zu den wichtigsten Themen der Ophthalmologie – so behalten Sie das gesamte Fach im Blick!, Originalien mit den neuesten Entwicklungen, Übersichten zu den relevanten Themen.
期刊最新文献
Histopathological Findings of Epithelial Invasion after Arcuate and Radial Keratotomies. Automated Measurement of Strabismus Angle Using a Commercial Virtual Reality Headset. Evaluation of Reproducibility of Measurements of Peripapillary Blood Flow with Laser Speckle Flowgraphy in Healthy and POAG Subjects. Standardized Assessment of Health-Related Quality of Life in Patients with Congenital Aniridia. Comparison of Ocular Biometry Measurements from Two Swept-Source OCT Devices: Eyestar 900 vs. Anterion.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1