Biomechanical Evaluation of a Femoral Implant for Hip Resurfacing Arthroplasty in Dogs: An Ex Vivo Study.

IF 1 2区 农林科学 Q3 VETERINARY SCIENCES Veterinary and Comparative Orthopaedics and Traumatology Pub Date : 2024-12-10 DOI:10.1055/s-0044-1800818
Daniel M G Perez Neto, Leonardo A L Muzzi, Fábio L Santos, Ruthnea A L Muzzi, Fernando Y K Kawamoto, Luciane R Mesquita, Bruno W Minto
{"title":"Biomechanical Evaluation of a Femoral Implant for Hip Resurfacing Arthroplasty in Dogs: An Ex Vivo Study.","authors":"Daniel M G Perez Neto, Leonardo A L Muzzi, Fábio L Santos, Ruthnea A L Muzzi, Fernando Y K Kawamoto, Luciane R Mesquita, Bruno W Minto","doi":"10.1055/s-0044-1800818","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong> The objective of this study was to assess the stability of the femoral component of a hip resurfacing arthroplasty prosthesis in canine cadaveric bone using a biomechanical test.</p><p><strong>Study design: </strong> Twenty adult dog cadavers were utilized, from which both femoral bones were extracted and preserved for experimentation. For each pair of femurs from the same animal, one femur underwent implantation of the hip resurfacing arthroplasty prosthesis, while the contralateral femur remained intact. All femoral specimens underwent flexo-compression biomechanical testing, with variables including maximum load (ML), load at collapse (LC), displacement at maximum load (DML), displacement at collapse (DC), and stiffness (k) being analysed. Subsequent to the biomechanical evaluation, the femoral specimens were radiographed to assess failure behaviour.</p><p><strong>Results: </strong> The prosthesis group showed a reduction of 22% (<i>p</i> ≤ 0.050) for the ML variable and 27% (<i>p</i> ≤ 0.05) for LC values. The prosthetic group supported a mean ML equivalent to six times the body weight of the animals whose bones were tested. No significant disparities were detected between the groups for the other analysed variables (DML, DC, and k). Similar failure patterns, such as fractures in the femoral neck region, were observed in specimens from both groups.</p><p><strong>Conclusion: </strong> The hip resurfacing arthroplasty prosthesis demonstrated biomechanical performance with no differences in DML, DC, and k but showed inferior performance in ML and LC when compared with intact femurs in the <i>ex vivo</i> setting.</p>","PeriodicalId":51204,"journal":{"name":"Veterinary and Comparative Orthopaedics and Traumatology","volume":" ","pages":""},"PeriodicalIF":1.0000,"publicationDate":"2024-12-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Veterinary and Comparative Orthopaedics and Traumatology","FirstCategoryId":"97","ListUrlMain":"https://doi.org/10.1055/s-0044-1800818","RegionNum":2,"RegionCategory":"农林科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"VETERINARY SCIENCES","Score":null,"Total":0}
引用次数: 0

Abstract

Objective:  The objective of this study was to assess the stability of the femoral component of a hip resurfacing arthroplasty prosthesis in canine cadaveric bone using a biomechanical test.

Study design:  Twenty adult dog cadavers were utilized, from which both femoral bones were extracted and preserved for experimentation. For each pair of femurs from the same animal, one femur underwent implantation of the hip resurfacing arthroplasty prosthesis, while the contralateral femur remained intact. All femoral specimens underwent flexo-compression biomechanical testing, with variables including maximum load (ML), load at collapse (LC), displacement at maximum load (DML), displacement at collapse (DC), and stiffness (k) being analysed. Subsequent to the biomechanical evaluation, the femoral specimens were radiographed to assess failure behaviour.

Results:  The prosthesis group showed a reduction of 22% (p ≤ 0.050) for the ML variable and 27% (p ≤ 0.05) for LC values. The prosthetic group supported a mean ML equivalent to six times the body weight of the animals whose bones were tested. No significant disparities were detected between the groups for the other analysed variables (DML, DC, and k). Similar failure patterns, such as fractures in the femoral neck region, were observed in specimens from both groups.

Conclusion:  The hip resurfacing arthroplasty prosthesis demonstrated biomechanical performance with no differences in DML, DC, and k but showed inferior performance in ML and LC when compared with intact femurs in the ex vivo setting.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
犬髋关节置换术中股骨植入物的生物力学评价:离体研究。
目的:本研究的目的是通过生物力学试验评估犬尸体骨髋关节置换术假体股骨部分的稳定性。研究设计:使用20具成年狗尸体,从中提取并保存两根股骨用于实验。对于同一只动物的每一对股骨,其中一根股骨植入髋关节表面置换假体,而对侧股骨保持完整。所有股骨标本均进行了屈曲压缩生物力学测试,分析了最大载荷(ML)、崩溃载荷(LC)、最大载荷位移(DML)、崩溃位移(DC)和刚度(k)等变量。在生物力学评估之后,对股骨标本进行x线摄影以评估失效行为。结果:假体组ML变量降低22% (p≤0.050),LC值降低27% (p≤0.05)。假体组支持的平均ML相当于接受骨骼测试的动物体重的六倍。其他分析变量(DML, DC和k)在两组之间没有发现显著差异。在两组标本中观察到类似的失败模式,例如股骨颈区域骨折。结论:髋关节置换术假体的生物力学性能在DML、DC和k方面没有差异,但在ML和LC方面与完整股骨相比在离体情况下表现较差。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
CiteScore
2.00
自引率
15.40%
发文量
49
审稿时长
18-36 weeks
期刊介绍: Veterinary and Comparative Orthopaedics and Traumatology (VCOT) is the most important single source for clinically relevant information in orthopaedics and neurosurgery available anywhere in the world today. It is unique in that it is truly comparative and there is an unrivalled mix of review articles and basic science amid the information that is immediately clinically relevant in veterinary surgery today.
期刊最新文献
Computed Tomographic Characteristics of Greyhound Central Tarsal Bone Fractures. Patient-specific Guides Improve the Accuracy and Safety of Transcondylar Screw Placement-a Cadaveric Study in the Canine Humerus. Clinical and Radiographic Outcomes of 1.5-mm Locking Plate Fixation for 30 Radial and Ulnar Fractures in Dogs. Feline Shoulder Arthrodesis Using 3D-printed Patient-specific Guides. Arthroscopic Removal of Traumatic Fractures of the Proximal Medial Trochlear Ridge of the Talus: A Retrospective Analysis of 18 Horses.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1