Pascal Schipper, Constance Weber, Ke Lu, Siqi Fan, Verena Prokosch, Frank G Holz, Karl Mercieca
{"title":"Anterior segment OCT for imaging PAUL<sup>®</sup> glaucoma implant patch grafts: a useful method for follow-up and risk management.","authors":"Pascal Schipper, Constance Weber, Ke Lu, Siqi Fan, Verena Prokosch, Frank G Holz, Karl Mercieca","doi":"10.1007/s00417-024-06708-2","DOIUrl":null,"url":null,"abstract":"<p><strong>Aim: </strong>To evaluate a useful, non-contact method for the follow-up of pericardium patch graft changes in patients undergoing PAUL<sup>®</sup> Glaucoma Implant (PGI) surgery using high-resolution anterior segment optical coherence tomography (OCT) to predict tube erosions.</p><p><strong>Methods: </strong>Prospective analysis over six months of tube pericardium patch graft thickness of PGI surgical cases at the University Eye Hospital Bonn, Germany, from November 2021 to August 2022. In all eyes, Tutopatch<sup>®</sup> (RTI Surgical, United States) pericardium was used to cover the implant intra-operatively. Anterior segment OCT (AS-OCT, Heidelberg ANTERION<sup>®</sup> Swept-Source-OCT) examinations were performed following a standardized protocol to measure quantitative and qualitative aspects of the patch grafts before surgery, and at three and six months after surgery.</p><p><strong>Results: </strong>Twenty-six eyes of 26 patients were included. Thickness of the patch material was 1188 µm (IQR 415 µm) directly after implantation and decreased over time to 1068 µm (IQR 478 µm) at 3 months and 846 µm (IQR 677 µm) at 6 months. No significant differences between groups were shown concerning gender (p = 0.128), ethnicity (p = 1.000), age (p = 0.741), glaucoma type (p = 0.173), other concurrent diseases (p = 0.302), former glaucoma surgeries (p = 1.000) and the quadrant of implantation (p = 0.555). Five eyes developed implant exposure. When comparing eyes with and without tube exposure, no significant differences were shown in average patch thickness above the tube directly after implantation (p = 0.476). However, significant differences in average thickness were observed at 3 months (p = 0.013) and 6 months (p = 0.005).</p><p><strong>Conclusions: </strong>Pericardial patch grafts tend to thin over time which can be assessed by AS-OCT, the latter proving to be a useful method to follow-up patients who undergo patch graft implantation during PGI surgery. This investigation could potentially help identify patients at risk of tube exposure which in turn could lead to modification of patient management. It could also possibly be used in future studies to find more suitable patch materials.</p>","PeriodicalId":12795,"journal":{"name":"Graefe’s Archive for Clinical and Experimental Ophthalmology","volume":" ","pages":""},"PeriodicalIF":2.4000,"publicationDate":"2024-12-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Graefe’s Archive for Clinical and Experimental Ophthalmology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00417-024-06708-2","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Aim: To evaluate a useful, non-contact method for the follow-up of pericardium patch graft changes in patients undergoing PAUL® Glaucoma Implant (PGI) surgery using high-resolution anterior segment optical coherence tomography (OCT) to predict tube erosions.
Methods: Prospective analysis over six months of tube pericardium patch graft thickness of PGI surgical cases at the University Eye Hospital Bonn, Germany, from November 2021 to August 2022. In all eyes, Tutopatch® (RTI Surgical, United States) pericardium was used to cover the implant intra-operatively. Anterior segment OCT (AS-OCT, Heidelberg ANTERION® Swept-Source-OCT) examinations were performed following a standardized protocol to measure quantitative and qualitative aspects of the patch grafts before surgery, and at three and six months after surgery.
Results: Twenty-six eyes of 26 patients were included. Thickness of the patch material was 1188 µm (IQR 415 µm) directly after implantation and decreased over time to 1068 µm (IQR 478 µm) at 3 months and 846 µm (IQR 677 µm) at 6 months. No significant differences between groups were shown concerning gender (p = 0.128), ethnicity (p = 1.000), age (p = 0.741), glaucoma type (p = 0.173), other concurrent diseases (p = 0.302), former glaucoma surgeries (p = 1.000) and the quadrant of implantation (p = 0.555). Five eyes developed implant exposure. When comparing eyes with and without tube exposure, no significant differences were shown in average patch thickness above the tube directly after implantation (p = 0.476). However, significant differences in average thickness were observed at 3 months (p = 0.013) and 6 months (p = 0.005).
Conclusions: Pericardial patch grafts tend to thin over time which can be assessed by AS-OCT, the latter proving to be a useful method to follow-up patients who undergo patch graft implantation during PGI surgery. This investigation could potentially help identify patients at risk of tube exposure which in turn could lead to modification of patient management. It could also possibly be used in future studies to find more suitable patch materials.
期刊介绍:
Graefe''s Archive for Clinical and Experimental Ophthalmology is a distinguished international journal that presents original clinical reports and clini-cally relevant experimental studies. Founded in 1854 by Albrecht von Graefe to serve as a source of useful clinical information and a stimulus for discussion, the journal has published articles by leading ophthalmologists and vision research scientists for more than a century. With peer review by an international Editorial Board and prompt English-language publication, Graefe''s Archive provides rapid dissemination of clinical and clinically related experimental information.