Accuracy of acetabular cup positioning in robotic-assisted total hip arthroplasty: a CT-based evaluation.

IF 1.8 Q2 ORTHOPEDICS SICOT-J Pub Date : 2024-01-01 Epub Date: 2024-12-20 DOI:10.1051/sicotj/2024057
Ashish Singh, Purushotam Kumar, Kanukuntla Kalyan, Akash Chandrashekar Gundalli, Sudhir Shankar Mane, Himanshu Swarnkar, Lavanya Singh
{"title":"Accuracy of acetabular cup positioning in robotic-assisted total hip arthroplasty: a CT-based evaluation.","authors":"Ashish Singh, Purushotam Kumar, Kanukuntla Kalyan, Akash Chandrashekar Gundalli, Sudhir Shankar Mane, Himanshu Swarnkar, Lavanya Singh","doi":"10.1051/sicotj/2024057","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Robot-assisted instrumentation during total hip arthroplasty (THA) has the potential to improve acetabular cup positioning. This study aimed to evaluate the precision of robotic-arm-assisted THA (rTHA) and assess whether the system can provide accurate cup positioning comparable to conventional THA (cTHA).</p><p><strong>Methods: </strong>A single-center prospective cohort study consisting of 151 patients who underwent THA (108 rTHA and 43 cTHA). The robotically assisted system was used to match the postoperative computed tomography (CT) image of the pelvis with the planned and intraoperative anatomical landmarks. The cTHA cohort underwent hip replacement using the standard manual procedure, with acetabular component locations assessed during and after surgery.</p><p><strong>Results: </strong>The rTHA cohort was significantly younger, but no other significant differences were found between the two cohorts in preoperative baseline data. In rTHA cohort, the planned inclination (40.0 ± 0.3°) closely matched the intraoperative (40.2 ± 2.7°; p = 0.54) and postoperative (40.7 ± 4.0°; p = 0.07) measurements. However, anteversion showed a significant increase from planned (19.4 ± 1.5°) to postoperative CT scan (28.7 ± 7.0°; p < 0.001). There was evidence of proportional bias in the measurements (p < 0.001). In the cTHA cohort, the mean inclination (43.1 ± 5.1°) did not show any significant change between the preoperative plans and postoperative assessments (p = 0.12); however, there was a remarkable change in the mean anteversion (17.6 ± 6.4°) between postoperative measurements and the preoperative plans (p < 0.001). The average anteversion in the preoperative plans did not differ remarkably between the rTHA and cTHA cohorts. However, the average inclination was substantially different between the two cohorts (p < 0.001). Both groups had no significant differences in the proportion of cups outside the referenced safe zones.</p><p><strong>Conclusion: </strong>The results suggest that while robotic-assisted guidance ensures consistent cup inclination, there may be more variability in achieving the planned anteversion, which warrants further investigation into the factors influencing postoperative changes in acetabular orientation.</p>","PeriodicalId":46378,"journal":{"name":"SICOT-J","volume":"10 ","pages":"57"},"PeriodicalIF":1.8000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"SICOT-J","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1051/sicotj/2024057","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/12/20 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Robot-assisted instrumentation during total hip arthroplasty (THA) has the potential to improve acetabular cup positioning. This study aimed to evaluate the precision of robotic-arm-assisted THA (rTHA) and assess whether the system can provide accurate cup positioning comparable to conventional THA (cTHA).

Methods: A single-center prospective cohort study consisting of 151 patients who underwent THA (108 rTHA and 43 cTHA). The robotically assisted system was used to match the postoperative computed tomography (CT) image of the pelvis with the planned and intraoperative anatomical landmarks. The cTHA cohort underwent hip replacement using the standard manual procedure, with acetabular component locations assessed during and after surgery.

Results: The rTHA cohort was significantly younger, but no other significant differences were found between the two cohorts in preoperative baseline data. In rTHA cohort, the planned inclination (40.0 ± 0.3°) closely matched the intraoperative (40.2 ± 2.7°; p = 0.54) and postoperative (40.7 ± 4.0°; p = 0.07) measurements. However, anteversion showed a significant increase from planned (19.4 ± 1.5°) to postoperative CT scan (28.7 ± 7.0°; p < 0.001). There was evidence of proportional bias in the measurements (p < 0.001). In the cTHA cohort, the mean inclination (43.1 ± 5.1°) did not show any significant change between the preoperative plans and postoperative assessments (p = 0.12); however, there was a remarkable change in the mean anteversion (17.6 ± 6.4°) between postoperative measurements and the preoperative plans (p < 0.001). The average anteversion in the preoperative plans did not differ remarkably between the rTHA and cTHA cohorts. However, the average inclination was substantially different between the two cohorts (p < 0.001). Both groups had no significant differences in the proportion of cups outside the referenced safe zones.

Conclusion: The results suggest that while robotic-assisted guidance ensures consistent cup inclination, there may be more variability in achieving the planned anteversion, which warrants further investigation into the factors influencing postoperative changes in acetabular orientation.

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
机器人辅助全髋关节置换术中髋臼杯定位的准确性:基于 CT 的评估。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 去求助
来源期刊
SICOT-J
SICOT-J ORTHOPEDICS-
CiteScore
3.20
自引率
12.50%
发文量
44
审稿时长
14 weeks
期刊最新文献
Adapting hip arthroplasty practices during the COVID-19 pandemic: Assessing the impact of outpatient care sudden increase on early complications and clinical outcomes. Outcomes of total hip arthroplasty using dual mobility cups following failed internal fixation of proximal femoral fractures at a mean follow-up of 6 years. Prior medial meniscus arthroscopy is not associated with worst functional outcomes in patients undergoing primary total knee arthroplasty: A retrospective single-center study with a minimum follow-up of 5 years. Short-term functional outcomes of robotic-assisted TKA are better with functional alignment compared to adjusted mechanical alignment. Delayed presentation of lower cervical facet dislocations: What to learn from past reports?
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1