Xinbin Chen, Huina Huang, Ying Su, Xiaojian Zhou, Menghua Liao, Yanhua Li, Zidong Cheng, Muzi Jin, Jie Tian, Haibo Wang
{"title":"Validation and evaluation of the performance of QIAstat-Dx gastrointestinal panel in pooled stool samples.","authors":"Xinbin Chen, Huina Huang, Ying Su, Xiaojian Zhou, Menghua Liao, Yanhua Li, Zidong Cheng, Muzi Jin, Jie Tian, Haibo Wang","doi":"10.1016/j.actatropica.2024.107514","DOIUrl":null,"url":null,"abstract":"<p><p>In the current study, the analytical sensitivity, analytical specificity, reproducibility, anti-interferences ability, and clinical performance of the QIAstat-Dx Gastrointestinal Panel (GIP) system were evaluated using pooled stool samples. Results showed that the pooled sample test detected the selected ten targets exclusively, with no cross reaction with any other targets of common enteropathogens. The analytical sensitivity of the pooled sample test on QIAstat-Dx GIP system was 10<sup>2</sup> CFU/ml for Shigella spp., 10<sup>3</sup> CFU/ml for Salmonella spp., Y. enterocolitica, Enterotoxigenic Escherichia coli, Enteropathogenic E. coli, 10<sup>4</sup> CFU/ml for V. cholerae, 10<sup>2</sup> copies/ml for Norovirus, 10<sup>3</sup> copies/ml for Rotavirus, Astrovirus, Sapovirus, respectively. The Coefficients of variation (CV) during the detection of V. cholerae, Salmonella spp., Y. enterocolitica, Enterotoxigenic E. coli, Enteropathogenic E. coli, Shigella spp., Rotavirus, Norovirus, Astrovirus, and Sapovirus detection was 2.3 %, 2.7 %, 3.9 %, 4.2 %, 1.7 %, 2.6 %, 6.7 %, 1.4 %, 1.3 % and 2.7 %, respectively, indicating the high reproducibility of the pooled sample test, except for Rotavirus. When potentially interfering agents were added, the shifted Ct value was less than the cut off value, suggesting the good anti-interferences ability. During clinical evaluation, the pooled sample test was 97.8 % concordant with gold standard methods (bacterial culture for bacteria and qPCR for viruses). Our results suggest that QIAstat-Dx GIP system could be used for pooled sample test for enteropathogens screening, which would be more economical and could improve throughput while provide comparable test performance.</p>","PeriodicalId":7240,"journal":{"name":"Acta tropica","volume":" ","pages":"107514"},"PeriodicalIF":2.1000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta tropica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.actatropica.2024.107514","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/12/26 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"PARASITOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
In the current study, the analytical sensitivity, analytical specificity, reproducibility, anti-interferences ability, and clinical performance of the QIAstat-Dx Gastrointestinal Panel (GIP) system were evaluated using pooled stool samples. Results showed that the pooled sample test detected the selected ten targets exclusively, with no cross reaction with any other targets of common enteropathogens. The analytical sensitivity of the pooled sample test on QIAstat-Dx GIP system was 102 CFU/ml for Shigella spp., 103 CFU/ml for Salmonella spp., Y. enterocolitica, Enterotoxigenic Escherichia coli, Enteropathogenic E. coli, 104 CFU/ml for V. cholerae, 102 copies/ml for Norovirus, 103 copies/ml for Rotavirus, Astrovirus, Sapovirus, respectively. The Coefficients of variation (CV) during the detection of V. cholerae, Salmonella spp., Y. enterocolitica, Enterotoxigenic E. coli, Enteropathogenic E. coli, Shigella spp., Rotavirus, Norovirus, Astrovirus, and Sapovirus detection was 2.3 %, 2.7 %, 3.9 %, 4.2 %, 1.7 %, 2.6 %, 6.7 %, 1.4 %, 1.3 % and 2.7 %, respectively, indicating the high reproducibility of the pooled sample test, except for Rotavirus. When potentially interfering agents were added, the shifted Ct value was less than the cut off value, suggesting the good anti-interferences ability. During clinical evaluation, the pooled sample test was 97.8 % concordant with gold standard methods (bacterial culture for bacteria and qPCR for viruses). Our results suggest that QIAstat-Dx GIP system could be used for pooled sample test for enteropathogens screening, which would be more economical and could improve throughput while provide comparable test performance.
期刊介绍:
Acta Tropica, is an international journal on infectious diseases that covers public health sciences and biomedical research with particular emphasis on topics relevant to human and animal health in the tropics and the subtropics.